Variety of femoral anteversion and its measurement in cementless total hip arthroplasty : Does robotic technology improve accuracy?
© 2024. The Author(s)..
BACKGROUND: High-performance total hip arthroplasty (THA) depends on the accurate position of components. However, femoral anteversion is variable, and current studies only used traditional instruments to evaluate it, such as protractor and spirit level with limited cases. This study aimed to identify the variability in the measured femoral native anteversion and intraoperative stem anteversion under different measurement methods, including intraoperative robotic method. We hypothesized that robotic technology was more accurate than traditional instruments for femoral anteversion evaluation.
METHODS: This study included 117 hips of patients who underwent robotic-assisted THA between November 2019 and March 2021. Preoperative native femoral anteversion was measured using a robotic system. Intraoperative femoral stem anteversion was evaluated visually, and then measured with a goniometer and a robotic system, respectively. Variability in the measured femoral native anteversion and intraoperative femoral stem anteversion was calculated and compared. Intraclass correlation coefficient (ICC) and Pearson correlation analysis were used to assess the consistency and correlation of anteversion of different measurements and postoperative CT-measured stem anteversion, respectively.
RESULTS: The result of measurement for preoperative native femoral anteversion was more variable than the intraoperative robotic-measured stem anteversion. Intraoperative robotic-measured stem version showed the highest correlation with postoperative CT measurement of stem version (r = 0.806, P < 0.001), while intraoperative surgeon estimation had the lowest correlation coefficient (r = 0.281, P = 0.025). As for the consistency with postoperative CT measurement of femoral stem anteversion, the intraoperative robotic-measured femoral stem version also had the highest value (ICC = 0.892, P < 0.001).
CONCLUSION: Native femoral anteversion was variable preoperatively. Using cementless stems, anteversion was also highly variable. Robotic assessment for stem anteversion during surgery was more consistent with the final position than the preoperative assessment and conventional intraoperative estimation.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2024 |
---|---|
Erschienen: |
2024 |
Enthalten in: |
Zur Gesamtaufnahme - volume:19 |
---|---|
Enthalten in: |
Journal of orthopaedic surgery and research - 19(2024), 1 vom: 09. Jan., Seite 47 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Shao, Hongyi [VerfasserIn] |
---|
Links: |
---|
Themen: |
Cementless total hip arthroplasty |
---|
Anmerkungen: |
Date Completed 11.01.2024 Date Revised 11.01.2024 published: Electronic Citation Status MEDLINE |
---|
doi: |
10.1186/s13018-024-04527-z |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM366861093 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM366861093 | ||
003 | DE-627 | ||
005 | 20240114233935.0 | ||
007 | cr uuu---uuuuu | ||
008 | 240114s2024 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1186/s13018-024-04527-z |2 doi | |
028 | 5 | 2 | |a pubmed24n1256.xml |
035 | |a (DE-627)NLM366861093 | ||
035 | |a (NLM)38195511 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Shao, Hongyi |e verfasserin |4 aut | |
245 | 1 | 0 | |a Variety of femoral anteversion and its measurement in cementless total hip arthroplasty |b Does robotic technology improve accuracy? |
264 | 1 | |c 2024 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 11.01.2024 | ||
500 | |a Date Revised 11.01.2024 | ||
500 | |a published: Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a © 2024. The Author(s). | ||
520 | |a BACKGROUND: High-performance total hip arthroplasty (THA) depends on the accurate position of components. However, femoral anteversion is variable, and current studies only used traditional instruments to evaluate it, such as protractor and spirit level with limited cases. This study aimed to identify the variability in the measured femoral native anteversion and intraoperative stem anteversion under different measurement methods, including intraoperative robotic method. We hypothesized that robotic technology was more accurate than traditional instruments for femoral anteversion evaluation | ||
520 | |a METHODS: This study included 117 hips of patients who underwent robotic-assisted THA between November 2019 and March 2021. Preoperative native femoral anteversion was measured using a robotic system. Intraoperative femoral stem anteversion was evaluated visually, and then measured with a goniometer and a robotic system, respectively. Variability in the measured femoral native anteversion and intraoperative femoral stem anteversion was calculated and compared. Intraclass correlation coefficient (ICC) and Pearson correlation analysis were used to assess the consistency and correlation of anteversion of different measurements and postoperative CT-measured stem anteversion, respectively | ||
520 | |a RESULTS: The result of measurement for preoperative native femoral anteversion was more variable than the intraoperative robotic-measured stem anteversion. Intraoperative robotic-measured stem version showed the highest correlation with postoperative CT measurement of stem version (r = 0.806, P < 0.001), while intraoperative surgeon estimation had the lowest correlation coefficient (r = 0.281, P = 0.025). As for the consistency with postoperative CT measurement of femoral stem anteversion, the intraoperative robotic-measured femoral stem version also had the highest value (ICC = 0.892, P < 0.001) | ||
520 | |a CONCLUSION: Native femoral anteversion was variable preoperatively. Using cementless stems, anteversion was also highly variable. Robotic assessment for stem anteversion during surgery was more consistent with the final position than the preoperative assessment and conventional intraoperative estimation | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a Cementless total hip arthroplasty | |
650 | 4 | |a Femoral anteversion | |
650 | 4 | |a Robotic-assisted THA | |
650 | 4 | |a Stem anteversion | |
700 | 1 | |a Huang, Yong |e verfasserin |4 aut | |
700 | 1 | |a Yang, Dejin |e verfasserin |4 aut | |
700 | 1 | |a Deng, Wang |e verfasserin |4 aut | |
700 | 1 | |a Wu, Xiang-Dong |e verfasserin |4 aut | |
700 | 1 | |a Zhou, Yixin |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Journal of orthopaedic surgery and research |d 2006 |g 19(2024), 1 vom: 09. Jan., Seite 47 |w (DE-627)NLM167010948 |x 1749-799X |7 nnns |
773 | 1 | 8 | |g volume:19 |g year:2024 |g number:1 |g day:09 |g month:01 |g pages:47 |
856 | 4 | 0 | |u http://dx.doi.org/10.1186/s13018-024-04527-z |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 19 |j 2024 |e 1 |b 09 |c 01 |h 47 |