Oxygen delivery failure due to improper installation of carbon dioxide absorbent canister : a case report
Copyright © 2021 Sociedade Brasileira de Anestesiologia. Published by Elsevier Editora Ltda. All rights reserved..
The CLIC system in the Dräger Apollo anesthesia workstation allows a successful pre-use machine checkout without the presence of a carbon dioxide absorbent canister. It also allows the canister to be changed without interrupting controlled ventilation. However, this canister can be easily installed improperly with the CLIC adapter. We report a case in which a patient could not be ventilated by mask after the induction of general anesthesia, resulting in oxygen desaturation before successful ventilation was achieved with a bag valve mask. This case illustrates the importance of a leak test after components of the breathing circuit are changed.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2021 |
---|---|
Erschienen: |
2021 |
Enthalten in: |
Zur Gesamtaufnahme - volume:71 |
---|---|
Enthalten in: |
Brazilian journal of anesthesiology (Elsevier) - 71(2021), 4 vom: 17. Juli, Seite 443-446 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Pai, Sher-Lu [VerfasserIn] |
---|
Links: |
---|
Themen: |
142M471B3J |
---|
Anmerkungen: |
Date Completed 25.10.2021 Date Revised 16.08.2022 published: Print-Electronic Citation Status MEDLINE |
---|
doi: |
10.1016/j.bjane.2021.03.023 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM324834845 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM324834845 | ||
003 | DE-627 | ||
005 | 20231225191145.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231225s2021 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1016/j.bjane.2021.03.023 |2 doi | |
028 | 5 | 2 | |a pubmed24n1082.xml |
035 | |a (DE-627)NLM324834845 | ||
035 | |a (NLM)33930338 | ||
035 | |a (PII)S0104-0014(21)00167-6 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Pai, Sher-Lu |e verfasserin |4 aut | |
245 | 1 | 0 | |a Oxygen delivery failure due to improper installation of carbon dioxide absorbent canister |b a case report |
264 | 1 | |c 2021 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 25.10.2021 | ||
500 | |a Date Revised 16.08.2022 | ||
500 | |a published: Print-Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a Copyright © 2021 Sociedade Brasileira de Anestesiologia. Published by Elsevier Editora Ltda. All rights reserved. | ||
520 | |a The CLIC system in the Dräger Apollo anesthesia workstation allows a successful pre-use machine checkout without the presence of a carbon dioxide absorbent canister. It also allows the canister to be changed without interrupting controlled ventilation. However, this canister can be easily installed improperly with the CLIC adapter. We report a case in which a patient could not be ventilated by mask after the induction of general anesthesia, resulting in oxygen desaturation before successful ventilation was achieved with a bag valve mask. This case illustrates the importance of a leak test after components of the breathing circuit are changed | ||
650 | 4 | |a Case Reports | |
650 | 4 | |a Anesthesiology | |
650 | 4 | |a Carbon dioxide | |
650 | 4 | |a Positive pressure respiration | |
650 | 4 | |a Soda lime | |
650 | 4 | |a Ventilators, mechanical | |
650 | 7 | |a Carbon Dioxide |2 NLM | |
650 | 7 | |a 142M471B3J |2 NLM | |
650 | 7 | |a Oxygen |2 NLM | |
650 | 7 | |a S88TT14065 |2 NLM | |
700 | 1 | |a Robards, Christopher B |e verfasserin |4 aut | |
700 | 1 | |a Riutort, Kevin T |e verfasserin |4 aut | |
700 | 1 | |a Torp, Klaus D |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Brazilian journal of anesthesiology (Elsevier) |d 2013 |g 71(2021), 4 vom: 17. Juli, Seite 443-446 |w (DE-627)NLM225306182 |x 2352-2291 |7 nnns |
773 | 1 | 8 | |g volume:71 |g year:2021 |g number:4 |g day:17 |g month:07 |g pages:443-446 |
856 | 4 | 0 | |u http://dx.doi.org/10.1016/j.bjane.2021.03.023 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 71 |j 2021 |e 4 |b 17 |c 07 |h 443-446 |