Trueness and precision of an intraoral scanner on abutments with subgingival vertical margins : An in vitro study
Copyright © 2024. Published by Elsevier Ltd..
OBJECTIVES: This study aimed to evaluate the accuracy of an intraoral scanner (IOS - Medit i700) on tooth abutments with vertical preparations at 2 depths below the free gingival margin, and to determine if the IOS can reproduce the area beyond the finish surface of the tested preparation geometry.
METHODS: Two abutments for a maxillary first molar were designed by means of CAD software, with vertical preparations set at 1 and 2 mm below the gingiva. These abutments were subsequently printed in resin and placed on a reference model. The reference files consisted of scans made using a metrological machine on these abutments. Ten scans were made with the tested IOS on each sample, resulting in two study groups. The scans from the experimental groups were labeled "V-1″ for vertical preparation at 1 mm below the gingival margin and "V-2″ for 2 mm below. The analysis of these scans was performed using Geomagic Control X (3D SYSTEMS) to assess their trueness and precision in µm. Descriptive statistics with a 95 % confidence interval were employed, alongside independent sample tests, to ascertain any differences between the groups (α=0.05).
RESULTS: Statistically significant differences were not found both for trueness (p=.104) and precision (p=.409), between the tested geometries. The mean values for trueness were V-1 = 37.5[31.4-43.6]; V-2 = 32.6[30.6-34.6]. About the precision, the mean values were V-1 = 20.5[8.4-32.5]; V-2 = 18.4[8.2-28.5]. In both the study groups, it was possible to detect the surface beyond the finish area.
CONCLUSIONS: Within the limitations of this study, vertical preparation design allows for registration of the tooth anatomy beyond the finish area with IOS. Moreover, the mean accuracy values were clinically acceptable at both 1 and 2 mm below the gingival margin.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2024 |
---|---|
Erschienen: |
2024 |
Enthalten in: |
Zur Gesamtaufnahme - volume:144 |
---|---|
Enthalten in: |
Journal of dentistry - 144(2024) vom: 16. Apr., Seite 104943 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Sorrentino, Roberto [VerfasserIn] |
---|
Links: |
---|
Themen: |
CAD |
---|
Anmerkungen: |
Date Completed 24.04.2024 Date Revised 24.04.2024 published: Print-Electronic Citation Status MEDLINE |
---|
doi: |
10.1016/j.jdent.2024.104943 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM369837282 |
---|
LEADER | 01000caa a22002652 4500 | ||
---|---|---|---|
001 | NLM369837282 | ||
003 | DE-627 | ||
005 | 20240425233202.0 | ||
007 | cr uuu---uuuuu | ||
008 | 240318s2024 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1016/j.jdent.2024.104943 |2 doi | |
028 | 5 | 2 | |a pubmed24n1386.xml |
035 | |a (DE-627)NLM369837282 | ||
035 | |a (NLM)38494043 | ||
035 | |a (PII)S0300-5712(24)00113-1 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Sorrentino, Roberto |e verfasserin |4 aut | |
245 | 1 | 0 | |a Trueness and precision of an intraoral scanner on abutments with subgingival vertical margins |b An in vitro study |
264 | 1 | |c 2024 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 24.04.2024 | ||
500 | |a Date Revised 24.04.2024 | ||
500 | |a published: Print-Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a Copyright © 2024. Published by Elsevier Ltd. | ||
520 | |a OBJECTIVES: This study aimed to evaluate the accuracy of an intraoral scanner (IOS - Medit i700) on tooth abutments with vertical preparations at 2 depths below the free gingival margin, and to determine if the IOS can reproduce the area beyond the finish surface of the tested preparation geometry | ||
520 | |a METHODS: Two abutments for a maxillary first molar were designed by means of CAD software, with vertical preparations set at 1 and 2 mm below the gingiva. These abutments were subsequently printed in resin and placed on a reference model. The reference files consisted of scans made using a metrological machine on these abutments. Ten scans were made with the tested IOS on each sample, resulting in two study groups. The scans from the experimental groups were labeled "V-1″ for vertical preparation at 1 mm below the gingival margin and "V-2″ for 2 mm below. The analysis of these scans was performed using Geomagic Control X (3D SYSTEMS) to assess their trueness and precision in µm. Descriptive statistics with a 95 % confidence interval were employed, alongside independent sample tests, to ascertain any differences between the groups (α=0.05) | ||
520 | |a RESULTS: Statistically significant differences were not found both for trueness (p=.104) and precision (p=.409), between the tested geometries. The mean values for trueness were V-1 = 37.5[31.4-43.6]; V-2 = 32.6[30.6-34.6]. About the precision, the mean values were V-1 = 20.5[8.4-32.5]; V-2 = 18.4[8.2-28.5]. In both the study groups, it was possible to detect the surface beyond the finish area | ||
520 | |a CONCLUSIONS: Within the limitations of this study, vertical preparation design allows for registration of the tooth anatomy beyond the finish area with IOS. Moreover, the mean accuracy values were clinically acceptable at both 1 and 2 mm below the gingival margin | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a CAD | |
650 | 4 | |a CAM | |
650 | 4 | |a Digital dentistry | |
650 | 4 | |a Intraoral scanning | |
650 | 4 | |a Subgingival tooth preparation | |
650 | 4 | |a Vertical tooth preparation | |
700 | 1 | |a Ruggiero, Gennaro |e verfasserin |4 aut | |
700 | 1 | |a Leone, Renato |e verfasserin |4 aut | |
700 | 1 | |a Cagidiaco, Edoardo Ferrari |e verfasserin |4 aut | |
700 | 1 | |a Mauro, Maria Irene Di |e verfasserin |4 aut | |
700 | 1 | |a Ferrari, Marco |e verfasserin |4 aut | |
700 | 1 | |a Zarone, Fernando |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Journal of dentistry |d 1972 |g 144(2024) vom: 16. Apr., Seite 104943 |w (DE-627)NLM000213578 |x 1879-176X |7 nnns |
773 | 1 | 8 | |g volume:144 |g year:2024 |g day:16 |g month:04 |g pages:104943 |
856 | 4 | 0 | |u http://dx.doi.org/10.1016/j.jdent.2024.104943 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 144 |j 2024 |b 16 |c 04 |h 104943 |