Evaluation of the TraumaGuard Balloon-in-Balloon Catheter Design for Intra-Abdominal Pressure Monitoring : Insights from Pig and Human Cadaver Studies
Introduction: Intra-abdominal pressure (IAP) monitoring is crucial for the detection and prevention of intra-abdominal hypertension (IAH) and abdominal compartment syndrome (ACS). In the 1970s, air-filled catheters (AFCs) for urodynamic studies were introduced as a solution to overcome the limitations of water-perfused catheters. Recent studies have shown that for correct IAP measurement with traditional AFC, the bladder needs to be primed with 25 mL of saline solution to allow pressure wave transmission to the transducer outside of the body, which limits continuous IAP monitoring. Methods: In this study, a novel triple balloon, air-filled TraumaGuard (TG) catheter system from Sentinel Medical Technologies (Jacksonville, FL, USA) with a unique balloon-in-balloon design was evaluated in a porcine and cadaver model of IAH via laparoscopy (IAPgold). Results: In total, 27 and 86 paired IAP measurements were performed in two pigs and one human cadaver, respectively. The mean IAPTG was 20.7 ± 10.7 mmHg compared to IAPgold of 20.3 ± 10.3 mmHg in the porcine study. In the cadaver investigation, the mean IAPTG was 15.6 ± 10.8 mmHg compared to IAPgold of 14.4 ± 10.4 mmHg. The correlation, concordance, bias, precision, limits of agreement, and percentage error were all in accordance with the WSACS (Abdominal Compartment Society) recommendations and guidelines for research. Conclusions: These findings support the use of the TG catheter for continuous IAP monitoring, providing early detection of elevated IAP, thus enabling the potential for prevention of IAH and ACS. Confirmation studies with the TraumaGuard system in critically ill patients are warranted to further validate these findings.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2023 |
---|---|
Erschienen: |
2023 |
Enthalten in: |
Zur Gesamtaufnahme - volume:23 |
---|---|
Enthalten in: |
Sensors (Basel, Switzerland) - 23(2023), 21 vom: 29. Okt. |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Tayebi, Salar [VerfasserIn] |
---|
Links: |
---|
Themen: |
Abdominal compartment syndrome |
---|
Anmerkungen: |
Date Completed 15.11.2023 Date Revised 17.11.2023 published: Electronic Citation Status MEDLINE |
---|
doi: |
10.3390/s23218806 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM364521473 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM364521473 | ||
003 | DE-627 | ||
005 | 20231226095618.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231226s2023 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.3390/s23218806 |2 doi | |
028 | 5 | 2 | |a pubmed24n1215.xml |
035 | |a (DE-627)NLM364521473 | ||
035 | |a (NLM)37960507 | ||
035 | |a (PII)8806 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Tayebi, Salar |e verfasserin |4 aut | |
245 | 1 | 0 | |a Evaluation of the TraumaGuard Balloon-in-Balloon Catheter Design for Intra-Abdominal Pressure Monitoring |b Insights from Pig and Human Cadaver Studies |
264 | 1 | |c 2023 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 15.11.2023 | ||
500 | |a Date Revised 17.11.2023 | ||
500 | |a published: Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a Introduction: Intra-abdominal pressure (IAP) monitoring is crucial for the detection and prevention of intra-abdominal hypertension (IAH) and abdominal compartment syndrome (ACS). In the 1970s, air-filled catheters (AFCs) for urodynamic studies were introduced as a solution to overcome the limitations of water-perfused catheters. Recent studies have shown that for correct IAP measurement with traditional AFC, the bladder needs to be primed with 25 mL of saline solution to allow pressure wave transmission to the transducer outside of the body, which limits continuous IAP monitoring. Methods: In this study, a novel triple balloon, air-filled TraumaGuard (TG) catheter system from Sentinel Medical Technologies (Jacksonville, FL, USA) with a unique balloon-in-balloon design was evaluated in a porcine and cadaver model of IAH via laparoscopy (IAPgold). Results: In total, 27 and 86 paired IAP measurements were performed in two pigs and one human cadaver, respectively. The mean IAPTG was 20.7 ± 10.7 mmHg compared to IAPgold of 20.3 ± 10.3 mmHg in the porcine study. In the cadaver investigation, the mean IAPTG was 15.6 ± 10.8 mmHg compared to IAPgold of 14.4 ± 10.4 mmHg. The correlation, concordance, bias, precision, limits of agreement, and percentage error were all in accordance with the WSACS (Abdominal Compartment Society) recommendations and guidelines for research. Conclusions: These findings support the use of the TG catheter for continuous IAP monitoring, providing early detection of elevated IAP, thus enabling the potential for prevention of IAH and ACS. Confirmation studies with the TraumaGuard system in critically ill patients are warranted to further validate these findings | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a abdominal compartment syndrome | |
650 | 4 | |a air-filled catheter | |
650 | 4 | |a balloon-in-balloon design | |
650 | 4 | |a continuous measurement | |
650 | 4 | |a intra-abdominal hypertension | |
650 | 4 | |a intra-abdominal pressure | |
650 | 4 | |a supine position | |
700 | 1 | |a McKinney, Tim |e verfasserin |4 aut | |
700 | 1 | |a McKinney, Cynthia |e verfasserin |4 aut | |
700 | 1 | |a Delvadia, Dipak |e verfasserin |4 aut | |
700 | 1 | |a Levine, Marc-Alan |e verfasserin |4 aut | |
700 | 1 | |a Spofford, Edward S |c Jr |e verfasserin |4 aut | |
700 | 1 | |a Malbrain, Luca |e verfasserin |4 aut | |
700 | 1 | |a Stiens, Johan |e verfasserin |4 aut | |
700 | 1 | |a Dabrowski, Wojciech |e verfasserin |4 aut | |
700 | 1 | |a Malbrain, Manu L N G |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Sensors (Basel, Switzerland) |d 2007 |g 23(2023), 21 vom: 29. Okt. |w (DE-627)NLM187985170 |x 1424-8220 |7 nnns |
773 | 1 | 8 | |g volume:23 |g year:2023 |g number:21 |g day:29 |g month:10 |
856 | 4 | 0 | |u http://dx.doi.org/10.3390/s23218806 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 23 |j 2023 |e 21 |b 29 |c 10 |