Mechanical Evaluation Of Unity Elevated Vacuum Suspension System
Copyright (c) 2020 Hossein Gholizadeh, Edward D Lemaire, Rasool Salekrostam..
BACKGROUND: Small residual limb-socket displacement is a good indicator of prosthetic suspension system quality. Active vacuum suspension systems can decrease vertical movement inside the socket, compared to non-active suction systems. This study mechanically evaluated limb-socket displacement with the Össur Unity active vacuum system.
METHOD: Forty-eight conditions were evaluated: four cylindrical and four conical sockets (polypropylene, polyethylene terephthalate glycol-modified (PETG), thermoset resin (acrylic), Thermolyn soft materials); two Iceross Seal-In V liners (standard, high profile); three vacuum conditions (active vacuum, inactive vacuum, no suction with valve open). An Instron 4428 test machine applied 0-100N linear ramped tensile loads to each positive mold, with the socket secured in place, while displacement between the mold and socket was recorded. Following the displacement tests, the load before failure (i.e., 10 mm displacement) was measured.
RESULTS: Average and standard deviations for movement between the mold and sockets were small. The displacement average for all conditions was 0.30±0.16mm for active vacuum, 0.32±0.16mm for inactive vacuum, and 0.39±0.22mm for no suction. Across all trials, active vacuum systems tolerated significantly (p<0.001) more load before failure (812±221N) compared to inactive vacuum (727±213N), and no suction (401±184N). The maximum load before failure (1142±53N) was for the cylindrical polypropylene socket and high-profile liner.
CONCLUSION: The Unity system successfully controlled pistoning inside the socket for regular activity loads and also controlled the greatest traction loads. While relative movement was smallest for Unity, all conditions (inactive vacuum, no suction) were viable for loads less than 100N. Furthermore, similar results can be achieved when using different socket fabrication materials.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2019 |
---|---|
Erschienen: |
2019 |
Enthalten in: |
Zur Gesamtaufnahme - volume:2 |
---|---|
Enthalten in: |
Canadian prosthetics & orthotics journal - 2(2019), 2 vom: 01., Seite 32941 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Gholizadeh, H [VerfasserIn] |
---|
Links: |
---|
Themen: |
Amputation |
---|
Anmerkungen: |
Date Revised 25.08.2023 published: Electronic-eCollection Citation Status PubMed-not-MEDLINE |
---|
doi: |
10.33137/cpoj.v2i2.32941 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM361131739 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM361131739 | ||
003 | DE-627 | ||
005 | 20231226084507.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231226s2019 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.33137/cpoj.v2i2.32941 |2 doi | |
028 | 5 | 2 | |a pubmed24n1203.xml |
035 | |a (DE-627)NLM361131739 | ||
035 | |a (NLM)37614770 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Gholizadeh, H |e verfasserin |4 aut | |
245 | 1 | 0 | |a Mechanical Evaluation Of Unity Elevated Vacuum Suspension System |
264 | 1 | |c 2019 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Revised 25.08.2023 | ||
500 | |a published: Electronic-eCollection | ||
500 | |a Citation Status PubMed-not-MEDLINE | ||
520 | |a Copyright (c) 2020 Hossein Gholizadeh, Edward D Lemaire, Rasool Salekrostam. | ||
520 | |a BACKGROUND: Small residual limb-socket displacement is a good indicator of prosthetic suspension system quality. Active vacuum suspension systems can decrease vertical movement inside the socket, compared to non-active suction systems. This study mechanically evaluated limb-socket displacement with the Össur Unity active vacuum system | ||
520 | |a METHOD: Forty-eight conditions were evaluated: four cylindrical and four conical sockets (polypropylene, polyethylene terephthalate glycol-modified (PETG), thermoset resin (acrylic), Thermolyn soft materials); two Iceross Seal-In V liners (standard, high profile); three vacuum conditions (active vacuum, inactive vacuum, no suction with valve open). An Instron 4428 test machine applied 0-100N linear ramped tensile loads to each positive mold, with the socket secured in place, while displacement between the mold and socket was recorded. Following the displacement tests, the load before failure (i.e., 10 mm displacement) was measured | ||
520 | |a RESULTS: Average and standard deviations for movement between the mold and sockets were small. The displacement average for all conditions was 0.30±0.16mm for active vacuum, 0.32±0.16mm for inactive vacuum, and 0.39±0.22mm for no suction. Across all trials, active vacuum systems tolerated significantly (p<0.001) more load before failure (812±221N) compared to inactive vacuum (727±213N), and no suction (401±184N). The maximum load before failure (1142±53N) was for the cylindrical polypropylene socket and high-profile liner | ||
520 | |a CONCLUSION: The Unity system successfully controlled pistoning inside the socket for regular activity loads and also controlled the greatest traction loads. While relative movement was smallest for Unity, all conditions (inactive vacuum, no suction) were viable for loads less than 100N. Furthermore, similar results can be achieved when using different socket fabrication materials | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a Amputation | |
650 | 4 | |a Elevated vacuum | |
650 | 4 | |a Prosthesis | |
650 | 4 | |a Prosthetic limb | |
650 | 4 | |a Prosthetic suspension system | |
650 | 4 | |a Rehabilitation | |
650 | 4 | |a sociodemographics | |
700 | 1 | |a Lemaire, E D |e verfasserin |4 aut | |
700 | 1 | |a Salekrostam, R |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Canadian prosthetics & orthotics journal |d 2019 |g 2(2019), 2 vom: 01., Seite 32941 |w (DE-627)NLM361128053 |x 2561-987X |7 nnns |
773 | 1 | 8 | |g volume:2 |g year:2019 |g number:2 |g day:01 |g pages:32941 |
856 | 4 | 0 | |u http://dx.doi.org/10.33137/cpoj.v2i2.32941 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 2 |j 2019 |e 2 |b 01 |h 32941 |