A stretchable hardness sensor for the assessment of skin disease in systemic sclerosis
© Author(s) (or their employer(s)) 2023. Re-use permitted under CC BY-NC. No commercial re-use. See rights and permissions. Published by BMJ..
OBJECTIVE: To determine the validity of a hardness sensor to objectively assess skin induration in patients with systemic sclerosis, and to compare the hardness sensor with the modified Rodnan skin score (MRSS) and a durometer.
METHODS: The skin induration was measured in two assessments: a Latin square experiment to examine the hardness sensor's intraobserver and interobserver reliability; and a longitudinal cohort to evaluate the distribution of hardness sensor measurements, the correlation between hardness sensor, durometer and MRSS, and the sensitivity to change in skin hardness. Other outcome data collected included the health assessment questionnaire (HAQ) disability index and Keitel function test (KTF) score.
RESULTS: The reliability of the hardness sensor was excellent, with high intraobserver and interobserver intraclass correlation coefficients (0.97; 0.96), which was higher than MRSS (0.86; 0.74). Interobserver reproducibility of hardness sensor was only poor in abdomen (0.38), yet for durometer it was poor in face (0.11) and abdomen (0.33). The hardness sensor score provided a greater dynamic evaluation range than MRSS. Total hardness sensor score correlated well with MRSS (r=0.90, p<0.001), total durometer score (r=0.95, p<0.001), HAQ disability index (r=0.70, p<0.001) and KTF score (r=0.66, p<0.001). Change in hardness sensor score also correlated with change in MRSS (r=0.78, p<0.001), total durometer score (r=0.85, p<0.001), HAQ disability index (r=0.76, p<0.001) and KTF score (r=0.67, p<0.001).
CONCLUSION: The hardness sensor showed greater reproducibility and accuracy than MRSS, and more application sites than durometer; it can also reflect patients' self-assessments and function test outcomes.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2023 |
---|---|
Erschienen: |
2023 |
Enthalten in: |
Zur Gesamtaufnahme - volume:9 |
---|---|
Enthalten in: |
RMD open - 9(2023), 4 vom: 23. Nov. |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Wang, Xiuyuan [VerfasserIn] |
---|
Links: |
---|
Themen: |
Journal Article |
---|
Anmerkungen: |
Date Completed 27.11.2023 Date Revised 11.12.2023 published: Electronic Citation Status MEDLINE |
---|
doi: |
10.1136/rmdopen-2023-003512 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM364873841 |
---|
LEADER | 01000caa a22002652 4500 | ||
---|---|---|---|
001 | NLM364873841 | ||
003 | DE-627 | ||
005 | 20231227132049.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231226s2023 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1136/rmdopen-2023-003512 |2 doi | |
028 | 5 | 2 | |a pubmed24n1226.xml |
035 | |a (DE-627)NLM364873841 | ||
035 | |a (NLM)37996124 | ||
035 | |a (PII)e003512 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Wang, Xiuyuan |e verfasserin |4 aut | |
245 | 1 | 2 | |a A stretchable hardness sensor for the assessment of skin disease in systemic sclerosis |
264 | 1 | |c 2023 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 27.11.2023 | ||
500 | |a Date Revised 11.12.2023 | ||
500 | |a published: Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a © Author(s) (or their employer(s)) 2023. Re-use permitted under CC BY-NC. No commercial re-use. See rights and permissions. Published by BMJ. | ||
520 | |a OBJECTIVE: To determine the validity of a hardness sensor to objectively assess skin induration in patients with systemic sclerosis, and to compare the hardness sensor with the modified Rodnan skin score (MRSS) and a durometer | ||
520 | |a METHODS: The skin induration was measured in two assessments: a Latin square experiment to examine the hardness sensor's intraobserver and interobserver reliability; and a longitudinal cohort to evaluate the distribution of hardness sensor measurements, the correlation between hardness sensor, durometer and MRSS, and the sensitivity to change in skin hardness. Other outcome data collected included the health assessment questionnaire (HAQ) disability index and Keitel function test (KTF) score | ||
520 | |a RESULTS: The reliability of the hardness sensor was excellent, with high intraobserver and interobserver intraclass correlation coefficients (0.97; 0.96), which was higher than MRSS (0.86; 0.74). Interobserver reproducibility of hardness sensor was only poor in abdomen (0.38), yet for durometer it was poor in face (0.11) and abdomen (0.33). The hardness sensor score provided a greater dynamic evaluation range than MRSS. Total hardness sensor score correlated well with MRSS (r=0.90, p<0.001), total durometer score (r=0.95, p<0.001), HAQ disability index (r=0.70, p<0.001) and KTF score (r=0.66, p<0.001). Change in hardness sensor score also correlated with change in MRSS (r=0.78, p<0.001), total durometer score (r=0.85, p<0.001), HAQ disability index (r=0.76, p<0.001) and KTF score (r=0.67, p<0.001) | ||
520 | |a CONCLUSION: The hardness sensor showed greater reproducibility and accuracy than MRSS, and more application sites than durometer; it can also reflect patients' self-assessments and function test outcomes | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a Outcome and Process Assessment, Health Care | |
650 | 4 | |a Scleroderma, Systemic | |
650 | 4 | |a Systemic Sclerosis | |
700 | 1 | |a Liu, Mengyang |e verfasserin |4 aut | |
700 | 1 | |a Ye, Tianbao |e verfasserin |4 aut | |
700 | 1 | |a Huang, Junxia |e verfasserin |4 aut | |
700 | 1 | |a Xu, Xinzhi |e verfasserin |4 aut | |
700 | 1 | |a Li, Ming |e verfasserin |4 aut | |
700 | 1 | |a Zhao, Xuefeng |e verfasserin |4 aut | |
700 | 1 | |a Lu, Hongliang |e verfasserin |4 aut | |
700 | 1 | |a Yang, Ji |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t RMD open |d 2015 |g 9(2023), 4 vom: 23. Nov. |w (DE-627)NLM254089658 |x 2056-5933 |7 nnns |
773 | 1 | 8 | |g volume:9 |g year:2023 |g number:4 |g day:23 |g month:11 |
856 | 4 | 0 | |u http://dx.doi.org/10.1136/rmdopen-2023-003512 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 9 |j 2023 |e 4 |b 23 |c 11 |