Practical approaches to the detection of macrotroponin
INTRODUCTION: Macrotroponin is increasingly recognised as a cause of confusion in interpreting high-sensitivity cardiac troponin (hs-cTnI) results. In this study, we sought to evaluate two practical approaches to detecting macrotroponin. These two approaches are PEG precipitation and SVM (support vector machine) analysis to classify discrepancies between hs-cTn assays.
METHOD: Residual serum and heparin plasma specimens (n = 483) with initially elevated hs-cTnI from hospital and community laboratories were retested on multiple hs-cTn platforms before and after PEG precipitation and Protein A immunoglobulin depletion. SVM analysis was conducted to identify a linear equation that best discriminated specimens with macrotroponin using a combination of results from two different hs-cTn assays.
FINDINGS: The diagnostic performance of PEG precipitation was carried out using Protein A immunoglobulin depletion as the reference comparator. When a cutoff residual activity after PEG precipitation of ≤ 20% was used, this threshold carried a high specificity of 92% (confidence interval 83-98%; n = 189) using the Siemens hs-cTnI Vista assay and 95% specificity (86%-98%; n = 242) using the Abbott hs-cTnI Architect assay. SVM analysis generated a linear equation identifying macrotroponin specimens from results obtained on two hs-cTn assays. This approach can be highly specific, comparable to PEG precipitation when certain assay combinations and concentrations are used.
CONCLUSION: We describe and identify practical alternatives to detecting macrotroponin. These approaches can be optimised for high specificity, reducing the need for more complex laboratory methods.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2024 |
---|---|
Erschienen: |
2024 |
Enthalten in: |
Zur Gesamtaufnahme - volume:61 |
---|---|
Enthalten in: |
Annals of clinical biochemistry - 61(2024), 2 vom: 22. März, Seite 122-132 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Lam, Leo [VerfasserIn] |
---|
Links: |
---|
Themen: |
Antibody interference |
---|
Anmerkungen: |
Date Completed 15.03.2024 Date Revised 15.03.2024 published: Print-Electronic Citation Status MEDLINE |
---|
doi: |
10.1177/00045632231197301 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM360775543 |
---|
LEADER | 01000caa a22002652 4500 | ||
---|---|---|---|
001 | NLM360775543 | ||
003 | DE-627 | ||
005 | 20240315232716.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231226s2024 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1177/00045632231197301 |2 doi | |
028 | 5 | 2 | |a pubmed24n1330.xml |
035 | |a (DE-627)NLM360775543 | ||
035 | |a (NLM)37578158 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Lam, Leo |e verfasserin |4 aut | |
245 | 1 | 0 | |a Practical approaches to the detection of macrotroponin |
264 | 1 | |c 2024 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 15.03.2024 | ||
500 | |a Date Revised 15.03.2024 | ||
500 | |a published: Print-Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a INTRODUCTION: Macrotroponin is increasingly recognised as a cause of confusion in interpreting high-sensitivity cardiac troponin (hs-cTnI) results. In this study, we sought to evaluate two practical approaches to detecting macrotroponin. These two approaches are PEG precipitation and SVM (support vector machine) analysis to classify discrepancies between hs-cTn assays | ||
520 | |a METHOD: Residual serum and heparin plasma specimens (n = 483) with initially elevated hs-cTnI from hospital and community laboratories were retested on multiple hs-cTn platforms before and after PEG precipitation and Protein A immunoglobulin depletion. SVM analysis was conducted to identify a linear equation that best discriminated specimens with macrotroponin using a combination of results from two different hs-cTn assays | ||
520 | |a FINDINGS: The diagnostic performance of PEG precipitation was carried out using Protein A immunoglobulin depletion as the reference comparator. When a cutoff residual activity after PEG precipitation of ≤ 20% was used, this threshold carried a high specificity of 92% (confidence interval 83-98%; n = 189) using the Siemens hs-cTnI Vista assay and 95% specificity (86%-98%; n = 242) using the Abbott hs-cTnI Architect assay. SVM analysis generated a linear equation identifying macrotroponin specimens from results obtained on two hs-cTn assays. This approach can be highly specific, comparable to PEG precipitation when certain assay combinations and concentrations are used | ||
520 | |a CONCLUSION: We describe and identify practical alternatives to detecting macrotroponin. These approaches can be optimised for high specificity, reducing the need for more complex laboratory methods | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a Troponin | |
650 | 4 | |a antibody interference | |
650 | 4 | |a macrotroponin | |
650 | 7 | |a Troponin I |2 NLM | |
650 | 7 | |a Immunoglobulins |2 NLM | |
650 | 7 | |a Troponin T |2 NLM | |
650 | 7 | |a Biomarkers |2 NLM | |
700 | 1 | |a Kyle, Campbell |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Annals of clinical biochemistry |d 1974 |g 61(2024), 2 vom: 22. März, Seite 122-132 |w (DE-627)NLM000057002 |x 1758-1001 |7 nnns |
773 | 1 | 8 | |g volume:61 |g year:2024 |g number:2 |g day:22 |g month:03 |g pages:122-132 |
856 | 4 | 0 | |u http://dx.doi.org/10.1177/00045632231197301 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 61 |j 2024 |e 2 |b 22 |c 03 |h 122-132 |