BACKGROUND: In recent years, mechanical power (MP) has emerged as an important concept that can significantly impact outcomes from mechanical ventilation. Several individual components of ventilatory support such as tidal volume (VT), breathing frequency, and PEEP have been shown to contribute to the extent of MP delivered from a mechanical ventilator to patients in respiratory distress/failure. The aim of this study was to identify which common individual setting of mechanical ventilation is more efficient in maintaining safe and protective levels of MP using different modes of ventilation in simulated subjects with ARDS.
METHODS: We used an interactive mathematical model of ventilator output during volume control ventilation (VCV) with either constant inspiratory flow (VCV-CF) or descending ramp inspiratory flow, as well as pressure control ventilation (PCV). MP values were determined for simulated subjects with mild, moderate, and severe ARDS; and whenever MP > 17 J/min, VT, breathing frequency, or PEEP was manipulated independently to bring back MP to ≤ 17 J/min. Finally, the optimum VT-breathing frequency combinations for MP = 17 J/min were determined with all 3 modes of ventilation.
RESULTS: VCV-CF always resulted in the lowest MPs while PCV resulted in highest MPs. Reductions in VT were the most efficient for maintaining safer and protective MP. At targeted MPs of 17 J/min and maximized minute ventilation, the optimum VT-breathing frequency combinations were 250-350 mL for VT and 32-35 breaths/min for breathing frequency in mild ARDS, 200-350 mL for VT and 34-40 breaths/min for breathing frequency in moderate ARDS, and 200-300 mL for VT and 37-45 breaths/min for breathing frequency for severe ARDS.
CONCLUSIONS: VCV-CF resulted in the lowest MP. VT was the most efficient for maintaining safe and protective MP in a mathematical simulation of subjects with ARDS. In the context of maintaining low and safe MPs, ventilatory strategies with lower-than-normal VT and higher-than-normal breathing frequency will need to be implemented in patients with ARDS.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2024 |
---|---|
Erschienen: |
2024 |
Enthalten in: |
Zur Gesamtaufnahme - volume:69 |
---|---|
Enthalten in: |
Respiratory care - 69(2024), 4 vom: 27. März, Seite 449-462 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
El-Khatib, Mohamad F [VerfasserIn] |
---|
Links: |
---|
Themen: |
ARDS |
---|
Anmerkungen: |
Date Completed 29.03.2024 Date Revised 29.03.2024 published: Electronic Citation Status MEDLINE |
---|
doi: |
10.4187/respcare.11470 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM370275594 |
---|
LEADER | 01000caa a22002652 4500 | ||
---|---|---|---|
001 | NLM370275594 | ||
003 | DE-627 | ||
005 | 20240330001856.0 | ||
007 | cr uuu---uuuuu | ||
008 | 240329s2024 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.4187/respcare.11470 |2 doi | |
028 | 5 | 2 | |a pubmed24n1355.xml |
035 | |a (DE-627)NLM370275594 | ||
035 | |a (NLM)38538014 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a El-Khatib, Mohamad F |e verfasserin |4 aut | |
245 | 1 | 0 | |a Effect of Ventilator Settings on Mechanical Power During Simulated Mechanical Ventilation of Patients With ARDS |
264 | 1 | |c 2024 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 29.03.2024 | ||
500 | |a Date Revised 29.03.2024 | ||
500 | |a published: Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a Copyright © 2024 by Daedalus Enterprises. | ||
520 | |a BACKGROUND: In recent years, mechanical power (MP) has emerged as an important concept that can significantly impact outcomes from mechanical ventilation. Several individual components of ventilatory support such as tidal volume (VT), breathing frequency, and PEEP have been shown to contribute to the extent of MP delivered from a mechanical ventilator to patients in respiratory distress/failure. The aim of this study was to identify which common individual setting of mechanical ventilation is more efficient in maintaining safe and protective levels of MP using different modes of ventilation in simulated subjects with ARDS | ||
520 | |a METHODS: We used an interactive mathematical model of ventilator output during volume control ventilation (VCV) with either constant inspiratory flow (VCV-CF) or descending ramp inspiratory flow, as well as pressure control ventilation (PCV). MP values were determined for simulated subjects with mild, moderate, and severe ARDS; and whenever MP > 17 J/min, VT, breathing frequency, or PEEP was manipulated independently to bring back MP to ≤ 17 J/min. Finally, the optimum VT-breathing frequency combinations for MP = 17 J/min were determined with all 3 modes of ventilation | ||
520 | |a RESULTS: VCV-CF always resulted in the lowest MPs while PCV resulted in highest MPs. Reductions in VT were the most efficient for maintaining safer and protective MP. At targeted MPs of 17 J/min and maximized minute ventilation, the optimum VT-breathing frequency combinations were 250-350 mL for VT and 32-35 breaths/min for breathing frequency in mild ARDS, 200-350 mL for VT and 34-40 breaths/min for breathing frequency in moderate ARDS, and 200-300 mL for VT and 37-45 breaths/min for breathing frequency for severe ARDS | ||
520 | |a CONCLUSIONS: VCV-CF resulted in the lowest MP. VT was the most efficient for maintaining safe and protective MP in a mathematical simulation of subjects with ARDS. In the context of maintaining low and safe MPs, ventilatory strategies with lower-than-normal VT and higher-than-normal breathing frequency will need to be implemented in patients with ARDS | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a ARDS | |
650 | 4 | |a breathing frequency | |
650 | 4 | |a mechanical power | |
650 | 4 | |a mechanical ventilation | |
650 | 4 | |a medical simulation | |
650 | 4 | |a tidal volume | |
700 | 1 | |a Zeinelddine, Salah M |e verfasserin |4 aut | |
700 | 1 | |a HajAli, Thuraya H |e verfasserin |4 aut | |
700 | 1 | |a Rizk, Marwan |e verfasserin |4 aut | |
700 | 1 | |a van der Staay, Matthias |e verfasserin |4 aut | |
700 | 1 | |a Chatburn, Robert L |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Respiratory care |d 1974 |g 69(2024), 4 vom: 27. März, Seite 449-462 |w (DE-627)NLM07486520X |x 1943-3654 |7 nnns |
773 | 1 | 8 | |g volume:69 |g year:2024 |g number:4 |g day:27 |g month:03 |g pages:449-462 |
856 | 4 | 0 | |u http://dx.doi.org/10.4187/respcare.11470 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 69 |j 2024 |e 4 |b 27 |c 03 |h 449-462 |