Detection of incorrect manufacturer labelling of hip components
Abstract We describe the case of a 53-year-old man who underwent a left metal-on-metal hip resurfacing in 2015. Component size mismatch (CSM) was suspected because of the patient’s immediate post-operative mechanical symptoms and high metal ion levels. Surgical notes indicated the appropriate combinations of implants were used. However, we detected a mismatch using computed tomography. Revision was performed and subsequent measurements of explanted components confirmed the mismatch. To our knowledge, this case is the first report of a CT method being used in a patient to pre-operatively identify CSM..
Medienart: |
Artikel |
---|
Erscheinungsjahr: |
2016 |
---|---|
Erschienen: |
2016 |
Enthalten in: |
Zur Gesamtaufnahme - volume:46 |
---|---|
Enthalten in: |
Skeletal radiology - 46(2016), 1 vom: 22. Sept., Seite 105-109 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Durand-Hill, Matthieu [VerfasserIn] |
---|
Links: |
Volltext [lizenzpflichtig] |
---|
Themen: |
Arthroplasty |
---|
Anmerkungen: |
© The Author(s) 2016 |
---|
doi: |
10.1007/s00256-016-2478-4 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
OLC2062047428 |
---|
LEADER | 01000caa a22002652 4500 | ||
---|---|---|---|
001 | OLC2062047428 | ||
003 | DE-627 | ||
005 | 20230327090200.0 | ||
007 | tu | ||
008 | 200819s2016 xx ||||| 00| ||eng c | ||
024 | 7 | |a 10.1007/s00256-016-2478-4 |2 doi | |
035 | |a (DE-627)OLC2062047428 | ||
035 | |a (DE-He213)s00256-016-2478-4-p | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
082 | 0 | 4 | |a 610 |q VZ |
100 | 1 | |a Durand-Hill, Matthieu |e verfasserin |4 aut | |
245 | 1 | 0 | |a Detection of incorrect manufacturer labelling of hip components |
264 | 1 | |c 2016 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ohne Hilfsmittel zu benutzen |b n |2 rdamedia | ||
338 | |a Band |b nc |2 rdacarrier | ||
500 | |a © The Author(s) 2016 | ||
520 | |a Abstract We describe the case of a 53-year-old man who underwent a left metal-on-metal hip resurfacing in 2015. Component size mismatch (CSM) was suspected because of the patient’s immediate post-operative mechanical symptoms and high metal ion levels. Surgical notes indicated the appropriate combinations of implants were used. However, we detected a mismatch using computed tomography. Revision was performed and subsequent measurements of explanted components confirmed the mismatch. To our knowledge, this case is the first report of a CT method being used in a patient to pre-operatively identify CSM. | ||
650 | 4 | |a Computed tomography | |
650 | 4 | |a X-ray | |
650 | 4 | |a Arthroplasty | |
650 | 4 | |a Replacement | |
650 | 4 | |a Hip | |
650 | 4 | |a Prostheses and implants | |
650 | 4 | |a Postoperative care | |
650 | 4 | |a Hip prosthesis | |
700 | 1 | |a Henckel, Johann |4 aut | |
700 | 1 | |a Burwell, Matthew |4 aut | |
700 | 1 | |a Skinner, John |4 aut | |
700 | 1 | |a Hart, Alister |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Skeletal radiology |d Springer Berlin Heidelberg, 1976 |g 46(2016), 1 vom: 22. Sept., Seite 105-109 |w (DE-627)130184764 |w (DE-600)527592-1 |w (DE-576)015728625 |x 0364-2348 |7 nnns |
773 | 1 | 8 | |g volume:46 |g year:2016 |g number:1 |g day:22 |g month:09 |g pages:105-109 |
856 | 4 | 1 | |u https://doi.org/10.1007/s00256-016-2478-4 |z lizenzpflichtig |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a SYSFLAG_A | ||
912 | |a GBV_OLC | ||
912 | |a GBV_ILN_267 | ||
912 | |a GBV_ILN_2018 | ||
912 | |a GBV_ILN_4012 | ||
912 | |a GBV_ILN_4219 | ||
912 | |a GBV_ILN_4277 | ||
951 | |a AR | ||
952 | |d 46 |j 2016 |e 1 |b 22 |c 09 |h 105-109 |