Technical performance of a dual-energy CT system with a novel deep-learning based reconstruction process : Evaluation using an abdomen protocol
© 2022 The Authors. Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine..
BACKGROUND: A new tube voltage-switching dual-energy (DE) CT system using a novel deep-learning based reconstruction process has been introduced. Characterizing the performance of this DE approach can help demonstrate its benefits and potential drawbacks.
PURPOSE: To evaluate the technical performance of a novel DECT system and compare it to that of standard single-kV CT and a rotate/rotate DECT, for abdominal imaging.
METHODS: DE and single-kV images of four different phantoms were acquired on a kV-switching DECT system, and on a rotate/rotate DECT. The dose for the acquisitions of each phantom was set to that selected for the kV-switching DE mode by the automatic tube current modulation (ATCM) at manufacturer-recommended settings. The dose that the ATCM would have selected in single-kV mode was also recorded. Virtual monochromatic images (VMIs) from 40 to 130 keV, as well as iodine maps, were reconstructed from the DE data. Single-kV images, acquired at 120 kV, were reconstructed using body hybrid iterative reconstruction. All reconstructions were made at 0.5 mm section thickness. Task transfer functions (TTFs) were determined for a Teflon and LDPE rod. Noise magnitude (SD), and noise power spectrum (NPS) were calculated using 240 and 320 mm diameter water phantoms. Iodine quantification accuracy and contrast-to-noise ratios (CNRs) relative to water for 2, 5, 10, and 15 mg I/ml were determined using a multi-energy CT (MECT) phantom. Low-contrast visibility was determined and the presence of beam-hardening artifacts and inhomogeneities were evaluated.
RESULTS: The TTFs of the kV-switching DE VMIs were higher than that of the single-kV images for Teflon (20% TTF: 6.8 lp/cm at 40 keV, 6.2 lp/cm for single-kV), while for LDPE the DE TTFs at 70 keV and above were equivalent or higher than the single-kV TTF. All TTFs of the kV-switching DECT were higher than for the rotate/rotate DECT. The SD was lowest in the 70 keV VMI (12.0 HU), which was lower than that of single-kV (18.3 HU). The average NPS frequency varied between 2.3 lp/cm and 4.2 lp/cm for the kV-switching VMIs and was 2.2 lp/cm for single-kV. The error in iodine quantification was at maximum 1 mg I/ml (at 5 mg I/ml). The highest CNR for all iodine concentrations was at 60 keV, 2.5 times higher than the CNR for single-kV. At 70-90 keV, the number of visible low contrast objects was comparable to that in single-kV, while other VMIs showed fewer objects. At manufacturer-recommended ATCM settings, the CTDIvol for the DE acquisitions of the water and MECT phantoms were 12.6 and 15.4 mGy, respectively, and higher than that for single-kV. The 70 keV VMI had less severe beam hardening artifacts than single-kV images. Hyper- and hypo-dense blotches may appear in VMIs when object attenuation exceeds manufacturer recommended limits.
CONCLUSIONS: At manufacturer-recommended ATCM settings for abdominal imaging, this DE implementation results in higher CTDIvol compared to single-kV acquisitions. However, it can create sharper, lower noise VMIs with up to 2.5 times higher iodine CNR compared to single-kV images acquired at the same dose.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2023 |
---|---|
Erschienen: |
2023 |
Enthalten in: |
Zur Gesamtaufnahme - volume:50 |
---|---|
Enthalten in: |
Medical physics - 50(2023), 3 vom: 11. März, Seite 1378-1389 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Oostveen, Luuk J [VerfasserIn] |
---|
Links: |
---|
Themen: |
9002-88-4 |
---|
Anmerkungen: |
Date Completed 21.03.2023 Date Revised 21.03.2023 published: Print-Electronic Citation Status MEDLINE |
---|
doi: |
10.1002/mp.16151 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM350146969 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM350146969 | ||
003 | DE-627 | ||
005 | 20231226044323.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231226s2023 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1002/mp.16151 |2 doi | |
028 | 5 | 2 | |a pubmed24n1167.xml |
035 | |a (DE-627)NLM350146969 | ||
035 | |a (NLM)36502496 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Oostveen, Luuk J |e verfasserin |4 aut | |
245 | 1 | 0 | |a Technical performance of a dual-energy CT system with a novel deep-learning based reconstruction process |b Evaluation using an abdomen protocol |
