Pseudoaneurysm Formation After "Preclose"-Assisted Impella Insertion in a Patient With Cardiogenic Shock
Copyright 2022, Asemota et al..
The use of mechanical support devices such as the Impella CP (Abiomed, Danvers, MA) is a growing form of treatment for patients with cardiogenic shock (CS). Despite the increase in usage, there remains a dearth in literature regarding potential complications. Vascular complications such as pseudoaneurysms (PAs) are rare but important potential complications that can occur with use of the Impella. We present Impella-assisted percutaneous coronary intervention (PCI) in a patient with CS, "Preclosed" with the Perclose ProGlide (Abbott, Plymouth, MN) device complicated by development of a PA. A 62-year-old male patient with a history of diabetes and hypertension presented to our emergency room (ER) with chest pain and electrocardiogram (ECG) findings consistent with an acute anterior wall ST-elevation myocardial infarction (STEMI). This was further complicated by refractory CS. The patient was urgently taken to the cardiac catherization laboratory. After exchange of sequential dilators, a single Perclose device was used prior to the insertion of the Impella sheath. The patient then underwent a successful Impella-assisted PCI of his left anterior descending artery. Upon stabilization of hemodynamics, the patient was taken to the catheterization laboratory for Impella removal. After removal of Impella, imaging detected extravasation of contrast, without development of hematoma, later confirmed to be a PA via computed tomography (CT) scans and ultrasound Doppler imaging. The PA was successfully managed with injection of thrombin. The PA was likely caused by shearing forces of the dilators, the 14-F Impella sheath and foot of the device. We propose deploying the Perclose device earlier in the process of dilating the access site to avoid such complication. This is one of the first case reports that detail the occurrence and management of a PA with Impella insertion.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2022 |
---|---|
Erschienen: |
2022 |
Enthalten in: |
Zur Gesamtaufnahme - volume:13 |
---|---|
Enthalten in: |
Journal of medical cases - 13(2022), 5 vom: 11. Mai, Seite 202-206 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Asemota, Daniel [VerfasserIn] |
---|
Links: |
---|
Themen: |
Cardiogenic shock |
---|
Anmerkungen: |
Date Revised 16.09.2023 published: Print-Electronic Citation Status PubMed-not-MEDLINE |
---|
doi: |
10.14740/jmc3841 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM341775878 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM341775878 | ||
003 | DE-627 | ||
005 | 20231226012543.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231226s2022 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.14740/jmc3841 |2 doi | |
028 | 5 | 2 | |a pubmed24n1139.xml |
035 | |a (DE-627)NLM341775878 | ||
035 | |a (NLM)35655626 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Asemota, Daniel |e verfasserin |4 aut | |
245 | 1 | 0 | |a Pseudoaneurysm Formation After "Preclose"-Assisted Impella Insertion in a Patient With Cardiogenic Shock |
264 | 1 | |c 2022 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Revised 16.09.2023 | ||
500 | |a published: Print-Electronic | ||
500 | |a Citation Status PubMed-not-MEDLINE | ||
520 | |a Copyright 2022, Asemota et al. | ||
520 | |a The use of mechanical support devices such as the Impella CP (Abiomed, Danvers, MA) is a growing form of treatment for patients with cardiogenic shock (CS). Despite the increase in usage, there remains a dearth in literature regarding potential complications. Vascular complications such as pseudoaneurysms (PAs) are rare but important potential complications that can occur with use of the Impella. We present Impella-assisted percutaneous coronary intervention (PCI) in a patient with CS, "Preclosed" with the Perclose ProGlide (Abbott, Plymouth, MN) device complicated by development of a PA. A 62-year-old male patient with a history of diabetes and hypertension presented to our emergency room (ER) with chest pain and electrocardiogram (ECG) findings consistent with an acute anterior wall ST-elevation myocardial infarction (STEMI). This was further complicated by refractory CS. The patient was urgently taken to the cardiac catherization laboratory. After exchange of sequential dilators, a single Perclose device was used prior to the insertion of the Impella sheath. The patient then underwent a successful Impella-assisted PCI of his left anterior descending artery. Upon stabilization of hemodynamics, the patient was taken to the catheterization laboratory for Impella removal. After removal of Impella, imaging detected extravasation of contrast, without development of hematoma, later confirmed to be a PA via computed tomography (CT) scans and ultrasound Doppler imaging. The PA was successfully managed with injection of thrombin. The PA was likely caused by shearing forces of the dilators, the 14-F Impella sheath and foot of the device. We propose deploying the Perclose device earlier in the process of dilating the access site to avoid such complication. This is one of the first case reports that detail the occurrence and management of a PA with Impella insertion | ||
650 | 4 | |a Case Reports | |
650 | 4 | |a Cardiogenic shock | |
650 | 4 | |a Impella complication | |
650 | 4 | |a Pseudoaneurysm | |
700 | 1 | |a Kassam, Zain |e verfasserin |4 aut | |
700 | 1 | |a Voto, Christian |e verfasserin |4 aut | |
700 | 1 | |a Mangla, Aditya |e verfasserin |4 aut | |
700 | 1 | |a Coven, David |e verfasserin |4 aut | |
700 | 1 | |a Lasic, Zoran |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Journal of medical cases |d 2010 |g 13(2022), 5 vom: 11. Mai, Seite 202-206 |w (DE-627)NLM207988420 |x 1923-4163 |7 nnns |
773 | 1 | 8 | |g volume:13 |g year:2022 |g number:5 |g day:11 |g month:05 |g pages:202-206 |
856 | 4 | 0 | |u http://dx.doi.org/10.14740/jmc3841 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 13 |j 2022 |e 5 |b 11 |c 05 |h 202-206 |