Single limb electrocardiogram using vector mapping : Evaluation and validation of a novel medical device
Copyright © 2021 Elsevier Inc. All rights reserved..
INTRODUCTION: The Tiger Tech Warfighter Monitor (WFM) is a novel single-limb device for ECG acquisition. The WFM provides true (not derived) single limb Electrocardiogram monitoring (ECG) to provide heart rate and R-R interval monitoring between QRS complexes. Herein, we evaluate the diagnostic accuracy of the WFM heart rate, R-R interval monitoring, and heart rate variability monitoring in comparison to a 2‑lead chest ECG.
METHODS: Data was collected under Institutional Review Board (IRB) approval. Patients available within our institution's pre-operative holding unit were randomly selected to undergo simultaneous chest and WFM ECG monitoring. 3-5-min measurements were taken depending on the patient's availability. Data was saved to two separate mobile phones and time-stamped for synchronization. A proprietary Tiger Tech extraction algorithm was used to tag proper features on both the WFM 1-Limb ECG and Chest ECG data files. A separate algorithm was then used to compare the beat-to-beat variations between the ECGs.
RESULTS: Data was extracted and analyzed on 26 subjects. Linear regression of heart rate analysis revealed excellent correlations with an R2 of 0.99 (p < 0.05). Similar linear regression evaluation of R-R interval correlation demonstrated a mean R2 value of 0.95 (p < 0.05). Statistically significant correlation was achieved in all 26 included study participants. Heart rate variability also achieved excellent correlation (SDNN R2 = 0.997, RMSSD R2 = 0.995, LnRMSSD R2 = 0.992, p << 0.05).
CONCLUSION: Results demonstrate that the WFM achieves excellent correlation with chest ECG for heart rate, R-R internals, and heart rate variability.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2021 |
---|---|
Erschienen: |
2021 |
Enthalten in: |
Zur Gesamtaufnahme - volume:67 |
---|---|
Enthalten in: |
Journal of electrocardiology - 67(2021) vom: 01. Juli, Seite 136-141 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Peck, Jacquelin [VerfasserIn] |
---|
Links: |
---|
Themen: |
Clinical research |
---|
Anmerkungen: |
Date Completed 01.10.2021 Date Revised 01.10.2021 published: Print-Electronic Citation Status MEDLINE |
---|
doi: |
10.1016/j.jelectrocard.2021.06.003 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM327887877 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM327887877 | ||
003 | DE-627 | ||
005 | 20231225201758.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231225s2021 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1016/j.jelectrocard.2021.06.003 |2 doi | |
028 | 5 | 2 | |a pubmed24n1092.xml |
035 | |a (DE-627)NLM327887877 | ||
035 | |a (NLM)34242911 | ||
035 | |a (PII)S0022-0736(21)00121-7 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Peck, Jacquelin |e verfasserin |4 aut | |
245 | 1 | 0 | |a Single limb electrocardiogram using vector mapping |b Evaluation and validation of a novel medical device |
264 | 1 | |c 2021 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 01.10.2021 | ||
500 | |a Date Revised 01.10.2021 | ||
500 | |a published: Print-Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a Copyright © 2021 Elsevier Inc. All rights reserved. | ||
520 | |a INTRODUCTION: The Tiger Tech Warfighter Monitor (WFM) is a novel single-limb device for ECG acquisition. The WFM provides true (not derived) single limb Electrocardiogram monitoring (ECG) to provide heart rate and R-R interval monitoring between QRS complexes. Herein, we evaluate the diagnostic accuracy of the WFM heart rate, R-R interval monitoring, and heart rate variability monitoring in comparison to a 2‑lead chest ECG | ||
520 | |a METHODS: Data was collected under Institutional Review Board (IRB) approval. Patients available within our institution's pre-operative holding unit were randomly selected to undergo simultaneous chest and WFM ECG monitoring. 3-5-min measurements were taken depending on the patient's availability. Data was saved to two separate mobile phones and time-stamped for synchronization. A proprietary Tiger Tech extraction algorithm was used to tag proper features on both the WFM 1-Limb ECG and Chest ECG data files. A separate algorithm was then used to compare the beat-to-beat variations between the ECGs | ||
520 | |a RESULTS: Data was extracted and analyzed on 26 subjects. Linear regression of heart rate analysis revealed excellent correlations with an R2 of 0.99 (p < 0.05). Similar linear regression evaluation of R-R interval correlation demonstrated a mean R2 value of 0.95 (p < 0.05). Statistically significant correlation was achieved in all 26 included study participants. Heart rate variability also achieved excellent correlation (SDNN R2 = 0.997, RMSSD R2 = 0.995, LnRMSSD R2 = 0.992, p << 0.05) | ||
520 | |a CONCLUSION: Results demonstrate that the WFM achieves excellent correlation with chest ECG for heart rate, R-R internals, and heart rate variability | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a Research Support, U.S. Gov't, Non-P.H.S. | |
650 | 4 | |a Clinical research | |
650 | 4 | |a Device | |
650 | 4 | |a Electrocardiogram | |
700 | 1 | |a Wishon, Michael J |e verfasserin |4 aut | |
700 | 1 | |a Wittels, Harrison |e verfasserin |4 aut | |
700 | 1 | |a Lee, Stephen J |e verfasserin |4 aut | |
700 | 1 | |a Hendricks, Stephanie |e verfasserin |4 aut | |
700 | 1 | |a Davila, Hector |e verfasserin |4 aut | |
700 | 1 | |a Wittels, S Howard |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Journal of electrocardiology |d 1968 |g 67(2021) vom: 01. Juli, Seite 136-141 |w (DE-627)NLM000525014 |x 1532-8430 |7 nnns |
773 | 1 | 8 | |g volume:67 |g year:2021 |g day:01 |g month:07 |g pages:136-141 |
856 | 4 | 0 | |u http://dx.doi.org/10.1016/j.jelectrocard.2021.06.003 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 67 |j 2021 |b 01 |c 07 |h 136-141 |