An in vitro evaluation of the WIRELE-x electronic apex locator

© 2022 The Authors. Australian Endodontic Journal published by John Wiley & Sons Australia, Ltd on behalf of Australian Society of Endodontology Inc..

This study evaluated the accuracy of the Wirele-X (Forum Tec, Ashkelon, Israel), a novel Bluetooth-enabled wireless electronic apex locator. Thirty-one extracted teeth with mature apices were used. Under 10X magnification, the actual canal lengths were determined. The teeth were embedded in alginate and electronic canal lengths were obtained using the Root ZX II and Wirele-X electronic apex locators. The actual canal lengths and electronic canal lengths were compared with Student's t-test. The average distance from the file tip to the actual canal length was -0.11 mm (±0.16) for the Root ZX II, and - 0.07 mm (±0.21) for the Wirele-X. There were no statistically significant differences between the two electronic apex locators in their ability to determine the actual canal length (p > 0.05). The wireless apex locator (Wirele-X) and the wired apex locator (Root ZX II) were found to be equally accurate.

Medienart:

E-Artikel

Erscheinungsjahr:

2023

Erschienen:

2023

Enthalten in:

Zur Gesamtaufnahme - volume:49 Suppl 1

Enthalten in:

Australian endodontic journal : the journal of the Australian Society of Endodontology Inc - 49 Suppl 1(2023) vom: 22. Sept., Seite 41-45

Sprache:

Englisch

Beteiligte Personen:

Brand, Lucas [VerfasserIn]
Dunlap, Craig A [VerfasserIn]
Scott, Raymond [VerfasserIn]
Arias, Anna [VerfasserIn]
Peters, Ove [VerfasserIn]

Links:

Volltext

Themen:

Electronic apex locator
Journal Article
Root ZX II
Root canal length
Wirele-X
Working length