The accuracies of three intraoral scanners with regard to shade determination : An in vitro study
© 2022 by the American College of Prosthodontists..
PURPOSE: To compare the accuracies of three intraoral scanners for shade determination function in vitro, and to preliminarily investigate the shade-matching characteristics of the three intraoral scanners.
MATERIALS AND METHODS: The shade of the middle third region of each shade tab on the Vita Classical A1-D4 shade guide (VC) was measured with a spectrophotometer (Vita Easyshade V, VE) and three intraoral scanners, including CEREC Omnicam (OM), 3Shape TRIOS 3 (T3), and TRIOS 4 (T4). A conversion table between VC values and CIELAB values was established from the database of VE to analyze the trueness. The reproducibility of the instruments was then compared by repeating the measurements five times.
RESULTS: The mean color difference for each instrument was highest in the OM, followed by the T4, and lowest in the T3 and VE, respectively. The L* and a* value for OM, and the b* value for T4, were significantly different from those for VE (p <0.05). The reproducibility of the instrument was highest in the VE (Fleiss' kappa: 0.95), followed by the T3 (Fleiss' kappa: 0.89), T4 (Fleiss' kappa: 0.87), and OM (Fleiss' kappa: 0.78).
CONCLUSIONS: Of the three intraoral scanners, the trueness was best on the T3. The reproducibility of all the instruments was excellent.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2023 |
---|---|
Erschienen: |
2023 |
Enthalten in: |
Zur Gesamtaufnahme - volume:32 |
---|---|
Enthalten in: |
Journal of prosthodontics : official journal of the American College of Prosthodontists - 32(2023), 5 vom: 29. Juni, Seite 111-117 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Huang, Mingming [VerfasserIn] |
---|
Links: |
---|
Themen: |
Intraoral scanner |
---|
Anmerkungen: |
Date Completed 07.06.2023 Date Revised 07.06.2023 published: Print-Electronic Citation Status MEDLINE |
---|
doi: |
10.1111/jopr.13571 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM343690969 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM343690969 | ||
003 | DE-627 | ||
005 | 20231226021034.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231226s2023 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1111/jopr.13571 |2 doi | |
028 | 5 | 2 | |a pubmed24n1145.xml |
035 | |a (DE-627)NLM343690969 | ||
035 | |a (NLM)35848886 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Huang, Mingming |e verfasserin |4 aut | |
245 | 1 | 4 | |a The accuracies of three intraoral scanners with regard to shade determination |b An in vitro study |
264 | 1 | |c 2023 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 07.06.2023 | ||
500 | |a Date Revised 07.06.2023 | ||
500 | |a published: Print-Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a © 2022 by the American College of Prosthodontists. | ||
520 | |a PURPOSE: To compare the accuracies of three intraoral scanners for shade determination function in vitro, and to preliminarily investigate the shade-matching characteristics of the three intraoral scanners | ||
520 | |a MATERIALS AND METHODS: The shade of the middle third region of each shade tab on the Vita Classical A1-D4 shade guide (VC) was measured with a spectrophotometer (Vita Easyshade V, VE) and three intraoral scanners, including CEREC Omnicam (OM), 3Shape TRIOS 3 (T3), and TRIOS 4 (T4). A conversion table between VC values and CIELAB values was established from the database of VE to analyze the trueness. The reproducibility of the instruments was then compared by repeating the measurements five times | ||
520 | |a RESULTS: The mean color difference for each instrument was highest in the OM, followed by the T4, and lowest in the T3 and VE, respectively. The L* and a* value for OM, and the b* value for T4, were significantly different from those for VE (p <0.05). The reproducibility of the instrument was highest in the VE (Fleiss' kappa: 0.95), followed by the T3 (Fleiss' kappa: 0.89), T4 (Fleiss' kappa: 0.87), and OM (Fleiss' kappa: 0.78) | ||
520 | |a CONCLUSIONS: Of the three intraoral scanners, the trueness was best on the T3. The reproducibility of all the instruments was excellent | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a intraoral scanner | |
650 | 4 | |a shade guide | |
650 | 4 | |a shade matching | |
650 | 4 | |a spectrophotometer | |
700 | 1 | |a Wang, Yong |e verfasserin |4 aut | |
700 | 1 | |a Sun, Yuchun |e verfasserin |4 aut | |
700 | 1 | |a Zhou, Yongsheng |e verfasserin |4 aut | |
700 | 1 | |a Liu, Yunsong |e verfasserin |4 aut | |
700 | 1 | |a Ye, Hongqiang |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Journal of prosthodontics : official journal of the American College of Prosthodontists |d 1996 |g 32(2023), 5 vom: 29. Juni, Seite 111-117 |w (DE-627)NLM012891819 |x 1532-849X |7 nnns |
773 | 1 | 8 | |g volume:32 |g year:2023 |g number:5 |g day:29 |g month:06 |g pages:111-117 |
856 | 4 | 0 | |u http://dx.doi.org/10.1111/jopr.13571 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 32 |j 2023 |e 5 |b 29 |c 06 |h 111-117 |