Clinical Application of a Stereotactic Frame-Specific 3D-Printed Attachment for Deep Brain Stimulation Surgery
Copyright © 2022 Elsevier Inc. All rights reserved..
BACKGROUND: Mispositioning of microelectrodes during deep brain stimulation surgery can incur serious complications for patients. Current practice of creating a burr hole for introduction of the microelectrode is done freehand and can cause trajectory misalignment. We aimed to create a sterilizable surgical adjunct to minimize error from burr hole placement.
METHODS: We designed and demonstrated clinical use of a 3D-printed surgical jig that can be mounted to the current Cosman-Roberts-Wells stereotactic frame. The jig allowed accurate placement of the perforating burr for creation of the burr hole.
RESULTS: Intraoperative usage of the jig in 11 patients who underwent bilateral deep brain stimulation microelectrode placement for Parkinson disease demonstrated high accuracy of microelectrode placement, with an average 1.18 mm deviation (range, 0-2.7 mm) from intended trajectories. No intraoperative complications were encountered.
CONCLUSIONS: This proof-of-concept study highlights the utility of 3D-printed surgical adjuncts that are fully customizable and rapidly produced to improve current surgical practice. The jig reduced surgery duration, need for multiple trajectories, and risk of potentially devastating neurological complications. As demonstrated, 3D-printed devices are useful as surgical adjuncts to optimize safety and efficacy in deep brain stimulation surgeries.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2023 |
---|---|
Erschienen: |
2023 |
Enthalten in: |
Zur Gesamtaufnahme - volume:170 |
---|---|
Enthalten in: |
World neurosurgery - 170(2023) vom: 10. Feb., Seite e777-e783 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Ang, Jensen [VerfasserIn] |
---|
Links: |
---|
Themen: |
3D printed devices |
---|
Anmerkungen: |
Date Completed 15.02.2023 Date Revised 15.02.2023 published: Print-Electronic Citation Status MEDLINE |
---|
doi: |
10.1016/j.wneu.2022.11.121 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM349684618 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM349684618 | ||
003 | DE-627 | ||
005 | 20231226043220.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231226s2023 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1016/j.wneu.2022.11.121 |2 doi | |
028 | 5 | 2 | |a pubmed24n1165.xml |
035 | |a (DE-627)NLM349684618 | ||
035 | |a (NLM)36455844 | ||
035 | |a (PII)S1878-8750(22)01674-6 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Ang, Jensen |e verfasserin |4 aut | |
245 | 1 | 0 | |a Clinical Application of a Stereotactic Frame-Specific 3D-Printed Attachment for Deep Brain Stimulation Surgery |
264 | 1 | |c 2023 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 15.02.2023 | ||
500 | |a Date Revised 15.02.2023 | ||
500 | |a published: Print-Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a Copyright © 2022 Elsevier Inc. All rights reserved. | ||
520 | |a BACKGROUND: Mispositioning of microelectrodes during deep brain stimulation surgery can incur serious complications for patients. Current practice of creating a burr hole for introduction of the microelectrode is done freehand and can cause trajectory misalignment. We aimed to create a sterilizable surgical adjunct to minimize error from burr hole placement | ||
520 | |a METHODS: We designed and demonstrated clinical use of a 3D-printed surgical jig that can be mounted to the current Cosman-Roberts-Wells stereotactic frame. The jig allowed accurate placement of the perforating burr for creation of the burr hole | ||
520 | |a RESULTS: Intraoperative usage of the jig in 11 patients who underwent bilateral deep brain stimulation microelectrode placement for Parkinson disease demonstrated high accuracy of microelectrode placement, with an average 1.18 mm deviation (range, 0-2.7 mm) from intended trajectories. No intraoperative complications were encountered | ||
520 | |a CONCLUSIONS: This proof-of-concept study highlights the utility of 3D-printed surgical adjuncts that are fully customizable and rapidly produced to improve current surgical practice. The jig reduced surgery duration, need for multiple trajectories, and risk of potentially devastating neurological complications. As demonstrated, 3D-printed devices are useful as surgical adjuncts to optimize safety and efficacy in deep brain stimulation surgeries | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a 3D printed devices | |
650 | 4 | |a 3D printing | |
650 | 4 | |a Deep brain stimulation | |
700 | 1 | |a Zhang, John J Y |e verfasserin |4 aut | |
700 | 1 | |a Yam, Michael |e verfasserin |4 aut | |
700 | 1 | |a Maszczyk, Tomasz |e verfasserin |4 aut | |
700 | 1 | |a Ng, Wai Hoe |e verfasserin |4 aut | |
700 | 1 | |a Wan, Kai Rui |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t World neurosurgery |d 2010 |g 170(2023) vom: 10. Feb., Seite e777-e783 |w (DE-627)NLM197924441 |x 1878-8769 |7 nnns |
773 | 1 | 8 | |g volume:170 |g year:2023 |g day:10 |g month:02 |g pages:e777-e783 |
856 | 4 | 0 | |u http://dx.doi.org/10.1016/j.wneu.2022.11.121 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 170 |j 2023 |b 10 |c 02 |h e777-e783 |