Evaluating the in vitro wettability and coefficient of friction of a novel and contemporary reusable silicone hydrogel contact lens materials using an in vitro blink model
Copyright © 2024 The Author(s). Published by Elsevier Ltd.. All rights reserved..
PURPOSE: To evaluate the in vitro wettability and coefficient of friction of a novel amphiphilic polymeric surfactant (APS), poly(oxyethylene)-co-poly(oxybutylene) (PEO-PBO) releasing silicone hydrogel (SiHy) contact lens material (serafilcon A), compared to other reusable SiHy lens materials.
METHODS: The release of fluorescently-labelled nitrobenzoxadiazole (NBD)-PEO-PBO was evaluated from serafilcon A over 7 days in a vial. The wettability and coefficient of friction of serafilcon A and three contemporary SiHy contact lens materials (senofilcon A; samfilcon A; comfilcon A) were evaluated using an in vitro blink model over their recommended wearing period; t = 0, 1, 7, 14 days for all lens types and t = 30 days for samfilcon A and comfilcon A (n = 4). Sessile drop contact angles were determined and in vitro non-invasive keratographic break-up time (NIKBUT) measurements were assessed on a blink model via the OCULUS Keratograph 5 M. The coefficient of friction was measured using a nano tribometer.
RESULTS: The relative fluorescence of NBD-PEO-PBO decreased in serafilcon A by approximately 18 % after 7 days. The amount of NBD-PEO-PBO released on day 7 was 50 % less than the amount released on day 1 (6.5±1.0 vs 3.4±0.5 µg/lens). The reduction in PEO-PBO in the lens also coincided with an increase in contact angles for serafilcon A after 7 days (p < 0.05), although there were no changes in NIKBUT or coefficient of friction (p > 0.05). The other contact lens materials had stable contact angles and NIKBUT over their recommended wearing period (p > 0.05), with the exception of samfilcon A, which had an increase in contact angle after 14 days as compared to t = 0 (p < 0.05). Senofilcon A and samfilcon A also showed an increase in coefficient of friction at 14 and 30 days, respectively, compared to their blister pack values (p < 0.05).
CONCLUSION: The results indicate that serafilcon A gradually depletes its reserve of PEO-PBO over 1 week, but this decrease did not significantly change the lens performance in vitro during this time frame.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2024 |
---|---|
Erschienen: |
2024 |
Enthalten in: |
Zur Gesamtaufnahme - volume:47 |
---|---|
Enthalten in: |
Contact lens & anterior eye : the journal of the British Contact Lens Association - 47(2024), 2 vom: 29. März, Seite 102129 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Phan, Chau-Minh [VerfasserIn] |
---|
Links: |
---|
Themen: |
---|
Anmerkungen: |
Date Completed 25.03.2024 Date Revised 25.03.2024 published: Print-Electronic Citation Status MEDLINE |
---|
doi: |
10.1016/j.clae.2024.102129 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM369137639 |
---|
LEADER | 01000caa a22002652 4500 | ||
---|---|---|---|
001 | NLM369137639 | ||
003 | DE-627 | ||
005 | 20240325235016.0 | ||
007 | cr uuu---uuuuu | ||
008 | 240301s2024 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1016/j.clae.2024.102129 |2 doi | |
028 | 5 | 2 | |a pubmed24n1346.xml |
035 | |a (DE-627)NLM369137639 | ||
035 | |a (NLM)38423868 | ||
035 | |a (PII)S1367-0484(24)00012-2 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Phan, Chau-Minh |e verfasserin |4 aut | |
245 | 1 | 0 | |a Evaluating the in vitro wettability and coefficient of friction of a novel and contemporary reusable silicone hydrogel contact lens materials using an in vitro blink model |
