Clinical Evaluation of a COVID-19 Antibody Lateral Flow Assay using Point of Care Samples

Introduction: The ongoing SARS-CoV-2 pandemic has spurred the development of numerous point of care (PoC) immunoassays. Assessments of performance of available kits are necessary to determine their clinical utility. Previous studies have mostly performed these assessments in a laboratory setting, which raises concerns of translating findings for PoC use. The aim of this study was to assess the performance of a lateral flow immunoassay for the detection of SARS-CoV-2 antibodies using samples collected at PoC.

Method: One lateral flow immunoassay (Humasis® COVID-19 IgG/IgM) was tested. In total, 50 PCR RT-PCR positive and 52 RT-PCR negative samples were collected at PoC. Fifty serum specimens from Dec 2018 to Feb 2019 were used as controls for specificity. Serum samples collected between Dec 2019 to Feb 2020 were used as additional comparators. Clinical data including symptom onset date was collected from patient history and the medical record.

Results: The overall sensitivity for the kit was 74% (95% CI: 59.7% - 85.4%). The sensitivity for IgM and IgG detection >14 days after date of onset was 88% (95% CI: 68.8% - 97.5%) and 84% (95% CI: 63.9% - 95.5%), with a negative predictive value (NPV) of 94% for IgM (95% CI: 83.5% - 98.8%) and 93% for IgG (95% CI: 81.8% - 97.9%). The overall specificity was 94% (95% CI: 83.5% - 98.8%). The Immunoglobulin specific specificity was 94% for IgM (95% CI: 83.5% - 98.8%) and 98% for IgG (95% CI: 89.4% - 100.0%), with a positive predictive value (PPV) of 88% for IgM (95% CI: 68.8% - 97.5%) and 95% for IgG (95% CI: 77.2% - 99.9%) respectively for samples collected from patients >14 days after date of onset. Specimen collected during early phase of COVID-19 pandemic (Dec 2019 to Feb 2020) showed 11.8% antibody positivity, and 11.3% of PCR-negative patients demonstrated antibody positivity.

Discussion: Humasis® COVID-19 IgG/IgM LFA demonstrates greater than 90% PPV and NPV for samples collected 14 days after the onset of symptoms using samples collected at PoC. While not practical for the diagnosis of acute infection, the use of the lateral flow assays with high specificity may have utility for determining seroprevalence or seroconversion in longitudinal studies.

Medienart:

E-Artikel

Erscheinungsjahr:

2020

Erschienen:

2020

Enthalten in:

Zur Gesamtaufnahme - year:2020

Enthalten in:

medRxiv : the preprint server for health sciences - (2020) vom: 04. Dez.

Sprache:

Englisch

Beteiligte Personen:

Lee, Won [VerfasserIn]
Straube, Steven [VerfasserIn]
Sincic, Ryan [VerfasserIn]
Noble, Jeanne A [VerfasserIn]
Montoy, Juan Carlos [VerfasserIn]
Kornblith, Aaron E [VerfasserIn]
Prakash, Arun [VerfasserIn]
Wang, Ralph [VerfasserIn]
Bainton, Roland J [VerfasserIn]
Kurien, Philip A [VerfasserIn]

Links:

Volltext

Themen:

Preprint

Anmerkungen:

Date Revised 19.10.2023

published: Electronic

Citation Status PubMed-not-MEDLINE

doi:

10.1101/2020.12.02.20242750

funding:

Förderinstitution / Projekttitel:

PPN (Katalog-ID):

NLM318654121