Quantitative Image Quality Metrics of the Low-Dose 2D/3D Slot Scanner Compared to Two Conventional Digital Radiography X-ray Imaging Systems
The aim of this study was to determine the quantitative image quality metrics of the low-dose 2D/3D EOS slot scanner X-ray imaging system (LDSS) compared with conventional digital radiography (DR) X-ray imaging systems. The effective detective quantum efficiency (eDQE) and effective noise quantum equivalent (eNEQ) were measured using chest and knee protocols.
METHODS: A Nationwide Evaluation of X-ray Trends (NEXT) of a chest adult phantom and a PolyMethylmethacrylate (PMMA) phantom were used for the chest and knee protocols, respectively. Quantitative image quality metrics, including effective normalised noise power spectrum (eNNPS), effective modulation transfer function (eMTF), eDQE and eNEQ of the LDSS and DR imaging systems were assessed and compared.
RESULTS: In the chest acquisition, the LDSS imaging system achieved significantly higher eNEQ and eDQE than the DR imaging systems at lower and higher spatial frequencies (0.001 ≤ p ≤ 0.044). For the knee acquisition, the LDSS imaging system also achieved significantly higher eNEQ and eDQE than the DR imaging systems at lower and higher spatial frequencies (0.001 ≤ p ≤ 0.002). However, there was no significant difference in eNEQ and eDQE between DR systems 1 and 2 at lower and higher spatial frequencies (0.10 < p < 1.00) for either chest or knee protocols.
CONCLUSION: The LDSS imaging system performed well compared to the DR systems. Thus, we have demonstrated that the LDSS imaging system has the potential to be used for clinical diagnostic purposes.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2021 |
---|---|
Erschienen: |
2021 |
Enthalten in: |
Zur Gesamtaufnahme - volume:11 |
---|---|
Enthalten in: |
Diagnostics (Basel, Switzerland) - 11(2021), 9 vom: 17. Sept. |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Abdi, Ahmed Jibril [VerfasserIn] |
---|
Links: |
---|
Themen: |
Chest X-ray |
---|
Anmerkungen: |
Date Revised 07.11.2023 published: Electronic Citation Status PubMed-not-MEDLINE |
---|
doi: |
10.3390/diagnostics11091699 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM331156504 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM331156504 | ||
003 | DE-627 | ||
005 | 20231225212835.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231225s2021 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.3390/diagnostics11091699 |2 doi | |
028 | 5 | 2 | |a pubmed24n1103.xml |
035 | |a (DE-627)NLM331156504 | ||
035 | |a (NLM)34574041 | ||
035 | |a (PII)1699 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Abdi, Ahmed Jibril |e verfasserin |4 aut | |
245 | 1 | 0 | |a Quantitative Image Quality Metrics of the Low-Dose 2D/3D Slot Scanner Compared to Two Conventional Digital Radiography X-ray Imaging Systems |
264 | 1 | |c 2021 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Revised 07.11.2023 | ||
500 | |a published: Electronic | ||
500 | |a Citation Status PubMed-not-MEDLINE | ||
520 | |a The aim of this study was to determine the quantitative image quality metrics of the low-dose 2D/3D EOS slot scanner X-ray imaging system (LDSS) compared with conventional digital radiography (DR) X-ray imaging systems. The effective detective quantum efficiency (eDQE) and effective noise quantum equivalent (eNEQ) were measured using chest and knee protocols | ||
520 | |a METHODS: A Nationwide Evaluation of X-ray Trends (NEXT) of a chest adult phantom and a PolyMethylmethacrylate (PMMA) phantom were used for the chest and knee protocols, respectively. Quantitative image quality metrics, including effective normalised noise power spectrum (eNNPS), effective modulation transfer function (eMTF), eDQE and eNEQ of the LDSS and DR imaging systems were assessed and compared | ||
520 | |a RESULTS: In the chest acquisition, the LDSS imaging system achieved significantly higher eNEQ and eDQE than the DR imaging systems at lower and higher spatial frequencies (0.001 ≤ p ≤ 0.044). For the knee acquisition, the LDSS imaging system also achieved significantly higher eNEQ and eDQE than the DR imaging systems at lower and higher spatial frequencies (0.001 ≤ p ≤ 0.002). However, there was no significant difference in eNEQ and eDQE between DR systems 1 and 2 at lower and higher spatial frequencies (0.10 < p < 1.00) for either chest or knee protocols | ||
520 | |a CONCLUSION: The LDSS imaging system performed well compared to the DR systems. Thus, we have demonstrated that the LDSS imaging system has the potential to be used for clinical diagnostic purposes | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a chest X-ray | |
650 | 4 | |a knee X-ray | |
650 | 4 | |a quantitative image quality metrics | |
700 | 1 | |a Mussmann, Bo R |e verfasserin |4 aut | |
700 | 1 | |a Mackenzie, Alistair |e verfasserin |4 aut | |
700 | 1 | |a Gerke, Oke |e verfasserin |4 aut | |
700 | 1 | |a Klaerke, Benedikte |e verfasserin |4 aut | |
700 | 1 | |a Andersen, Poul Erik |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Diagnostics (Basel, Switzerland) |d 2011 |g 11(2021), 9 vom: 17. Sept. |w (DE-627)NLM250111136 |x 2075-4418 |7 nnns |
773 | 1 | 8 | |g volume:11 |g year:2021 |g number:9 |g day:17 |g month:09 |
856 | 4 | 0 | |u http://dx.doi.org/10.3390/diagnostics11091699 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 11 |j 2021 |e 9 |b 17 |c 09 |