Triple-sensitivity high-spatial-resolution X-ray computed tomography using a cadmium-telluride detector and its beam-hardening effect

Copyright © 2020 Elsevier Ltd. All rights reserved..

To observe blood vessels at high contrasts, we constructed a first-generation triple-sensitivity X-ray computed tomography (TS-CT) scanner using a cadmium-telluride (CdTe) detector and a triple-amplifying system. X-ray photons are absorbed by the CdTe crystal, and the electric charges produced by photons are converted into voltages using a current-to-voltage (I-V) amplifier, and the I-V output is amplified by a voltage-to-voltage (V-V) amplifier. The V-V output 1 is sent to a dual V-V amplifier through a 5.0-m-length coaxial cable and amplified to two-different outputs of 2 and 3. The three outputs 1-3 are sent to a personal computer through an analog-to-digital converter to reconstruct three tomograms simultaneously. In the TS-CT, the scattering photons from the object are extremely reduced using a 0.5-mm-diameter lead pinhole behind the object. The translation and rotation steps were 0.1 mm and 0.5°, respectively, and the spatial resolutions were 0.25 × 0.25 mm2. The scanning time was 19.6 min, and blood vessels were visible using gadolinium contrast media. In particular, the effective photon energy increased with increasing amplification factor of the amplifier caused by beam hardening of the object.

Medienart:

E-Artikel

Erscheinungsjahr:

2020

Erschienen:

2020

Enthalten in:

Zur Gesamtaufnahme - volume:159

Enthalten in:

Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine - 159(2020) vom: 22. Mai, Seite 109089

Sprache:

Englisch

Beteiligte Personen:

Yoshida, Sohei [VerfasserIn]
Sato, Eiichi [VerfasserIn]
Oda, Yasuyuki [VerfasserIn]
Yoshioka, Kunihiro [VerfasserIn]
Moriyama, Hodaka [VerfasserIn]
Watanabe, Manabu [VerfasserIn]

Links:

Volltext

Themen:

Cadmium Compounds
Cadmium telluride
High spatial resolution
Journal Article
Low-noise CdTe detector
NQA0O090ZJ
STG188WO13
Tellurium
Triple amplifier
Triple energy
Triple sensitivity
X-ray CT

Anmerkungen:

Date Completed 08.03.2021

Date Revised 08.03.2021

published: Print-Electronic

Citation Status MEDLINE

doi:

10.1016/j.apradiso.2020.109089

funding:

Förderinstitution / Projekttitel:

PPN (Katalog-ID):

NLM308361318