Reduced fibrous capsule elastic fibers from biologic ECM-enveloped CIEDs in minipigs, supported with a novel compression mechanics model
BACKGROUND: Fibrous capsules (Fb) in response to cardiovascular implantable electronic devices (CIEDs), including a pacemaker (P) system, can produce patient discomfort and difficulties in revision surgery due partially to their increased compressive strength, previously linked to elevated tissue fibers.
OBJECTIVE: A preliminary study to quantify structural proteins, determine if biologic extracellular matrix-enveloped CIEDs (PECM) caused differential Fb properties, and to implement a realistic mechanical model.
METHODS: Retrieved Fb (-P and -PECM) from minipigs were subjected to biomechanical (shear oscillation and uniaxial compression) and histological (collagen I and elastin) analyses.
RESULTS: Fb-PECM showed significant decreases compared to Fb-P in: low strain-loss modulus (390 vs. 541 Pa) across angular frequencies, high strain-compressive elastic modulus (1043 vs. 2042 kPa), and elastic fiber content (1.92 vs. 3.15 μg/mg tissue). Decreases in elastin were particularly noted closer to the implant's surface (Fb-PECM = 71% vs. Fb-P = 143% relative to dermal elastin at mid-tangential sections) and verified with a solid mechanics hyperelasticity with direction-dependent fiber viscoelasticity compression simulation (r2 ≥ 98.9%).
CONCLUSIONS: The biologic envelope composed of decellularized porcine small intestine submucosa ECM for CIEDs promoted fibrous tissues with less elastic fibers. Novel compression modeling analyses directly correlated this singular reduction to more desirable subcutaneous tissue mechanics.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2023 |
---|---|
Erschienen: |
2023 |
Enthalten in: |
Zur Gesamtaufnahme - volume:34 |
---|---|
Enthalten in: |
Bio-medical materials and engineering - 34(2023), 4 vom: 04., Seite 289-304 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
de Guzman, Roche C [VerfasserIn] |
---|
Links: |
---|
Anmerkungen: |
Date Completed 11.07.2023 Date Revised 22.07.2023 published: Print Citation Status MEDLINE |
---|
doi: |
10.3233/BME-221488 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM351288643 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM351288643 | ||
003 | DE-627 | ||
005 | 20231226051011.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231226s2023 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.3233/BME-221488 |2 doi | |
028 | 5 | 2 | |a pubmed24n1170.xml |
035 | |a (DE-627)NLM351288643 | ||
035 | |a (NLM)36617774 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a de Guzman, Roche C |e verfasserin |4 aut | |
245 | 1 | 0 | |a Reduced fibrous capsule elastic fibers from biologic ECM-enveloped CIEDs in minipigs, supported with a novel compression mechanics model |
264 | 1 | |c 2023 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 11.07.2023 | ||
500 | |a Date Revised 22.07.2023 | ||
500 | |a published: Print | ||
500 | |a Citation Status MEDLINE | ||
520 | |a BACKGROUND: Fibrous capsules (Fb) in response to cardiovascular implantable electronic devices (CIEDs), including a pacemaker (P) system, can produce patient discomfort and difficulties in revision surgery due partially to their increased compressive strength, previously linked to elevated tissue fibers | ||
520 | |a OBJECTIVE: A preliminary study to quantify structural proteins, determine if biologic extracellular matrix-enveloped CIEDs (PECM) caused differential Fb properties, and to implement a realistic mechanical model | ||
520 | |a METHODS: Retrieved Fb (-P and -PECM) from minipigs were subjected to biomechanical (shear oscillation and uniaxial compression) and histological (collagen I and elastin) analyses | ||
520 | |a RESULTS: Fb-PECM showed significant decreases compared to Fb-P in: low strain-loss modulus (390 vs. 541 Pa) across angular frequencies, high strain-compressive elastic modulus (1043 vs. 2042 kPa), and elastic fiber content (1.92 vs. 3.15 μg/mg tissue). Decreases in elastin were particularly noted closer to the implant's surface (Fb-PECM = 71% vs. Fb-P = 143% relative to dermal elastin at mid-tangential sections) and verified with a solid mechanics hyperelasticity with direction-dependent fiber viscoelasticity compression simulation (r2 ≥ 98.9%) | ||
520 | |a CONCLUSIONS: The biologic envelope composed of decellularized porcine small intestine submucosa ECM for CIEDs promoted fibrous tissues with less elastic fibers. Novel compression modeling analyses directly correlated this singular reduction to more desirable subcutaneous tissue mechanics | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a 3D distribution | |
650 | 4 | |a COMSOL multiphysics hyperelastic simulation | |
650 | 4 | |a Minipig study | |
650 | 4 | |a biologic envelope | |
650 | 4 | |a cardiovascular implantable electronic devices | |
650 | 4 | |a compression biomechanics | |
650 | 4 | |a decellularized tissue ECM | |
650 | 4 | |a elastin and elastic fibers | |
650 | 4 | |a fibrous tissue encapsulation | |
650 | 4 | |a foreign body reaction | |
650 | 4 | |a oscillation rheology | |
650 | 4 | |a pacemaker | |
650 | 7 | |a Elastin |2 NLM | |
650 | 7 | |a 9007-58-3 |2 NLM | |
650 | 7 | |a Biological Products |2 NLM | |
700 | 1 | |a Meer, Allison S |e verfasserin |4 aut | |
700 | 1 | |a Mathews, Aidan A |e verfasserin |4 aut | |
700 | 1 | |a Israel, Atara R |e verfasserin |4 aut | |
700 | 1 | |a Moses, Michael T |e verfasserin |4 aut | |
700 | 1 | |a Sams, Clarence M |e verfasserin |4 aut | |
700 | 1 | |a Deegan, Daniel B |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Bio-medical materials and engineering |d 1996 |g 34(2023), 4 vom: 04., Seite 289-304 |w (DE-627)NLM013151207 |x 1878-3619 |7 nnns |
773 | 1 | 8 | |g volume:34 |g year:2023 |g number:4 |g day:04 |g pages:289-304 |
856 | 4 | 0 | |u http://dx.doi.org/10.3233/BME-221488 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 34 |j 2023 |e 4 |b 04 |h 289-304 |