Development and Clinical Performance of InteliSwab® COVID-19 Rapid Test : Evaluation of Antigen Test for the Diagnosis of SARS-CoV-2 and Analytical Sensitivity to Detect Variants of Concern Including Omicron and Subvariants

BACKGROUND AND OBJECTIVES: Timely detection of SARS-CoV-2 infection with subsequent contact tracing and rapid isolation are considered critical to containing the pandemic, which continues with the emergence of new variants. Hence, there is an ongoing need for reliable point-of-care antigen rapid diagnostic tests (Ag-RDT). This report describes the development, evaluation, and analytical sensitivity of the diagnostic performance of the InteliSwab® COVID-19 Rapid Test. Methods: Samples from 165 symptomatic subjects were tested with InteliSwab® and the results were compared to RT-PCR to determine the antigen test performance. The analytical sensitivity of InteliSwab® for the detection of different variants was assessed by limit of detection (LOD) determination using recombinant nucleocapsid proteins (NPs) and testing with virus isolates. Western immunoblot independently confirmed that each monoclonal Ab is capable of binding to all variants tested thus far.

RESULTS: The overall positivity rate by RT-PCR was 37% for the 165 symptomatic subjects. Based on RT-PCR results as the reference standard, InteliSwab® showed clinical sensitivity and specificity of 85.2% (95% CI, 74.3-92.0%) and 98.1% (95% CI, 93.3-99.7%), respectively. The overall agreement was 93.3% (Kappa index value 0.85; 95% CI, 0.77-0.74) between RT-PCR and InteliSwab® test results. Furthermore, the evaluation of analytical sensitivity for different SARS-CoV-2 variants by InteliSwab® was comparable in the detection of all the variants tested, including Omicron subvariants, BA.4, BA.5, and BQ.1.

CONCLUSIONS: Due to the surge of infections caused by different variants from time to time, there is a critical need to evaluate the sensitivity of rapid antigen-detecting tests for new variants. The study findings showed the robust diagnostic performance of InteliSwab® and analytical sensitivity in detecting different SARS-CoV-2 variants, including the Omicron subvariants. With the integrated swab and excellent sensitivity and variant detection, this test has high potential as a point-of-care Ag-RDT in various settings when molecular assays are in limited supply and rapid diagnosis of SARS-CoV-2 is necessary.

Medienart:

E-Artikel

Erscheinungsjahr:

2023

Erschienen:

2023

Enthalten in:

Zur Gesamtaufnahme - volume:16

Enthalten in:

Viruses - 16(2023), 1 vom: 29. Dez.

Sprache:

Englisch

Beteiligte Personen:

Fischl, Mark J [VerfasserIn]
Young, Janean [VerfasserIn]
Kardos, Keith [VerfasserIn]
Roehler, Michele [VerfasserIn]
Miller, Tiffany [VerfasserIn]
Wooten, Melinda [VerfasserIn]
Holmes, Natalie [VerfasserIn]
Gula, Nicole [VerfasserIn]
Baglivo, Mia [VerfasserIn]
Steen, Justin [VerfasserIn]
Zelenz, Nori [VerfasserIn]
Joyee, Antony George [VerfasserIn]
Munster, Vincent [VerfasserIn]
Weishampel, Zack [VerfasserIn]
Yinda, Claude Kwe [VerfasserIn]
Rouse, Kevin G [VerfasserIn]
Gvozden, Cathy [VerfasserIn]
Wever, David [VerfasserIn]
Yanez, Giralt [VerfasserIn]
Anderson, Marc [VerfasserIn]
Yu, Song [VerfasserIn]
Bearie, Brian [VerfasserIn]
Young, Stephen [VerfasserIn]
Berry, Jody D [VerfasserIn]

Links:

Volltext

Themen:

COVID-19
Diagnosis
Journal Article
Omicron
Rapid antigen test
SARS-CoV-2
Variants of concern

Anmerkungen:

Date Completed 24.01.2024

Date Revised 29.01.2024

published: Electronic

Citation Status MEDLINE

doi:

10.3390/v16010061

funding:

Förderinstitution / Projekttitel:

PPN (Katalog-ID):

NLM367482479