1.5-T MR relaxometry in quantifying splenic and pancreatic iron : retrospective comparison of a commercial 3D-Dixon sequence and an established 2D multi-gradient echo sequence

© 2023. The Author(s)..

OBJECTIVES: To compare the quantitative measurement of splenic and pancreatic iron content using a commercial 3D-Dixon sequence (qDixon) versus an established fat-saturated R2* relaxometry method (ME-GRE).

METHODS: We analyzed splenic and pancreatic iron levels in 143 MR examinations (1.5 T) using the qDixon and a ME-GRE sequence (108 patients: 65 males, 43 females, mean age 61.31 years). Splenic and pancreatic R2* values were compared between both methods using Bland-Altman plots, concordance correlation coefficients (CCC), and linear regression analyses. Iron overload (R2* > 50 1/s) was defined for both organs and compared using contingency tables, overall agreement, and Gwet's AC1 coefficient.

RESULTS: Of all analyzable examinations, the median splenic R2* using the qDixon sequence was 25.75 1/s (range: 5.6-433) and for the ME-GRE sequence 35.35 1/s (range: 10.9-400.8) respectively. Concerning the pancreas, a median R2* of 29.93 1/s (range: 14-111.45) for the qDixon and 31.25 1/s (range: 14-97) for the ME-GRE sequence was found. Bland-Altman analysis showed a mean R2* difference of 2.12 1/s with a CCC of 0.934 for the spleen and of 0.29 1/s with a CCC of 0.714 for the pancreas. Linear regression for the spleen/pancreas resulted in a correlation coefficient of 0.94 (p < 0.001)/0.725 (p < 0.001). Concerning iron overload, the proportion of overall agreement between the two methods was 91.43% for the spleen and 93.18% for the pancreas.

CONCLUSIONS: Our data show good concordance between R2* values obtained with a commercial qDixon sequence and a validated ME-GRE relaxometry method. The 3D-qDixon sequence, originally intended for liver assessment, seems to be a reliable tool for non-invasive evaluation of iron content also in the spleen and the pancreas.

KEY POINTS: • A 3D chemical shift imaging sequence and 2D multi-gradient echo sequence show good conformity quantifying splenic and pancreatic R2* values. • The 3D chemical shift imaging sequence allows a reliable analysis also of splenic and pancreatic iron status. • In addition to the liver, the analysis of the spleen and pancreas is often helpful for further differential diagnostic clarification and patient guidance regarding the iron status.

Medienart:

E-Artikel

Erscheinungsjahr:

2023

Erschienen:

2023

Enthalten in:

Zur Gesamtaufnahme - volume:33

Enthalten in:

European radiology - 33(2023), 7 vom: 17. Juli, Seite 4973-4980

Sprache:

Englisch

Beteiligte Personen:

Plaikner, Michaela [VerfasserIn]
Lanser, Lukas [VerfasserIn]
Kremser, Christian [VerfasserIn]
Weiss, Günter [VerfasserIn]
Henninger, Benjamin [VerfasserIn]

Links:

Volltext

Themen:

E1UOL152H7
Iron
Iron overload
Journal Article
Liver
Pancreas
Spleen

Anmerkungen:

Date Completed 26.06.2023

Date Revised 10.11.2023

published: Print-Electronic

Citation Status MEDLINE

doi:

10.1007/s00330-023-09451-0

funding:

Förderinstitution / Projekttitel:

PPN (Katalog-ID):

NLM353064777