Delivery catheter system carries more physiological right ventricular septal pacing than stylet system
© 2024 Wiley Periodicals LLC..
INTRODUCTION: The Mt. FUJI multicenter trial demonstrated that a delivery catheter system had a higher rate of successful right ventricular (RV) lead deployment on the RV septum (RVS) than a conventional stylet system. In this subanalysis of the Mt. FUJI trial, we assessed the differences in electrocardiogram (ECG) parameters during RV pacing between a delivery catheter system and a stylet system and their associations with the lead tip positions.
METHODS: Among 70 patients enrolled in the Mt FUJI trial, ECG parameters, RV lead tip positions, and lead depth inside the septum assessed by computed tomography were compared between the catheter group (n = 36) and stylet group (n = 34).
RESULTS: The paced QRS duration (QRS-d), corrected paced QT (QTc), and JT interval (JTc) were significantly shorter in the catheter group than in the stylet group (QRS-d: 130 ± 19 vs. 142 ± 15 ms, p = .004; QTc: 476 ± 25 vs. 514 ± 20 ms, p < .001; JTc: 347 ± 24 vs. 372 ± 17 ms, p < .001). This superiority of the catheter group was maintained in a subgroup analysis of patients with an RV lead tip position at the septum. The lead depth inside the septum was greater in the catheter group than in the stylet group, and there was a significant negative correlation between the paced QRS-d and the lead depth.
CONCLUSION: Using a delivery catheter system carries more physiological depolarization and repolarization during RVS pacing and deeper screw penetration in the septum in comparison to conventional stylet system. The lead depth could have a more impact on the ECG parameters rather than the type of pacing lead.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2024 |
---|---|
Erschienen: |
2024 |
Enthalten in: |
Zur Gesamtaufnahme - volume:35 |
---|---|
Enthalten in: |
Journal of cardiovascular electrophysiology - 35(2024), 4 vom: 26. Apr., Seite 802-810 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Hasebe, Hideyuki [VerfasserIn] |
---|
Links: |
---|
Themen: |
Clinical Trial |
---|
Anmerkungen: |
Date Completed 10.04.2024 Date Revised 25.04.2024 published: Print-Electronic Citation Status MEDLINE |
---|
doi: |
10.1111/jce.16226 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM368998606 |
---|
LEADER | 01000caa a22002652 4500 | ||
---|---|---|---|
001 | NLM368998606 | ||
003 | DE-627 | ||
005 | 20240425233000.0 | ||
007 | cr uuu---uuuuu | ||
008 | 240229s2024 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1111/jce.16226 |2 doi | |
028 | 5 | 2 | |a pubmed24n1386.xml |
035 | |a (DE-627)NLM368998606 | ||
035 | |a (NLM)38409896 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Hasebe, Hideyuki |e verfasserin |4 aut | |
245 | 1 | 0 | |a Delivery catheter system carries more physiological right ventricular septal pacing than stylet system |
264 | 1 | |c 2024 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 10.04.2024 | ||
500 | |a Date Revised 25.04.2024 | ||
500 | |a published: Print-Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a © 2024 Wiley Periodicals LLC. | ||
520 | |a INTRODUCTION: The Mt. FUJI multicenter trial demonstrated that a delivery catheter system had a higher rate of successful right ventricular (RV) lead deployment on the RV septum (RVS) than a conventional stylet system. In this subanalysis of the Mt. FUJI trial, we assessed the differences in electrocardiogram (ECG) parameters during RV pacing between a delivery catheter system and a stylet system and their associations with the lead tip positions | ||
520 | |a METHODS: Among 70 patients enrolled in the Mt FUJI trial, ECG parameters, RV lead tip positions, and lead depth inside the septum assessed by computed tomography were compared between the catheter group (n = 36) and stylet group (n = 34) | ||
520 | |a RESULTS: The paced QRS duration (QRS-d), corrected paced QT (QTc), and JT interval (JTc) were significantly shorter in the catheter group than in the stylet group (QRS-d: 130 ± 19 vs. 142 ± 15 ms, p = .004; QTc: 476 ± 25 vs. 514 ± 20 ms, p < .001; JTc: 347 ± 24 vs. 372 ± 17 ms, p < .001). This superiority of the catheter group was maintained in a subgroup analysis of patients with an RV lead tip position at the septum. The lead depth inside the septum was greater in the catheter group than in the stylet group, and there was a significant negative correlation between the paced QRS-d and the lead depth | ||
520 | |a CONCLUSION: Using a delivery catheter system carries more physiological depolarization and repolarization during RVS pacing and deeper screw penetration in the septum in comparison to conventional stylet system. The lead depth could have a more impact on the ECG parameters rather than the type of pacing lead | ||
650 | 4 | |a Clinical Trial | |
650 | 4 | |a Journal Article | |
650 | 4 | |a Multicenter Study | |
650 | 4 | |a delivery catheter | |
650 | 4 | |a depolarization | |
650 | 4 | |a repolarization | |
650 | 4 | |a right ventricular septal pacing | |
700 | 1 | |a Naruse, Yoshihisa |e verfasserin |4 aut | |
700 | 1 | |a Sano, Makoto |e verfasserin |4 aut | |
700 | 1 | |a Kurebayashi, Nobutake |e verfasserin |4 aut | |
700 | 1 | |a Morikawa, Shuji |e verfasserin |4 aut | |
700 | 1 | |a Tsurumi, Naoki |e verfasserin |4 aut | |
700 | 1 | |a Shiozawa, Tomoyuki |e verfasserin |4 aut | |
700 | 1 | |a Takano, Shintaro |e verfasserin |4 aut | |
700 | 1 | |a Ogano, Michio |e verfasserin |4 aut | |
700 | 1 | |a Kimura, Kei |e verfasserin |4 aut | |
700 | 1 | |a Miyajima, Keisuke |e verfasserin |4 aut | |
700 | 1 | |a Sugiura, Ryo |e verfasserin |4 aut | |
700 | 1 | |a Henmi, Ryuta |e verfasserin |4 aut | |
700 | 1 | |a Muto, Masahiro |e verfasserin |4 aut | |
700 | 1 | |a Hosoya, Natsuko |e verfasserin |4 aut | |
700 | 1 | |a Mizukami, Akira |e verfasserin |4 aut | |
700 | 1 | |a Iguchi, Keisuke |e verfasserin |4 aut | |
700 | 1 | |a Atsumi, Akiko |e verfasserin |4 aut | |
700 | 1 | |a Odagiri, Keiichi |e verfasserin |4 aut | |
700 | 1 | |a Kiyama, Yumi |e verfasserin |4 aut | |
700 | 1 | |a Maekawa, Yuichiro |e verfasserin |4 aut | |
700 | 0 | |a Mt. FUJI trial investigators |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Journal of cardiovascular electrophysiology |d 1996 |g 35(2024), 4 vom: 26. Apr., Seite 802-810 |w (DE-627)NLM074905481 |x 1540-8167 |7 nnns |
773 | 1 | 8 | |g volume:35 |g year:2024 |g number:4 |g day:26 |g month:04 |g pages:802-810 |
856 | 4 | 0 | |u http://dx.doi.org/10.1111/jce.16226 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 35 |j 2024 |e 4 |b 26 |c 04 |h 802-810 |