Validation of a Monte Carlo simulated cardiac phantom for planar and SPECT studies
Copyright © 2023 Associazione Italiana di Fisica Medica e Sanitaria. Published by Elsevier Ltd. All rights reserved..
PURPOSE: This work aimed to validate Monte Carlo (MC) simulated cardiac phantoms for the evaluation of planar- and SPECT-gated-blood-pool (GBP-P and GBP-S) studies.
METHODS: A comparison of gamma camera system performance criteria measurements (energy resolution, spatial resolution, sensitivity) with MC simulations was conducted. Furthermore, the accuracy of measured and simulated volumes of two stereolithography-printed cardiac phantoms (based on 4D-XCAT phantoms) was assessed. Finally, the simulated GBP-P and GBP-S XCAT studies were validated by comparing calculated left ventricular ejection fraction (LVEF) and ventricle volume values with known parameters.
RESULTS: The simulated performance criteria compared well with measured values (energy resolution difference: 0.1 ± 0.10%; spatial resolution (full width at half maximum) difference ≤ 0.5 ± 0.8 mm and system sensitivity difference ≤ 6.2 ± 0.62cps/MBq). The measured and simulated cardiac phantoms were in good agreement; the left anterior oblique views compared well. This is supported by line profiles through these phantoms and on average, simulated counts were 5.8% lower than measured counts. The LVEF values calculated from the GBP-P and GBP-S simulated data differ from known values (2.8 ± 0.64% and 0.8 ± 0.52%). The differences between the known XCAT LV volumes and simulated GBP-S calculated volumes were -1.2 ± 1.91 ml and -1.5 ± 0.96 ml for the end-diastolic and end-systolic volumes.
CONCLUSION: The MC-simulated cardiac phantom has been validated successfully. Stereolithography-printing allows researchers to create clinically realistic organ phantoms and is a valuable tool for validating MC simulations and clinical software. By conducting GBP simulation studies with various XCAT models, the user will be able to generate GBP-P and GBP-S databases for future software evaluation.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2023 |
---|---|
Erschienen: |
2023 |
Enthalten in: |
Zur Gesamtaufnahme - volume:111 |
---|---|
Enthalten in: |
Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) - 111(2023) vom: 20. Juli, Seite 102617 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Pieters, Hané [VerfasserIn] |
---|
Links: |
---|
Themen: |
4D-XCAT |
---|
Anmerkungen: |
Date Completed 23.06.2023 Date Revised 23.06.2023 published: Print-Electronic Citation Status MEDLINE |
---|
doi: |
10.1016/j.ejmp.2023.102617 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM357920279 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM357920279 | ||
003 | DE-627 | ||
005 | 20231226073633.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231226s2023 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1016/j.ejmp.2023.102617 |2 doi | |
028 | 5 | 2 | |a pubmed24n1193.xml |
035 | |a (DE-627)NLM357920279 | ||
035 | |a (NLM)37290226 | ||
035 | |a (PII)S1120-1797(23)00094-7 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Pieters, Hané |e verfasserin |4 aut | |
245 | 1 | 0 | |a Validation of a Monte Carlo simulated cardiac phantom for planar and SPECT studies |
264 | 1 | |c 2023 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 23.06.2023 | ||
500 | |a Date Revised 23.06.2023 | ||
500 | |a published: Print-Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a Copyright © 2023 Associazione Italiana di Fisica Medica e Sanitaria. Published by Elsevier Ltd. All rights reserved. | ||
520 | |a PURPOSE: This work aimed to validate Monte Carlo (MC) simulated cardiac phantoms for the evaluation of planar- and SPECT-gated-blood-pool (GBP-P and GBP-S) studies | ||
520 | |a METHODS: A comparison of gamma camera system performance criteria measurements (energy resolution, spatial resolution, sensitivity) with MC simulations was conducted. Furthermore, the accuracy of measured and simulated volumes of two stereolithography-printed cardiac phantoms (based on 4D-XCAT phantoms) was assessed. Finally, the simulated GBP-P and GBP-S XCAT studies were validated by comparing calculated left ventricular ejection fraction (LVEF) and ventricle volume values with known parameters | ||
520 | |a RESULTS: The simulated performance criteria compared well with measured values (energy resolution difference: 0.1 ± 0.10%; spatial resolution (full width at half maximum) difference ≤ 0.5 ± 0.8 mm and system sensitivity difference ≤ 6.2 ± 0.62cps/MBq). The measured and simulated cardiac phantoms were in good agreement; the left anterior oblique views compared well. This is supported by line profiles through these phantoms and on average, simulated counts were 5.8% lower than measured counts. The LVEF values calculated from the GBP-P and GBP-S simulated data differ from known values (2.8 ± 0.64% and 0.8 ± 0.52%). The differences between the known XCAT LV volumes and simulated GBP-S calculated volumes were -1.2 ± 1.91 ml and -1.5 ± 0.96 ml for the end-diastolic and end-systolic volumes | ||
520 | |a CONCLUSION: The MC-simulated cardiac phantom has been validated successfully. Stereolithography-printing allows researchers to create clinically realistic organ phantoms and is a valuable tool for validating MC simulations and clinical software. By conducting GBP simulation studies with various XCAT models, the user will be able to generate GBP-P and GBP-S databases for future software evaluation | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a 4D-XCAT | |
650 | 4 | |a Gated blood pool studies | |
650 | 4 | |a SIMIND | |
650 | 4 | |a Stereolithography 3D printed phantom | |
700 | 1 | |a van Staden, Johannes A |e verfasserin |4 aut | |
700 | 1 | |a du Plessis, Frederik C P |e verfasserin |4 aut | |
700 | 1 | |a du Raan, Hanlie |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) |d 1997 |g 111(2023) vom: 20. Juli, Seite 102617 |w (DE-627)NLM092554024 |x 1724-191X |7 nnns |
773 | 1 | 8 | |g volume:111 |g year:2023 |g day:20 |g month:07 |g pages:102617 |
856 | 4 | 0 | |u http://dx.doi.org/10.1016/j.ejmp.2023.102617 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 111 |j 2023 |b 20 |c 07 |h 102617 |