Artefactually low creatinine by Beckman Coulter enzymatic method due to immunoglobulin M paraprotein interference
An 81-year-old man was admitted to hospital with symptomatic coronavirus disease (COVID-19) infection. He had a background of progressive chronic inflammatory demyelinating polyneuropathy associated with Waldenstrom's macroglobulinaemia. His plasma creatinine on four separate samples was inconceivably low (all ≤13 μmol/L), as measured by a Beckman Coulter enzymatic assay) after being 72 μmol/L 3 months earlier. On further investigation, his serum immunoglobulin M (IgM) was 15.4 g/L and his plasma creatinine measured by Roche enzymatic and Roche Jaffe methods was 62 μmol/L and 64 μmol/L, respectively. This was consistent with results post dilution studies and polyethylene glycol (PEG) precipitation on the Beckman Coulter assay. There was no evidence of similar interference when reviewing creatinine results from 10 other patients with IgM paraproteinaemia who had been tested in our laboratory. Clinicians and laboratorians are reminded that enzymatic creatinine is not free from interferences. IgM paraprotein negative interference of enzymatic creatinine is rare and specific to a patient's IgM and assay combination, but should be considered in patients with an unexplained low enzymatic creatinine result. Useful investigations to identify an interference include dilution studies, PEG precipitation and measuring creatinine on an alternative method such as Jaffe, mass spectrometry or an enzymatic method from a different platform.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2023 |
---|---|
Erschienen: |
2023 |
Enthalten in: |
Zur Gesamtaufnahme - volume:60 |
---|---|
Enthalten in: |
Annals of clinical biochemistry - 60(2023), 6 vom: 21. Nov., Seite 423-427 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Li, Bobby [VerfasserIn] |
---|
Links: |
---|
Themen: |
AYI8EX34EU |
---|
Anmerkungen: |
Date Completed 15.11.2023 Date Revised 24.11.2023 published: Print-Electronic Citation Status MEDLINE |
---|
doi: |
10.1177/00045632231190507 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM359828663 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM359828663 | ||
003 | DE-627 | ||
005 | 20231226081718.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231226s2023 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1177/00045632231190507 |2 doi | |
028 | 5 | 2 | |a pubmed24n1199.xml |
035 | |a (DE-627)NLM359828663 | ||
035 | |a (NLM)37482625 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Li, Bobby |e verfasserin |4 aut | |
245 | 1 | 0 | |a Artefactually low creatinine by Beckman Coulter enzymatic method due to immunoglobulin M paraprotein interference |
264 | 1 | |c 2023 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 15.11.2023 | ||
500 | |a Date Revised 24.11.2023 | ||
500 | |a published: Print-Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a An 81-year-old man was admitted to hospital with symptomatic coronavirus disease (COVID-19) infection. He had a background of progressive chronic inflammatory demyelinating polyneuropathy associated with Waldenstrom's macroglobulinaemia. His plasma creatinine on four separate samples was inconceivably low (all ≤13 μmol/L), as measured by a Beckman Coulter enzymatic assay) after being 72 μmol/L 3 months earlier. On further investigation, his serum immunoglobulin M (IgM) was 15.4 g/L and his plasma creatinine measured by Roche enzymatic and Roche Jaffe methods was 62 μmol/L and 64 μmol/L, respectively. This was consistent with results post dilution studies and polyethylene glycol (PEG) precipitation on the Beckman Coulter assay. There was no evidence of similar interference when reviewing creatinine results from 10 other patients with IgM paraproteinaemia who had been tested in our laboratory. Clinicians and laboratorians are reminded that enzymatic creatinine is not free from interferences. IgM paraprotein negative interference of enzymatic creatinine is rare and specific to a patient's IgM and assay combination, but should be considered in patients with an unexplained low enzymatic creatinine result. Useful investigations to identify an interference include dilution studies, PEG precipitation and measuring creatinine on an alternative method such as Jaffe, mass spectrometry or an enzymatic method from a different platform | ||
650 | 4 | |a Case Reports | |
650 | 4 | |a Journal Article | |
650 | 4 | |a Beckman Coulter | |
650 | 4 | |a IgM | |
650 | 4 | |a enzymatic creatinine | |
650 | 4 | |a immunoglobulin M | |
650 | 4 | |a paraprotein | |
650 | 7 | |a Paraproteins |2 NLM | |
650 | 7 | |a Immunoglobulin M |2 NLM | |
650 | 7 | |a Creatinine |2 NLM | |
650 | 7 | |a AYI8EX34EU |2 NLM | |
700 | 1 | |a Thompson, Simon |e verfasserin |4 aut | |
700 | 1 | |a Igot, Kareen |e verfasserin |4 aut | |
700 | 1 | |a King, Richard |e verfasserin |4 aut | |
700 | 1 | |a Doogue, Matthew |e verfasserin |4 aut | |
700 | 1 | |a Florkowski, Chris |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Annals of clinical biochemistry |d 1974 |g 60(2023), 6 vom: 21. Nov., Seite 423-427 |w (DE-627)NLM000057002 |x 1758-1001 |7 nnns |
773 | 1 | 8 | |g volume:60 |g year:2023 |g number:6 |g day:21 |g month:11 |g pages:423-427 |
856 | 4 | 0 | |u http://dx.doi.org/10.1177/00045632231190507 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 60 |j 2023 |e 6 |b 21 |c 11 |h 423-427 |