264 | 1 | |c 2024 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 25.03.2024 | ||
500 | |a Date Revised 25.03.2024 | ||
500 | |a published: Print-Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a Copyright © 2024 The Author(s). Published by Elsevier Ltd.. All rights reserved. | ||
520 | |a PURPOSE: To evaluate the in vitro wettability and coefficient of friction of a novel amphiphilic polymeric surfactant (APS), poly(oxyethylene)-co-poly(oxybutylene) (PEO-PBO) releasing silicone hydrogel (SiHy) contact lens material (serafilcon A), compared to other reusable SiHy lens materials | ||
520 | |a METHODS: The release of fluorescently-labelled nitrobenzoxadiazole (NBD)-PEO-PBO was evaluated from serafilcon A over 7 days in a vial. The wettability and coefficient of friction of serafilcon A and three contemporary SiHy contact lens materials (senofilcon A; samfilcon A; comfilcon A) were evaluated using an in vitro blink model over their recommended wearing period; t = 0, 1, 7, 14 days for all lens types and t = 30 days for samfilcon A and comfilcon A (n = 4). Sessile drop contact angles were determined and in vitro non-invasive keratographic break-up time (NIKBUT) measurements were assessed on a blink model via the OCULUS Keratograph 5 M. The coefficient of friction was measured using a nano tribometer | ||
520 | |a RESULTS: The relative fluorescence of NBD-PEO-PBO decreased in serafilcon A by approximately 18 % after 7 days. The amount of NBD-PEO-PBO released on day 7 was 50 % less than the amount released on day 1 (6.5±1.0 vs 3.4±0.5 µg/lens). The reduction in PEO-PBO in the lens also coincided with an increase in contact angles for serafilcon A after 7 days (p < 0.05), although there were no changes in NIKBUT or coefficient of friction (p > 0.05). The other contact lens materials had stable contact angles and NIKBUT over their recommended wearing period (p > 0.05), with the exception of samfilcon A, which had an increase in contact angle after 14 days as compared to t = 0 (p < 0.05). Senofilcon A and samfilcon A also showed an increase in coefficient of friction at 14 and 30 days, respectively, compared to their blister pack values (p < 0.05) | ||
520 | |a CONCLUSION: The results indicate that serafilcon A gradually depletes its reserve of PEO-PBO over 1 week, but this decrease did not significantly change the lens performance in vitro during this time frame | ||
650 | 4 | |a Journal Article | |
650 | 7 | |a Silicones |2 NLM | |
650 | 7 | |a Hydrogels |2 NLM | |
700 | 1 | |a Wy Chan, Vivian |e verfasserin |4 aut | |
700 | 1 | |a Drolle, Elizabeth |e verfasserin |4 aut | |
700 | 1 | |a Hui, Alex |e verfasserin |4 aut | |
700 | 1 | |a Ngo, William |e verfasserin |4 aut | |
700 | 1 | |a Bose, Susmita |e verfasserin |4 aut | |
700 | 1 | |a Shows, Amanda |e verfasserin |4 aut | |
700 | 1 | |a Liang, Shuang |e verfasserin |4 aut | |
700 | 1 | |a Sharma, Vinay |e verfasserin |4 aut | |
700 | 1 | |a Subbaraman, Lakshman |e verfasserin |4 aut | |
700 | 1 | |a Zheng, Ying |e verfasserin |4 aut | |
700 | 1 | |a Shi, Xinfeng |e verfasserin |4 aut | |
700 | 1 | |a Wu, James |e verfasserin |4 aut | |
700 | 1 | |a Jones, Lyndon |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Contact lens & anterior eye : the journal of the British Contact Lens Association |d 1997 |g 47(2024), 2 vom: 29. März, Seite 102129 |w (DE-627)NLM096520159 |x 1476-5411 |7 nnns |
773 | 1 | 8 | |g volume:47 |g year:2024 |g number:2 |g day:29 |g month:03 |g pages:102129 |
856 | 4 | 0 | |u http://dx.doi.org/10.1016/j.clae.2024.102129 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 47 |j 2024 |e 2 |b 29 |c 03 |h 102129 |