Applicability of the cobas 6800 System for Epstein-Barr viral load quantitation using whole-blood specimens
© The Author(s) 2023. Published by Oxford University Press on behalf of American Society for Clinical Pathology. All rights reserved. For permissions, please e-mail: journals.permissionsoup.com..
OBJECTIVES: This study aimed to evaluate the analytical performance of the cobas 6800 System (Roche Diagnostics) and assess the feasibility of using whole-blood specimens instead of plasma.
METHODS: The analytical performance of the cobas EBV test (Roche Diagnostics) was evaluated. Thereafter, 120 clinical samples were collected to compare the cobas EBV test and the artus EBV RG PCR Kit (Qiagen). The results of the cobas EBV test conducted using paired plasma as well as 5× and 10× diluted whole-blood specimens were compared with those of the artus EBV RG PCR Kit performed using whole blood.
RESULTS: The precision of the cobas EBV test was acceptable, and its linearity was confirmed to be within the range of 2.85 to 6.89 log IU/mL. Cross-reactivity was not observed. The best qualitative agreement (Cohen κ = 0.733) was observed using 5× diluted whole blood; the best quantitative correlation (Spearman correlation coefficient = 0.6865) was observed using 10× diluted whole blood.
CONCLUSIONS: A significant discrepancy was observed in the results obtained from the 2 assays because of the different specimens used. We observed, however, that diluting whole blood before conducting the cobas EBV test effectively resolved polymerase chain reaction inhibition and viscosity issues, leading to an acceptable correlation with the results from the artus EBV RG PCR Kit conducted using whole blood.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2024 |
---|---|
Erschienen: |
2024 |
Enthalten in: |
Zur Gesamtaufnahme - volume:161 |
---|---|
Enthalten in: |
American journal of clinical pathology - 161(2024), 3 vom: 01. März, Seite 273-282 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Song, Junhyup [VerfasserIn] |
---|
Links: |
---|
Themen: |
Analytical performance |
---|
Anmerkungen: |
Date Completed 04.03.2024 Date Revised 04.03.2024 published: Print Citation Status MEDLINE |
---|
doi: |
10.1093/ajcp/aqad146 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM364280158 |
---|
LEADER | 01000caa a22002652 4500 | ||
---|---|---|---|
001 | NLM364280158 | ||
003 | DE-627 | ||
005 | 20240304232025.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231226s2024 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1093/ajcp/aqad146 |2 doi | |
028 | 5 | 2 | |a pubmed24n1316.xml |
035 | |a (DE-627)NLM364280158 | ||
035 | |a (NLM)37936258 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Song, Junhyup |e verfasserin |4 aut | |
245 | 1 | 0 | |a Applicability of the cobas 6800 System for Epstein-Barr viral load quantitation using whole-blood specimens |
264 | 1 | |c 2024 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 04.03.2024 | ||
500 | |a Date Revised 04.03.2024 | ||
500 | |a published: Print | ||
500 | |a Citation Status MEDLINE | ||
520 | |a © The Author(s) 2023. Published by Oxford University Press on behalf of American Society for Clinical Pathology. All rights reserved. For permissions, please e-mail: journals.permissionsoup.com. | ||
520 | |a OBJECTIVES: This study aimed to evaluate the analytical performance of the cobas 6800 System (Roche Diagnostics) and assess the feasibility of using whole-blood specimens instead of plasma | ||
520 | |a METHODS: The analytical performance of the cobas EBV test (Roche Diagnostics) was evaluated. Thereafter, 120 clinical samples were collected to compare the cobas EBV test and the artus EBV RG PCR Kit (Qiagen). The results of the cobas EBV test conducted using paired plasma as well as 5× and 10× diluted whole-blood specimens were compared with those of the artus EBV RG PCR Kit performed using whole blood | ||
520 | |a RESULTS: The precision of the cobas EBV test was acceptable, and its linearity was confirmed to be within the range of 2.85 to 6.89 log IU/mL. Cross-reactivity was not observed. The best qualitative agreement (Cohen κ = 0.733) was observed using 5× diluted whole blood; the best quantitative correlation (Spearman correlation coefficient = 0.6865) was observed using 10× diluted whole blood | ||
520 | |a CONCLUSIONS: A significant discrepancy was observed in the results obtained from the 2 assays because of the different specimens used. We observed, however, that diluting whole blood before conducting the cobas EBV test effectively resolved polymerase chain reaction inhibition and viscosity issues, leading to an acceptable correlation with the results from the artus EBV RG PCR Kit conducted using whole blood | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a Epstein-Barr virus | |
650 | 4 | |a analytical performance | |
650 | 4 | |a cobas 6800 | |
650 | 4 | |a methods comparison | |
650 | 4 | |a whole blood | |
700 | 1 | |a Kim, Sinyoung |e verfasserin |4 aut | |
700 | 1 | |a Kwak, Eunmin |e verfasserin |4 aut | |
700 | 1 | |a Park, Younhee |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t American journal of clinical pathology |d 1945 |g 161(2024), 3 vom: 01. März, Seite 273-282 |w (DE-627)NLM000037001 |x 1943-7722 |7 nnns |
773 | 1 | 8 | |g volume:161 |g year:2024 |g number:3 |g day:01 |g month:03 |g pages:273-282 |
856 | 4 | 0 | |u http://dx.doi.org/10.1093/ajcp/aqad146 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 161 |j 2024 |e 3 |b 01 |c 03 |h 273-282 |