Use of current automatic smoke evacuation system in flexible gastrointestinal endoscopy : Its feasibility and potential usefulness
© 2024 The Authors. DEN Open published by John Wiley & Sons Australia, Ltd on behalf of Japan Gastroenterological Endoscopy Society..
Objectives: Flexible endoscopy does not have a system that can automatically evacuate surgical smoke generated in the gastrointestinal lumen. We aimed to investigate the feasibility and potential usefulness of automatic smoke evacuation systems in flexible endoscopy.
Methods: [Bench] After surgical smoke generated in the stomach was evacuated by the evacuator, the degree of residual smoke and gastric luminal collapse were evaluated to optimize the evacuator settings. [Animal] Insufflation, suction, and total operation time to complete the protocol of 10 cauterizations of the gastric mucosa were measured in three groups: "manual suction only," "manual suction with automatic evacuation (50% force)," and "manual suction with automatic evacuation (70% force)." The stability of endoscopic visualization and operability was evaluated by 10 endoscopists blinded to those suction settings, and the number of manual suctions, insufflations, and total operation time were measured.
Results: [Bench] The degree of residual smoke and gastric luminal collapse were inversely correlated. [Animal] When the automatic evacuator was partially used, there was no difference in the insufflation time, but the suction time (vs 50%; p = 0.011, vs. 70%; p = 0.011) and total operation time (vs. 50%; p = 0.012, vs. 70%; p = 0.036) were significantly reduced compared to manual operation only. Furthermore, manual suction with automatic evacuation (50% force) significantly improved the stability of endoscopic visualization and operability compared to manual operation only (p = 0.041, p = 0.0085).
Conclusions: The automatic smoke evacuation in flexible gastrointestinal endoscopy was potentially feasible and useful by improving the device setting.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2024 |
---|---|
Erschienen: |
2024 |
Enthalten in: |
Zur Gesamtaufnahme - volume:4 |
---|---|
Enthalten in: |
DEN open - 4(2024), 1 vom: 11. Apr., Seite e367 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Nose, Yohei [VerfasserIn] |
---|
Links: |
---|
Themen: |
Endoscopic submucosal dissection |
---|
Anmerkungen: |
Date Revised 13.04.2024 published: Electronic-eCollection Citation Status PubMed-not-MEDLINE |
---|
doi: |
10.1002/deo2.367 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM370953169 |
---|
LEADER | 01000caa a22002652 4500 | ||
---|---|---|---|
001 | NLM370953169 | ||
003 | DE-627 | ||
005 | 20240413232908.0 | ||
007 | cr uuu---uuuuu | ||
008 | 240412s2024 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1002/deo2.367 |2 doi | |
028 | 5 | 2 | |a pubmed24n1374.xml |
035 | |a (DE-627)NLM370953169 | ||
035 | |a (NLM)38605997 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Nose, Yohei |e verfasserin |4 aut | |
245 | 1 | 0 | |a Use of current automatic smoke evacuation system in flexible gastrointestinal endoscopy |b Its feasibility and potential usefulness |
264 | 1 | |c 2024 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Revised 13.04.2024 | ||
500 | |a published: Electronic-eCollection | ||
500 | |a Citation Status PubMed-not-MEDLINE | ||
520 | |a © 2024 The Authors. DEN Open published by John Wiley & Sons Australia, Ltd on behalf of Japan Gastroenterological Endoscopy Society. | ||
520 | |a Objectives: Flexible endoscopy does not have a system that can automatically evacuate surgical smoke generated in the gastrointestinal lumen. We aimed to investigate the feasibility and potential usefulness of automatic smoke evacuation systems in flexible endoscopy | ||
520 | |a Methods: [Bench] After surgical smoke generated in the stomach was evacuated by the evacuator, the degree of residual smoke and gastric luminal collapse were evaluated to optimize the evacuator settings. [Animal] Insufflation, suction, and total operation time to complete the protocol of 10 cauterizations of the gastric mucosa were measured in three groups: "manual suction only," "manual suction with automatic evacuation (50% force)," and "manual suction with automatic evacuation (70% force)." The stability of endoscopic visualization and operability was evaluated by 10 endoscopists blinded to those suction settings, and the number of manual suctions, insufflations, and total operation time were measured | ||
520 | |a Results: [Bench] The degree of residual smoke and gastric luminal collapse were inversely correlated. [Animal] When the automatic evacuator was partially used, there was no difference in the insufflation time, but the suction time (vs 50%; p = 0.011, vs. 70%; p = 0.011) and total operation time (vs. 50%; p = 0.012, vs. 70%; p = 0.036) were significantly reduced compared to manual operation only. Furthermore, manual suction with automatic evacuation (50% force) significantly improved the stability of endoscopic visualization and operability compared to manual operation only (p = 0.041, p = 0.0085) | ||
520 | |a Conclusions: The automatic smoke evacuation in flexible gastrointestinal endoscopy was potentially feasible and useful by improving the device setting | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a endoscopic submucosal dissection | |
650 | 4 | |a flexible endoscopy | |
650 | 4 | |a infection control | |
650 | 4 | |a smoke evacuator | |
650 | 4 | |a surgical smoke | |
700 | 1 | |a Kato, Motohiko |e verfasserin |4 aut | |
700 | 1 | |a Aoyagi, Shoma |e verfasserin |4 aut | |
700 | 1 | |a Akeo, Kazunori |e verfasserin |4 aut | |
700 | 1 | |a Yamashita, Kotaro |e verfasserin |4 aut | |
700 | 1 | |a Saito, Takuro |e verfasserin |4 aut | |
700 | 1 | |a Tanaka, Koji |e verfasserin |4 aut | |
700 | 1 | |a Yamamoto, Kazuyoshi |e verfasserin |4 aut | |
700 | 1 | |a Makino, Tomoki |e verfasserin |4 aut | |
700 | 1 | |a Takahashi, Tsuyoshi |e verfasserin |4 aut | |
700 | 1 | |a Kurokawa, Yukinori |e verfasserin |4 aut | |
700 | 1 | |a Eguchi, Hidetoshi |e verfasserin |4 aut | |
700 | 1 | |a Doki, Yuichiro |e verfasserin |4 aut | |
700 | 1 | |a Nakajima, Kiyokazu |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t DEN open |d 2021 |g 4(2024), 1 vom: 11. Apr., Seite e367 |w (DE-627)NLM336141726 |x 2692-4609 |7 nnns |
773 | 1 | 8 | |g volume:4 |g year:2024 |g number:1 |g day:11 |g month:04 |g pages:e367 |
856 | 4 | 0 | |u http://dx.doi.org/10.1002/deo2.367 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 4 |j 2024 |e 1 |b 11 |c 04 |h e367 |