Miniaturized and IoT Enabled Continuous-Flow-Based Microfluidic PCR Device for DNA Amplification
Herein, a continuous-flow driven microfluidic device has been designed and fabricated using the CO2 laser ablation method for polymerase chain reaction (PCR). The device consists of a polymethyl methacrylate (PMMA) microfluidic channel with 30 serpentine thermal cycles, an arduino board, two custom-made cartridge heaters, and thermocouple sensors. The portable thermal management system, with aluminium blocks placed on a wooden substrate, working on the PID controller principle, is low-cost, battery-powered, automated, integrated, and IoT-enabled. The device with dimensions 80×72×36 mm3 (L × W × H) has a temperature accuracy of ±0.2 °C. The IoT module enables accessing and storage of real-time temperature values directly onto the smartphone through ThingSpeak analytics. It was developed to achieve desirable accurate temperature at two thermal zones, denaturation and annealing (95 °C and 60 °C) on the microfluidic thermal management platform. A PCR mixture of [Formula: see text] was infused into the serpentine-based microchannel using a syringe pump. Amplification of DNA template with 594-base pair (bp) fragment of the rat GAPDH gene was successfully performed on the miniaturized thermal management system. The total time required for a complete PCR reaction was 32 min at an optimum flow rate of [Formula: see text]/min. The amplified sample of the target DNA obtained from the PCR microchannel was then separated by agarose gel electrophoresis and was further analyzed using a gel-doc system. Finally, the obtained results were compared to the conventional PCR instrument showing excellent performance.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2022 |
---|---|
Erschienen: |
2022 |
Enthalten in: |
Zur Gesamtaufnahme - volume:21 |
---|---|
Enthalten in: |
IEEE transactions on nanobioscience - 21(2022), 1 vom: 25. Jan., Seite 97-104 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Kulkarni, Madhusudan B [VerfasserIn] |
---|
Links: |
---|
Themen: |
---|
Anmerkungen: |
Date Completed 03.01.2022 Date Revised 03.01.2022 published: Print-Electronic Citation Status MEDLINE |
---|
doi: |
10.1109/TNB.2021.3092292 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM327179449 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM327179449 | ||
003 | DE-627 | ||
005 | 20231225200213.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231225s2022 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1109/TNB.2021.3092292 |2 doi | |
028 | 5 | 2 | |a pubmed24n1090.xml |
035 | |a (DE-627)NLM327179449 | ||
035 | |a (NLM)34170829 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Kulkarni, Madhusudan B |e verfasserin |4 aut | |
245 | 1 | 0 | |a Miniaturized and IoT Enabled Continuous-Flow-Based Microfluidic PCR Device for DNA Amplification |
264 | 1 | |c 2022 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 03.01.2022 | ||
500 | |a Date Revised 03.01.2022 | ||
500 | |a published: Print-Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a Herein, a continuous-flow driven microfluidic device has been designed and fabricated using the CO2 laser ablation method for polymerase chain reaction (PCR). The device consists of a polymethyl methacrylate (PMMA) microfluidic channel with 30 serpentine thermal cycles, an arduino board, two custom-made cartridge heaters, and thermocouple sensors. The portable thermal management system, with aluminium blocks placed on a wooden substrate, working on the PID controller principle, is low-cost, battery-powered, automated, integrated, and IoT-enabled. The device with dimensions 80×72×36 mm3 (L × W × H) has a temperature accuracy of ±0.2 °C. The IoT module enables accessing and storage of real-time temperature values directly onto the smartphone through ThingSpeak analytics. It was developed to achieve desirable accurate temperature at two thermal zones, denaturation and annealing (95 °C and 60 °C) on the microfluidic thermal management platform. A PCR mixture of [Formula: see text] was infused into the serpentine-based microchannel using a syringe pump. Amplification of DNA template with 594-base pair (bp) fragment of the rat GAPDH gene was successfully performed on the miniaturized thermal management system. The total time required for a complete PCR reaction was 32 min at an optimum flow rate of [Formula: see text]/min. The amplified sample of the target DNA obtained from the PCR microchannel was then separated by agarose gel electrophoresis and was further analyzed using a gel-doc system. Finally, the obtained results were compared to the conventional PCR instrument showing excellent performance | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a Research Support, Non-U.S. Gov't | |
650 | 7 | |a DNA |2 NLM | |
650 | 7 | |a 9007-49-2 |2 NLM | |
700 | 1 | |a Goyal, Srashti |e verfasserin |4 aut | |
700 | 1 | |a Dhar, Arti |e verfasserin |4 aut | |
700 | 1 | |a Sriram, D |e verfasserin |4 aut | |
700 | 1 | |a Goel, Sanket |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t IEEE transactions on nanobioscience |d 2002 |g 21(2022), 1 vom: 25. Jan., Seite 97-104 |w (DE-627)NLM150513186 |x 1558-2639 |7 nnns |
773 | 1 | 8 | |g volume:21 |g year:2022 |g number:1 |g day:25 |g month:01 |g pages:97-104 |
856 | 4 | 0 | |u http://dx.doi.org/10.1109/TNB.2021.3092292 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 21 |j 2022 |e 1 |b 25 |c 01 |h 97-104 |