The Role of Ultrasound in Pediatric Skull Fractures : Determination of Fracture and Optic Nerve Sheath Diameter Measurements
Copyright © 2023 Wolters Kluwer Health, Inc. All rights reserved..
BACKGROUND: The aim of the present study was to determine the accuracy of point-of-care ultrasound (POCUS) for detecting skull fractures and to evaluate sonographic measurements of optic nerve sheath diameter (ONSD) and ONSD/eyeball vertical diameter (EVD) ratios in children with head trauma.
METHODS: Children who presented with local signs of head trauma and underwent cranial computed tomography (CT) were enrolled. The suspected area was examined by POCUS to identify a skull fracture, and then the ONSD at 3 mm posterior to the globe and the EVD were measured. Ratios of ONSD measurement at 3 mm/EVD were reported. All ONSD measurements and ratios were calculated from cranial CT images.
RESULTS: There were 112 children enrolled in the study. The sensitivity and specificity of POCUS for skull fractures was 93.7% (95% confidence interval [CI], 82.8-98.6) and 96.8% (95% CI, 89.1-99.6), whereas the positive predictive value was 95.7% (95% CI, 85.1-98.8), and the negative predictive value was 95.3% (95% CI, 87.3-98.4). There was high agreement between POCUS and CT for identifying skull fractures (κ, 0.90 [±0.04]). In the group without elevated intracranial pressure findings on CT, patients with space-occupying lesions (SOLs) had higher sonographic ONSD measurements and ratios (P < 0.001) compared with cases without SOLs.
CONCLUSIONS: When used with clinical decision rules to minimize the risk for clinically important traumatic brain injury, POCUS seems to be a promising tool to detect skull fractures and calculate ONSD measurements and rates to predict the risk for SOLs and perform further risk stratification of children with minor head trauma.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2023 |
---|---|
Erschienen: |
2023 |
Enthalten in: |
Zur Gesamtaufnahme - volume:39 |
---|---|
Enthalten in: |
Pediatric emergency care - 39(2023), 2 vom: 01. Feb., Seite 91-97 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Şık, Nihan [VerfasserIn] |
---|
Links: |
---|
Themen: |
---|
Anmerkungen: |
Date Completed 06.02.2024 Date Revised 06.02.2024 published: Print-Electronic Citation Status MEDLINE |
---|
doi: |
10.1097/PEC.0000000000002895 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM352297697 |
---|
LEADER | 01000caa a22002652 4500 | ||
---|---|---|---|
001 | NLM352297697 | ||
003 | DE-627 | ||
005 | 20240206231838.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231226s2023 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1097/PEC.0000000000002895 |2 doi | |
028 | 5 | 2 | |a pubmed24n1282.xml |
035 | |a (DE-627)NLM352297697 | ||
035 | |a (NLM)36719390 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Şık, Nihan |e verfasserin |4 aut | |
245 | 1 | 4 | |a The Role of Ultrasound in Pediatric Skull Fractures |b Determination of Fracture and Optic Nerve Sheath Diameter Measurements |
264 | 1 | |c 2023 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 06.02.2024 | ||
500 | |a Date Revised 06.02.2024 | ||
500 | |a published: Print-Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a Copyright © 2023 Wolters Kluwer Health, Inc. All rights reserved. | ||
520 | |a BACKGROUND: The aim of the present study was to determine the accuracy of point-of-care ultrasound (POCUS) for detecting skull fractures and to evaluate sonographic measurements of optic nerve sheath diameter (ONSD) and ONSD/eyeball vertical diameter (EVD) ratios in children with head trauma | ||
520 | |a METHODS: Children who presented with local signs of head trauma and underwent cranial computed tomography (CT) were enrolled. The suspected area was examined by POCUS to identify a skull fracture, and then the ONSD at 3 mm posterior to the globe and the EVD were measured. Ratios of ONSD measurement at 3 mm/EVD were reported. All ONSD measurements and ratios were calculated from cranial CT images | ||
520 | |a RESULTS: There were 112 children enrolled in the study. The sensitivity and specificity of POCUS for skull fractures was 93.7% (95% confidence interval [CI], 82.8-98.6) and 96.8% (95% CI, 89.1-99.6), whereas the positive predictive value was 95.7% (95% CI, 85.1-98.8), and the negative predictive value was 95.3% (95% CI, 87.3-98.4). There was high agreement between POCUS and CT for identifying skull fractures (κ, 0.90 [±0.04]). In the group without elevated intracranial pressure findings on CT, patients with space-occupying lesions (SOLs) had higher sonographic ONSD measurements and ratios (P < 0.001) compared with cases without SOLs | ||
520 | |a CONCLUSIONS: When used with clinical decision rules to minimize the risk for clinically important traumatic brain injury, POCUS seems to be a promising tool to detect skull fractures and calculate ONSD measurements and rates to predict the risk for SOLs and perform further risk stratification of children with minor head trauma | ||
650 | 4 | |a Journal Article | |
700 | 1 | |a Öztürk, Ali |e verfasserin |4 aut | |
700 | 1 | |a Yılmaz, Durgül |e verfasserin |4 aut | |
700 | 1 | |a Duman, Murat |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Pediatric emergency care |d 1993 |g 39(2023), 2 vom: 01. Feb., Seite 91-97 |w (DE-627)NLM013326031 |x 1535-1815 |7 nnns |
773 | 1 | 8 | |g volume:39 |g year:2023 |g number:2 |g day:01 |g month:02 |g pages:91-97 |
856 | 4 | 0 | |u http://dx.doi.org/10.1097/PEC.0000000000002895 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 39 |j 2023 |e 2 |b 01 |c 02 |h 91-97 |