Comparative study on the accuracy of implant-borne single restoration by two production processes with Ti-base

PURPOSE: To study on the accuracy of implant-borne single restoration by two production processes with Ti-base to provide experimental data for proper processes of single implant-borne restoration.

METHODS: Thirty patients were selected with single posterior teeth missing from the Department of Oral Implantology of Shanghai Putuo District Eye Disease and Dental Disease Prevention and Treatment Institute. The patients were taken 2 traditional impressions clinically for two plaster model equipped with implant analogue. These models were then divided into 2 groups according to different production processes. The experimental group was scanned with the scan body installed in the model implant analogue, while the control group was scanned directly on the Ti-base abutment installed in the model implant analogue. The implant-borne single restorations of the two groups were cut along the buccal-lingual side and the distance between the measuring point to the Ti-Base abutment was observed by electron microscopy. In addition, the breaking limit of zirconia crown was observed, universal test machine was used to load direct force to the crown. SPSS 22.0 software package was used for data analysis.

RESULTS: The gap between the implant-borne single restoration to the Ti-base abutment of the experimental group was significantly smaller than that of the control group. The difference between the two groups was statistically significant (P<0.05). However, by testing the breaking limit of zirconia crown, there was no significant difference(P>0.05).

CONCLUSIONS: Using scan body to transfer the implant position and Ti-base abutment data information to the digital dental design software is more accurate and reliable than directly scanning the Ti-base on the model analogue. Using scan body is recommended for processing and manufacture of implant-borne singe restoration.

Medienart:

Artikel

Erscheinungsjahr:

2022

Erschienen:

2022

Enthalten in:

Zur Gesamtaufnahme - volume:31

Enthalten in:

Shanghai kou qiang yi xue = Shanghai journal of stomatology - 31(2022), 4 vom: 27. Aug., Seite 414-417

Sprache:

Chinesisch

Beteiligte Personen:

Cao, Wei-Yu [VerfasserIn]
Cao, Jie [VerfasserIn]
Liu, Bin-Yi [VerfasserIn]
Xu, Zhou [VerfasserIn]
Lu, Wei-Qing [VerfasserIn]

Themen:

C6V6S92N3C
D1JT611TNE
English Abstract
Journal Article
S38N85C5G0
Titanium
Zirconium
Zirconium oxide

Anmerkungen:

Date Completed 31.01.2023

Date Revised 02.02.2023

published: Print

Citation Status MEDLINE

Förderinstitution / Projekttitel:

PPN (Katalog-ID):

NLM352209712