Test-retest reliability and concurrent validity assessment of a novel high-frequency sensor device for anterior tibial translation measurement in loaded and unloaded condition : an exploratory cross-sectional study

© 2024. The Author(s)..

BACKGROUND: Magnetic resonance imaging (MRI) and manual tests remain the standard for diagnosing anterior cruciate ligament (ACL) rupture. Furthermore, the passive knee displacement, also described as anterior tibial translation (ATT), is used in order to make decisions about surgery or to assess rehabilitation outcomes. Unfortunately, these manual tests are limited to passive situations, and their application to assess knee stability in loaded, weight-bearing positions are missing. Therefore, a new device with high-performance sensors and a new sensor setting was developed. The aim of this exploratory cross-sectional study was to assess the test-retest reliability of this new device in a first step and the concurrent validity in a second step.

METHODS: A total of 20 healthy volunteers were measured. Measurement consistency of the new device was assessed on the basis of reliability during Lachman test setting and in loaded position by artificial knee perturbation in a test-retest procedure. In a second step, the concurrent validity was evaluated with the Lachmeter® as a reference instrument. Intraclass correlation coefficient (ICC), standard error of measurement (SEM), the minimal detectable change (MDC) and Bland-Altman analysis were evaluated to assess the quality criteria.

RESULTS: The measurements with the new device during the Lachman test provided a mean ATT of 5.46±2.22mm. The SEM ranged from 0.60 to 0.69mm resulting in an MDC between 1.67 and 1.93mm for the new device. In the loaded test situation, the mean ATT was 2.11±1.20mm, with test-retest reliability also showing good correlation (r>0.83). The comparison of the two measurement methods with an ICC of (r>0.89) showed good correlation, which also underlines the reasonable agreement of the Bland-Altman analysis.

CONCLUSIONS: The evaluation of the test-retest reliability of the new device during the knee stability testing in passive situation as well as in a functional, loaded situation presented good reliability. In addition, the new device demonstrated good agreement with the reference device and therefore good validity. Furthermore, the quality criteria demonstrated the ability of the new device to detect the cut-off value (3-5mm) described in the literature for the diagnosis of ACL-deficient knees, which underlines the clinical relevance of this new device as a reliable and valid tool.

Medienart:

E-Artikel

Erscheinungsjahr:

2024

Erschienen:

2024

Enthalten in:

Zur Gesamtaufnahme - volume:25

Enthalten in:

BMC musculoskeletal disorders - 25(2024), 1 vom: 16. März, Seite 218

Sprache:

Englisch

Beteiligte Personen:

Deiss, Valentin [VerfasserIn]
Bähler, Philippe [VerfasserIn]
Kolly, Pascal [VerfasserIn]
Schärer, Anton [VerfasserIn]
Henle, Philippe [VerfasserIn]
Eichelberger, Patric [VerfasserIn]
Lutz, Nathanael [VerfasserIn]
Baur, Heiner [VerfasserIn]

Links:

Volltext

Themen:

ACL injury
Anterior tibial translation
Functional joint stability
High frequency sensor device
Journal Article
Knee kinematics
Reproducibility of results

Anmerkungen:

Date Completed 18.03.2024

Date Revised 19.03.2024

published: Electronic

Citation Status MEDLINE

doi:

10.1186/s12891-024-07343-y

funding:

Förderinstitution / Projekttitel:

PPN (Katalog-ID):

NLM369810856