Test-retest repeatability and agreement of the quantitative contrast sensitivity function test : towards the validation of a new clinical endpoint
© 2023. The Author(s), under exclusive licence to Springer-Verlag GmbH Germany, part of Springer Nature..
PURPOSE: The purpose of this study is to investigate test-retest reliability and agreement of the quantitative contrast sensitivity function test (qCSF) in the retina clinic.
METHODS: A total of 121 right eyes of 121 patients were tested and consecutively re-tested with qCSF in the retina clinic. Outcomes included area under the logarithm of contrast sensitivity function curve (AULCSF), contrast acuity, and contrast sensitivity thresholds at 1-18 cycles per degree (cpd). Test-retest means were compared with paired t-test, variability was compared with the Brown-Forsythe test, and intraclass correlation coefficient (ICC) and Bland Altman plots evaluated reliability and agreement.
RESULTS: Mean test-retest differences for all qCSF metrics ranged from 0.02 to 0.05 log units without statistically significant differences in variability. Standard deviations ranged from 0.08 to 0.14. Coefficients of repeatability ranged from 0.16 to 0.27 log units. ICC > 0.9 for all metrics except 1cpd (ICC = 0.84, all p < 0.001); AULCSF ICC = 0.971.
CONCLUSION: qCSF-measured contrast sensitivity shows great test-retest repeatability and agreement in the retina clinic.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2024 |
---|---|
Erschienen: |
2024 |
Enthalten in: |
Zur Gesamtaufnahme - volume:262 |
---|---|
Enthalten in: |
Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie - 262(2024), 3 vom: 01. März, Seite 813-822 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Finn, Matthew [VerfasserIn] |
---|
Links: |
---|
Themen: |
Agreement |
---|
Anmerkungen: |
Date Completed 04.03.2024 Date Revised 04.03.2024 published: Print-Electronic Citation Status MEDLINE |
---|
doi: |
10.1007/s00417-023-06291-y |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM364473568 |
---|
LEADER | 01000caa a22002652 4500 | ||
---|---|---|---|
001 | NLM364473568 | ||
003 | DE-627 | ||
005 | 20240304232033.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231226s2024 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1007/s00417-023-06291-y |2 doi | |
028 | 5 | 2 | |a pubmed24n1316.xml |
035 | |a (DE-627)NLM364473568 | ||
035 | |a (NLM)37955702 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Finn, Matthew |e verfasserin |4 aut | |
245 | 1 | 0 | |a Test-retest repeatability and agreement of the quantitative contrast sensitivity function test |b towards the validation of a new clinical endpoint |
264 | 1 | |c 2024 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 04.03.2024 | ||
500 | |a Date Revised 04.03.2024 | ||
500 | |a published: Print-Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a © 2023. The Author(s), under exclusive licence to Springer-Verlag GmbH Germany, part of Springer Nature. | ||
520 | |a PURPOSE: The purpose of this study is to investigate test-retest reliability and agreement of the quantitative contrast sensitivity function test (qCSF) in the retina clinic | ||
520 | |a METHODS: A total of 121 right eyes of 121 patients were tested and consecutively re-tested with qCSF in the retina clinic. Outcomes included area under the logarithm of contrast sensitivity function curve (AULCSF), contrast acuity, and contrast sensitivity thresholds at 1-18 cycles per degree (cpd). Test-retest means were compared with paired t-test, variability was compared with the Brown-Forsythe test, and intraclass correlation coefficient (ICC) and Bland Altman plots evaluated reliability and agreement | ||
520 | |a RESULTS: Mean test-retest differences for all qCSF metrics ranged from 0.02 to 0.05 log units without statistically significant differences in variability. Standard deviations ranged from 0.08 to 0.14. Coefficients of repeatability ranged from 0.16 to 0.27 log units. ICC > 0.9 for all metrics except 1cpd (ICC = 0.84, all p < 0.001); AULCSF ICC = 0.971 | ||
520 | |a CONCLUSION: qCSF-measured contrast sensitivity shows great test-retest repeatability and agreement in the retina clinic | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a Agreement | |
650 | 4 | |a Clinical endpoints | |
650 | 4 | |a Contrast sensitivity | |
650 | 4 | |a Functional vision | |
650 | 4 | |a Quantitative contrast sensitivity function | |
650 | 4 | |a Repeatability | |
650 | 4 | |a Retina | |
650 | 4 | |a Validation | |
650 | 4 | |a Visual function | |
700 | 1 | |a Vingopoulos, Filippos |e verfasserin |4 aut | |
700 | 1 | |a Zhao, Yan |e verfasserin |4 aut | |
700 | 1 | |a Zhou, Paul |e verfasserin |4 aut | |
700 | 1 | |a Bannerman, Augustine |e verfasserin |4 aut | |
700 | 1 | |a Romano, Francesco |e verfasserin |4 aut | |
700 | 1 | |a Ding, Xinyi |e verfasserin |4 aut | |
700 | 1 | |a Hassan, Zakariyya |e verfasserin |4 aut | |
700 | 1 | |a Patel, Nimesh A |e verfasserin |4 aut | |
700 | 1 | |a Wu, David M |e verfasserin |4 aut | |
700 | 1 | |a Miller, John B |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie |d 1985 |g 262(2024), 3 vom: 01. März, Seite 813-822 |w (DE-627)NLM01296803X |x 1435-702X |7 nnns |
773 | 1 | 8 | |g volume:262 |g year:2024 |g number:3 |g day:01 |g month:03 |g pages:813-822 |
856 | 4 | 0 | |u http://dx.doi.org/10.1007/s00417-023-06291-y |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 262 |j 2024 |e 3 |b 01 |c 03 |h 813-822 |