Oxygen delivery failure due to improper installation of carbon dioxide absorbent canister : a case report

Copyright © 2021 Sociedade Brasileira de Anestesiologia. Published by Elsevier Editora Ltda. All rights reserved..

The CLIC system in the Dräger Apollo anesthesia workstation allows a successful pre-use machine checkout without the presence of a carbon dioxide absorbent canister. It also allows the canister to be changed without interrupting controlled ventilation. However, this canister can be easily installed improperly with the CLIC adapter. We report a case in which a patient could not be ventilated by mask after the induction of general anesthesia, resulting in oxygen desaturation before successful ventilation was achieved with a bag valve mask. This case illustrates the importance of a leak test after components of the breathing circuit are changed.

Medienart:

E-Artikel

Erscheinungsjahr:

2021

Erschienen:

2021

Enthalten in:

Zur Gesamtaufnahme - volume:71

Enthalten in:

Brazilian journal of anesthesiology (Elsevier) - 71(2021), 4 vom: 17. Juli, Seite 443-446

Sprache:

Englisch

Beteiligte Personen:

Pai, Sher-Lu [VerfasserIn]
Robards, Christopher B [VerfasserIn]
Riutort, Kevin T [VerfasserIn]
Torp, Klaus D [VerfasserIn]

Links:

Volltext

Themen:

142M471B3J
Anesthesiology
Carbon Dioxide
Carbon dioxide
Case Reports
Oxygen
Positive pressure respiration
S88TT14065
Soda lime
Ventilators, mechanical

Anmerkungen:

Date Completed 25.10.2021

Date Revised 16.08.2022

published: Print-Electronic

Citation Status MEDLINE

doi:

10.1016/j.bjane.2021.03.023

funding:

Förderinstitution / Projekttitel:

PPN (Katalog-ID):

NLM324834845