Improved test-retest reliability of

© 2023 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals LLC on behalf of International Society for Magnetic Resonance in Medicine..

PURPOSE: This study aimed to evaluate the potential of 3D EPI for improving the reliability of T 2 * $$ {\mathrm{T}}_2^{\ast } $$ -weighted data and quantification of R 2 * $$ {\mathrm{R}}_2^{\ast } $$ decay rate and susceptibility (χ) compared with conventional gradient-echo (GRE)-based acquisition.

METHODS: Eight healthy subjects in a wide age range were recruited. Each subject received repeated scans for both GRE and EPI acquisitions with an isotropic 1 mm resolution at 3 T. Maps of R 2 * $$ {\mathrm{R}}_2^{\ast } $$ and χ were quantified, and their interscan differences were used to evaluate the test-retest reliability. Interprotocol differences of R 2 * $$ {\mathrm{R}}_2^{\ast } $$ and χ between GRE and EPI were also measured voxel by voxel and in selected regions of interest to test the consistency between the two acquisition methods.

RESULTS: The quantifications of R 2 * $$ {\mathrm{R}}_2^{\ast } $$ and χ using EPI protocols showed increased test-retest reliability with higher EPI factors up to 5 as performed in the experiment and were consistent with those based on GRE.

CONCLUSION: The result suggests that multishot multi-echo 3D EPI can be a useful alternative acquisition method for T 2 * $$ {\mathrm{T}}_2^{\ast } $$ -weighted MRI and quantification of R 2 * $$ {\mathrm{R}}_2^{\ast } $$ and χ with reduced scan time, improved test-retest reliability, and similar accuracy compared with commonly used 3D GRE.

Errataetall:

UpdateOf: ArXiv. 2023 Aug 15;:. - PMID 37645047

Medienart:

E-Artikel

Erscheinungsjahr:

2024

Erschienen:

2024

Enthalten in:

Zur Gesamtaufnahme - volume:91

Enthalten in:

Magnetic resonance in medicine - 91(2024), 6 vom: 25. Apr., Seite 2310-2319

Sprache:

Englisch

Beteiligte Personen:

Huang, Yujia [VerfasserIn]
Chen, Lin [VerfasserIn]
Li, Xu [VerfasserIn]
Liu, Jiaen [VerfasserIn]

Links:

Volltext

Themen:

3D EPI
3D GRE
Journal Article
QSM
Test–retest reliability

Anmerkungen:

Date Completed 08.04.2024

Date Revised 08.04.2024

published: Print-Electronic

UpdateOf: ArXiv. 2023 Aug 15;:. - PMID 37645047

Citation Status MEDLINE

doi:

10.1002/mrm.29992

funding:

Förderinstitution / Projekttitel:

PPN (Katalog-ID):

NLM366475215