Closure of patent ductus arteriosus with OcclutechTM device, experience in an andean country
Background and Objective: To communicate the experience in an Andean country with the OcclutechTM Duct Occluder device for the closure of patent ductus arteriosus.
Method: observational, retrospective, cross-sectional study with basic statistical analysis. Period: December/2014 to December/2022. Data: medical chart, reports of catheterization.
Results: Forty-six patients, female 71.3%, male 28.7%; age: 0.6-38 years-old (median [Me]: 5.2); weight: 6.3-60 kg (Me: 16.5). Origin: Andean 91.3%, coast 8.7%. Types of patent ductus arteriosus: E 54.4%, A 32.6%, D 13%. Minimum ductal diameter: 1.8-11.8 mm (Me: 3.5). Mean pulmonary artery pressure prior to occlusion: 14-67 mmHg (Me: 27). Pulmonary vascular resistance index prior to occlusion: 0.28-4.9 WU/m2 (Me: 1.3). Six of them were classified as hypertensive patent ductus arteriosus. Occlusion rate: 47.8% immediate, 81% at 24 hours, 100% after six months. Fluoroscopy time: 2-13.8 minutes (Me: 4). Complications: a migrated device. Follow-up: 1-6.5 years.
Conclusions: OcclutechTM Duct Occluder device was effective and safe for the closure of patent ductus arteriosus type E, A and D in low-altitude and high-altitude dwellers, whether they were children or adults, even when these ductus arteriosus were hypertensive.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2024 |
---|---|
Erschienen: |
2024 |
Enthalten in: |
Zur Gesamtaufnahme - volume:94 |
---|---|
Enthalten in: |
Archivos de cardiologia de Mexico - 94(2024), 2 vom: 15. Feb., Seite 169-173 |
Sprache: |
Spanisch |
---|
Weiterer Titel: |
Cierre de conducto arterioso persistente con dispositivo OcclutechTM, experiencia en un país andino |
---|
Beteiligte Personen: |
Ríos-Méndez, Raúl E [VerfasserIn] |
---|
Links: |
---|
Themen: |
Andean |
---|
Anmerkungen: |
Date Revised 22.04.2024 published: Electronic Citation Status In-Process |
---|
doi: |
10.24875/ACM.23000126 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM368495159 |
---|
LEADER | 01000caa a22002652 4500 | ||
---|---|---|---|
001 | NLM368495159 | ||
003 | DE-627 | ||
005 | 20240423232137.0 | ||
007 | cr uuu---uuuuu | ||
008 | 240216s2024 xx |||||o 00| ||spa c | ||
024 | 7 | |a 10.24875/ACM.23000126 |2 doi | |
028 | 5 | 2 | |a pubmed24n1384.xml |
035 | |a (DE-627)NLM368495159 | ||
035 | |a (NLM)38359430 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a spa | ||
100 | 1 | |a Ríos-Méndez, Raúl E |e verfasserin |4 aut | |
245 | 1 | 0 | |a Closure of patent ductus arteriosus with OcclutechTM device, experience in an andean country |
246 | 3 | 3 | |a Cierre de conducto arterioso persistente con dispositivo OcclutechTM, experiencia en un país andino |
264 | 1 | |c 2024 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Revised 22.04.2024 | ||
500 | |a published: Electronic | ||
500 | |a Citation Status In-Process | ||
520 | |a Background and Objective: To communicate the experience in an Andean country with the OcclutechTM Duct Occluder device for the closure of patent ductus arteriosus | ||
520 | |a Method: observational, retrospective, cross-sectional study with basic statistical analysis. Period: December/2014 to December/2022. Data: medical chart, reports of catheterization | ||
520 | |a Results: Forty-six patients, female 71.3%, male 28.7%; age: 0.6-38 years-old (median [Me]: 5.2); weight: 6.3-60 kg (Me: 16.5). Origin: Andean 91.3%, coast 8.7%. Types of patent ductus arteriosus: E 54.4%, A 32.6%, D 13%. Minimum ductal diameter: 1.8-11.8 mm (Me: 3.5). Mean pulmonary artery pressure prior to occlusion: 14-67 mmHg (Me: 27). Pulmonary vascular resistance index prior to occlusion: 0.28-4.9 WU/m2 (Me: 1.3). Six of them were classified as hypertensive patent ductus arteriosus. Occlusion rate: 47.8% immediate, 81% at 24 hours, 100% after six months. Fluoroscopy time: 2-13.8 minutes (Me: 4). Complications: a migrated device. Follow-up: 1-6.5 years | ||
520 | |a Conclusions: OcclutechTM Duct Occluder device was effective and safe for the closure of patent ductus arteriosus type E, A and D in low-altitude and high-altitude dwellers, whether they were children or adults, even when these ductus arteriosus were hypertensive | ||
650 | 4 | |a English Abstract | |
650 | 4 | |a Journal Article | |
650 | 4 | |a Andean | |
650 | 4 | |a Catheterization | |
650 | 4 | |a High altitude | |
650 | 4 | |a Patent ductus arteriosus | |
650 | 4 | |a Pulmonary hypertension | |
700 | 1 | |a Araúz-Martínez, María E |e verfasserin |4 aut | |
700 | 1 | |a Oliveros-Rivero, Jorge A |e verfasserin |4 aut | |
700 | 1 | |a Crespo-Gutiérrez, Yeiber J |e verfasserin |4 aut | |
700 | 1 | |a Pérez-Vite, Yesenia W |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Archivos de cardiologia de Mexico |d 2001 |g 94(2024), 2 vom: 15. Feb., Seite 169-173 |w (DE-627)NLM114616272 |x 1665-1731 |7 nnns |
773 | 1 | 8 | |g volume:94 |g year:2024 |g number:2 |g day:15 |g month:02 |g pages:169-173 |
856 | 4 | 0 | |u http://dx.doi.org/10.24875/ACM.23000126 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 94 |j 2024 |e 2 |b 15 |c 02 |h 169-173 |