Ulnohumeral joint static cartilage compression is affected by radial head implant size
Copyright © 2021. Published by Elsevier Ltd..
INTRODUCTION: Radiocapitellar joint arthroplasty is a commonly performed procedure, which often leads to early failure or instability. Few studies assess the effect of radiocapitellar joint arthroplasty on the ulnohumeral joint. We hypothesized that static forces of contact (compressing cartilage, or cartilage relaxation contact force) would reveal the effect of varying radial head implant size and elbow position on the ulnohumeral joint.
METHODS: A minimally-invasive method of measuring cartilage relaxation contact force was utilized in 10 fresh-frozen human cadaveric specimens that did not require significant dissection or intraarticular sensor placement. Specimens were rigidly fixed in various positions of elbow flexion and forearm pronosupination with increasing radial head implant lengths. Uniaxial distracting forces were applied and displacement was repeatedly measured with resultant best-fit polynomial curves to determine inflections corresponding to the force required to overcome static cartilage relaxation as in previous work.
FINDINGS: Baseline mean (intra-cadaver) cartilage relaxation contact force was 11.8 N (standard error of the mean = 0.3) at 90° of elbow flexion and neutral rotation. There was little variation within specimens (Intraclass correlation coefficient > 0.94). Cartilage relaxation contact force increased at the ulnohumeral joint with radial head implant overstuffing (> 4 mm, P < 0.05) and elbow flexion (120°, P < 0.001). Pronosupination altered cartilage relaxation contact force in an implant-length independent manner (P < 0.05).
INTERPRETATION: Radiocapitellar joint arthroplasty implant length and elbow joint position independently contribute to increased cartilage relaxation contact force at the ulnohumeral joint. This further supports attempts at anatomic reconstruction of the radiocapitellar joint to prevent pathologic ulnohumeral joint loading.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2023 |
---|---|
Erschienen: |
2023 |
Enthalten in: |
Zur Gesamtaufnahme - volume:109 |
---|---|
Enthalten in: |
Clinical biomechanics (Bristol, Avon) - 109(2023) vom: 17. Okt., Seite 105290 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Elfar, John [VerfasserIn] |
---|
Links: |
---|
Anmerkungen: |
Date Completed 23.10.2023 Date Revised 22.11.2023 published: Print-Electronic Citation Status MEDLINE |
---|
doi: |
10.1016/j.clinbiomech.2021.105290 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM321690672 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM321690672 | ||
003 | DE-627 | ||
005 | 20231225180423.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231225s2023 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1016/j.clinbiomech.2021.105290 |2 doi | |
028 | 5 | 2 | |a pubmed24n1072.xml |
035 | |a (DE-627)NLM321690672 | ||
035 | |a (NLM)33610388 | ||
035 | |a (PII)S0268-0033(21)00020-6 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Elfar, John |e verfasserin |4 aut | |
245 | 1 | 0 | |a Ulnohumeral joint static cartilage compression is affected by radial head implant size |
264 | 1 | |c 2023 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 23.10.2023 | ||
500 | |a Date Revised 22.11.2023 | ||
500 | |a published: Print-Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a Copyright © 2021. Published by Elsevier Ltd. | ||
520 | |a INTRODUCTION: Radiocapitellar joint arthroplasty is a commonly performed procedure, which often leads to early failure or instability. Few studies assess the effect of radiocapitellar joint arthroplasty on the ulnohumeral joint. We hypothesized that static forces of contact (compressing cartilage, or cartilage relaxation contact force) would reveal the effect of varying radial head implant size and elbow position on the ulnohumeral joint | ||
520 | |a METHODS: A minimally-invasive method of measuring cartilage relaxation contact force was utilized in 10 fresh-frozen human cadaveric specimens that did not require significant dissection or intraarticular sensor placement. Specimens were rigidly fixed in various positions of elbow flexion and forearm pronosupination with increasing radial head implant lengths. Uniaxial distracting forces were applied and displacement was repeatedly measured with resultant best-fit polynomial curves to determine inflections corresponding to the force required to overcome static cartilage relaxation as in previous work | ||
520 | |a FINDINGS: Baseline mean (intra-cadaver) cartilage relaxation contact force was 11.8 N (standard error of the mean = 0.3) at 90° of elbow flexion and neutral rotation. There was little variation within specimens (Intraclass correlation coefficient > 0.94). Cartilage relaxation contact force increased at the ulnohumeral joint with radial head implant overstuffing (> 4 mm, P < 0.05) and elbow flexion (120°, P < 0.001). Pronosupination altered cartilage relaxation contact force in an implant-length independent manner (P < 0.05) | ||
520 | |a INTERPRETATION: Radiocapitellar joint arthroplasty implant length and elbow joint position independently contribute to increased cartilage relaxation contact force at the ulnohumeral joint. This further supports attempts at anatomic reconstruction of the radiocapitellar joint to prevent pathologic ulnohumeral joint loading | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a Research Support, N.I.H., Extramural | |
650 | 4 | |a Research Support, Non-U.S. Gov't | |
650 | 4 | |a Research Support, U.S. Gov't, Non-P.H.S. | |
650 | 4 | |a Cartilage relaxation | |
650 | 4 | |a Elbow | |
650 | 4 | |a Joint contact force | |
650 | 4 | |a Joint reaction force | |
650 | 4 | |a Radial head arthroplasty | |
700 | 1 | |a Caldwell, Michael |e verfasserin |4 aut | |
700 | 1 | |a Mason, Robert |e verfasserin |4 aut | |
700 | 1 | |a Schaffer, Joseph |e verfasserin |4 aut | |
700 | 1 | |a Koroneos, Zachary |e verfasserin |4 aut | |
700 | 1 | |a Maqsoodi, Noorullah |e verfasserin |4 aut | |
700 | 1 | |a Osman, Walid S |e verfasserin |4 aut | |
700 | 1 | |a Olles, Mark |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Clinical biomechanics (Bristol, Avon) |d 1986 |g 109(2023) vom: 17. Okt., Seite 105290 |w (DE-627)NLM085859559 |x 1879-1271 |7 nnns |
773 | 1 | 8 | |g volume:109 |g year:2023 |g day:17 |g month:10 |g pages:105290 |
856 | 4 | 0 | |u http://dx.doi.org/10.1016/j.clinbiomech.2021.105290 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 109 |j 2023 |b 17 |c 10 |h 105290 |