264 | 1 | |c 2023 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 21.03.2023 | ||
500 | |a Date Revised 21.03.2023 | ||
500 | |a published: Print-Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a © 2022 The Authors. Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine. | ||
520 | |a BACKGROUND: A new tube voltage-switching dual-energy (DE) CT system using a novel deep-learning based reconstruction process has been introduced. Characterizing the performance of this DE approach can help demonstrate its benefits and potential drawbacks | ||
520 | |a PURPOSE: To evaluate the technical performance of a novel DECT system and compare it to that of standard single-kV CT and a rotate/rotate DECT, for abdominal imaging | ||
520 | |a METHODS: DE and single-kV images of four different phantoms were acquired on a kV-switching DECT system, and on a rotate/rotate DECT. The dose for the acquisitions of each phantom was set to that selected for the kV-switching DE mode by the automatic tube current modulation (ATCM) at manufacturer-recommended settings. The dose that the ATCM would have selected in single-kV mode was also recorded. Virtual monochromatic images (VMIs) from 40 to 130 keV, as well as iodine maps, were reconstructed from the DE data. Single-kV images, acquired at 120 kV, were reconstructed using body hybrid iterative reconstruction. All reconstructions were made at 0.5 mm section thickness. Task transfer functions (TTFs) were determined for a Teflon and LDPE rod. Noise magnitude (SD), and noise power spectrum (NPS) were calculated using 240 and 320 mm diameter water phantoms. Iodine quantification accuracy and contrast-to-noise ratios (CNRs) relative to water for 2, 5, 10, and 15 mg I/ml were determined using a multi-energy CT (MECT) phantom. Low-contrast visibility was determined and the presence of beam-hardening artifacts and inhomogeneities were evaluated | ||
520 | |a RESULTS: The TTFs of the kV-switching DE VMIs were higher than that of the single-kV images for Teflon (20% TTF: 6.8 lp/cm at 40 keV, 6.2 lp/cm for single-kV), while for LDPE the DE TTFs at 70 keV and above were equivalent or higher than the single-kV TTF. All TTFs of the kV-switching DECT were higher than for the rotate/rotate DECT. The SD was lowest in the 70 keV VMI (12.0 HU), which was lower than that of single-kV (18.3 HU). The average NPS frequency varied between 2.3 lp/cm and 4.2 lp/cm for the kV-switching VMIs and was 2.2 lp/cm for single-kV. The error in iodine quantification was at maximum 1 mg I/ml (at 5 mg I/ml). The highest CNR for all iodine concentrations was at 60 keV, 2.5 times higher than the CNR for single-kV. At 70-90 keV, the number of visible low contrast objects was comparable to that in single-kV, while other VMIs showed fewer objects. At manufacturer-recommended ATCM settings, the CTDIvol for the DE acquisitions of the water and MECT phantoms were 12.6 and 15.4 mGy, respectively, and higher than that for single-kV. The 70 keV VMI had less severe beam hardening artifacts than single-kV images. Hyper- and hypo-dense blotches may appear in VMIs when object attenuation exceeds manufacturer recommended limits | ||
520 | |a CONCLUSIONS: At manufacturer-recommended ATCM settings for abdominal imaging, this DE implementation results in higher CTDIvol compared to single-kV acquisitions. However, it can create sharper, lower noise VMIs with up to 2.5 times higher iodine CNR compared to single-kV images acquired at the same dose | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a computed tomography | |
650 | 4 | |a dual energy and spectral | |
650 | 4 | |a technology assessment | |
650 | 7 | |a Polyethylene |2 NLM | |
650 | 7 | |a 9002-88-4 |2 NLM | |
650 | 7 | |a Iodine |2 NLM | |
650 | 7 | |a 9679TC07X4 |2 NLM | |
700 | 1 | |a Boedeker, Kirsten L |e verfasserin |4 aut | |
700 | 1 | |a Balta, Christiana |e verfasserin |4 aut | |
700 | 1 | |a Shin, Daniel |e verfasserin |4 aut | |
700 | 1 | |a de Lange, Frank |e verfasserin |4 aut | |
700 | 1 | |a Prokop, Mathias |e verfasserin |4 aut | |
700 | 1 | |a Sechopoulos, Ioannis |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Medical physics |d 1974 |g 50(2023), 3 vom: 11. März, Seite 1378-1389 |w (DE-627)NLM000294748 |x 2473-4209 |7 nnns |
773 | 1 | 8 | |g volume:50 |g year:2023 |g number:3 |g day:11 |g month:03 |g pages:1378-1389 |
856 | 4 | 0 | |u http://dx.doi.org/10.1002/mp.16151 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 50 |j 2023 |e 3 |b 11 |c 03 |h 1378-1389 |