Assessment of Anastomotic Viability With Spectroscopic Real-time Oxygen Saturation Measurement in a Porcine Study
OBJECTIVE: Anastomotic leakage (AL) is a severe complication following intestinal procedures. Intra.Ox™ by ViOptix Inc (Newark, CA, USA) is a novel, FDA-approved spectroscopic device which enables real-time measurement of mixed tissue oxygen saturation (StO2). Using a porcine model, this study explores the correlation between StO2 measurements and AL formation as well as investigates the applicability of Intra.Ox™ in the clinical setting.
METHODS: Eleven female swine were divided into 3 groups to explore AL formation in different ischemic conditions. Group 1: 100% mesenteric-vascular ligation, n = 3; Group 2: 50% ligation, n = 5; Group 3: No mesenteric ligation, n = 3. StO2 at the anastomotic line was measured before and after vessel ligation and anastomosis. Measurements were taken at 6 distinct locations along afferent and efferent loops. AL was evaluated on postoperative day 5 by re-laparotomy.
RESULTS: AL rate was 100%, 60% and 0% in groups 1, 2 and 3, respectively. Post-anastomotic StO2 in group 1 (22.9 ± 18.5%) and 2 (39.2 ± 20.1%) were significantly lower than in group 3 (53.1 ± 8.3%, p<.0001). Post-anastomotic StO2 readings ≤40% indicated AL potential with 100% sensitivity,+ 80% specificity, positive predictive value of 85.7% and negative predictive value of 100%.
CONCLUSION: This study demonstrates the value of Intra.Ox™ in assessing local perfusion and indicate the association between low StO2 and AL by providing accurate, real-time, noninvasive tissue oxygenation measurements at anastomotic sites. Further studies are required to investigate the clinical application of this novel device in intestinal surgery.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2023 |
---|---|
Erschienen: |
2023 |
Enthalten in: |
Zur Gesamtaufnahme - volume:30 |
---|---|
Enthalten in: |
Surgical innovation - 30(2023), 3 vom: 01. Juni, Seite 349-355 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Khosrawipour, Tanja [VerfasserIn] |
---|
Links: |
---|
Themen: |
Anastomotic leak |
---|
Anmerkungen: |
Date Completed 10.07.2023 Date Revised 18.07.2023 published: Print-Electronic Citation Status MEDLINE |
---|
doi: |
10.1177/15533506221127378 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM347700780 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM347700780 | ||
003 | DE-627 | ||
005 | 20231226034443.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231226s2023 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1177/15533506221127378 |2 doi | |
028 | 5 | 2 | |a pubmed24n1158.xml |
035 | |a (DE-627)NLM347700780 | ||
035 | |a (NLM)36255230 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Khosrawipour, Tanja |e verfasserin |4 aut | |
245 | 1 | 0 | |a Assessment of Anastomotic Viability With Spectroscopic Real-time Oxygen Saturation Measurement in a Porcine Study |
264 | 1 | |c 2023 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 10.07.2023 | ||
500 | |a Date Revised 18.07.2023 | ||
500 | |a published: Print-Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a OBJECTIVE: Anastomotic leakage (AL) is a severe complication following intestinal procedures. Intra.Ox™ by ViOptix Inc (Newark, CA, USA) is a novel, FDA-approved spectroscopic device which enables real-time measurement of mixed tissue oxygen saturation (StO2). Using a porcine model, this study explores the correlation between StO2 measurements and AL formation as well as investigates the applicability of Intra.Ox™ in the clinical setting | ||
520 | |a METHODS: Eleven female swine were divided into 3 groups to explore AL formation in different ischemic conditions. Group 1: 100% mesenteric-vascular ligation, n = 3; Group 2: 50% ligation, n = 5; Group 3: No mesenteric ligation, n = 3. StO2 at the anastomotic line was measured before and after vessel ligation and anastomosis. Measurements were taken at 6 distinct locations along afferent and efferent loops. AL was evaluated on postoperative day 5 by re-laparotomy | ||
520 | |a RESULTS: AL rate was 100%, 60% and 0% in groups 1, 2 and 3, respectively. Post-anastomotic StO2 in group 1 (22.9 ± 18.5%) and 2 (39.2 ± 20.1%) were significantly lower than in group 3 (53.1 ± 8.3%, p<.0001). Post-anastomotic StO2 readings ≤40% indicated AL potential with 100% sensitivity,+ 80% specificity, positive predictive value of 85.7% and negative predictive value of 100% | ||
520 | |a CONCLUSION: This study demonstrates the value of Intra.Ox™ in assessing local perfusion and indicate the association between low StO2 and AL by providing accurate, real-time, noninvasive tissue oxygenation measurements at anastomotic sites. Further studies are required to investigate the clinical application of this novel device in intestinal surgery | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a Intra.Ox | |
650 | 4 | |a anastomotic leak | |
650 | 4 | |a prediction | |
650 | 4 | |a tissue oximetry | |
650 | 4 | |a vioptix | |
700 | 1 | |a Li, Shiri |e verfasserin |4 aut | |
700 | 1 | |a Steward, Earl |e verfasserin |4 aut | |
700 | 1 | |a Chaudhry, Haris |e verfasserin |4 aut | |
700 | 1 | |a Nguyen, Danielle |e verfasserin |4 aut | |
700 | 1 | |a Khosrawipour, Veria |e verfasserin |4 aut | |
700 | 1 | |a Jafari, Mehraneh Dorna |e verfasserin |4 aut | |
700 | 1 | |a Pigazzi, Alessio |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Surgical innovation |d 2004 |g 30(2023), 3 vom: 01. Juni, Seite 349-355 |w (DE-627)NLM154108928 |x 1553-3514 |7 nnns |
773 | 1 | 8 | |g volume:30 |g year:2023 |g number:3 |g day:01 |g month:06 |g pages:349-355 |
856 | 4 | 0 | |u http://dx.doi.org/10.1177/15533506221127378 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 30 |j 2023 |e 3 |b 01 |c 06 |h 349-355 |