Triple-sensitivity high-spatial-resolution X-ray computed tomography using a cadmium-telluride detector and its beam-hardening effect
Copyright © 2020 Elsevier Ltd. All rights reserved..
To observe blood vessels at high contrasts, we constructed a first-generation triple-sensitivity X-ray computed tomography (TS-CT) scanner using a cadmium-telluride (CdTe) detector and a triple-amplifying system. X-ray photons are absorbed by the CdTe crystal, and the electric charges produced by photons are converted into voltages using a current-to-voltage (I-V) amplifier, and the I-V output is amplified by a voltage-to-voltage (V-V) amplifier. The V-V output 1 is sent to a dual V-V amplifier through a 5.0-m-length coaxial cable and amplified to two-different outputs of 2 and 3. The three outputs 1-3 are sent to a personal computer through an analog-to-digital converter to reconstruct three tomograms simultaneously. In the TS-CT, the scattering photons from the object are extremely reduced using a 0.5-mm-diameter lead pinhole behind the object. The translation and rotation steps were 0.1 mm and 0.5°, respectively, and the spatial resolutions were 0.25 × 0.25 mm2. The scanning time was 19.6 min, and blood vessels were visible using gadolinium contrast media. In particular, the effective photon energy increased with increasing amplification factor of the amplifier caused by beam hardening of the object.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2020 |
---|---|
Erschienen: |
2020 |
Enthalten in: |
Zur Gesamtaufnahme - volume:159 |
---|---|
Enthalten in: |
Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine - 159(2020) vom: 22. Mai, Seite 109089 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Yoshida, Sohei [VerfasserIn] |
---|
Links: |
---|
Themen: |
Cadmium Compounds |
---|
Anmerkungen: |
Date Completed 08.03.2021 Date Revised 08.03.2021 published: Print-Electronic Citation Status MEDLINE |
---|
doi: |
10.1016/j.apradiso.2020.109089 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM308361318 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM308361318 | ||
003 | DE-627 | ||
005 | 20231225131555.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231225s2020 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1016/j.apradiso.2020.109089 |2 doi | |
028 | 5 | 2 | |a pubmed24n1027.xml |
035 | |a (DE-627)NLM308361318 | ||
035 | |a (NLM)32250761 | ||
035 | |a (PII)S0969-8043(19)31021-8 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Yoshida, Sohei |e verfasserin |4 aut | |
245 | 1 | 0 | |a Triple-sensitivity high-spatial-resolution X-ray computed tomography using a cadmium-telluride detector and its beam-hardening effect |
264 | 1 | |c 2020 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 08.03.2021 | ||
500 | |a Date Revised 08.03.2021 | ||
500 | |a published: Print-Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a Copyright © 2020 Elsevier Ltd. All rights reserved. | ||
520 | |a To observe blood vessels at high contrasts, we constructed a first-generation triple-sensitivity X-ray computed tomography (TS-CT) scanner using a cadmium-telluride (CdTe) detector and a triple-amplifying system. X-ray photons are absorbed by the CdTe crystal, and the electric charges produced by photons are converted into voltages using a current-to-voltage (I-V) amplifier, and the I-V output is amplified by a voltage-to-voltage (V-V) amplifier. The V-V output 1 is sent to a dual V-V amplifier through a 5.0-m-length coaxial cable and amplified to two-different outputs of 2 and 3. The three outputs 1-3 are sent to a personal computer through an analog-to-digital converter to reconstruct three tomograms simultaneously. In the TS-CT, the scattering photons from the object are extremely reduced using a 0.5-mm-diameter lead pinhole behind the object. The translation and rotation steps were 0.1 mm and 0.5°, respectively, and the spatial resolutions were 0.25 × 0.25 mm2. The scanning time was 19.6 min, and blood vessels were visible using gadolinium contrast media. In particular, the effective photon energy increased with increasing amplification factor of the amplifier caused by beam hardening of the object | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a High spatial resolution | |
650 | 4 | |a Low-noise CdTe detector | |
650 | 4 | |a Triple amplifier | |
650 | 4 | |a Triple energy | |
650 | 4 | |a Triple sensitivity | |
650 | 4 | |a X-ray CT | |
650 | 7 | |a Cadmium Compounds |2 NLM | |
650 | 7 | |a Tellurium |2 NLM | |
650 | 7 | |a NQA0O090ZJ |2 NLM | |
650 | 7 | |a cadmium telluride |2 NLM | |
650 | 7 | |a STG188WO13 |2 NLM | |
700 | 1 | |a Sato, Eiichi |e verfasserin |4 aut | |
700 | 1 | |a Oda, Yasuyuki |e verfasserin |4 aut | |
700 | 1 | |a Yoshioka, Kunihiro |e verfasserin |4 aut | |
700 | 1 | |a Moriyama, Hodaka |e verfasserin |4 aut | |
700 | 1 | |a Watanabe, Manabu |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine |d 1994 |g 159(2020) vom: 22. Mai, Seite 109089 |w (DE-627)NLM074730231 |x 1872-9800 |7 nnns |
773 | 1 | 8 | |g volume:159 |g year:2020 |g day:22 |g month:05 |g pages:109089 |
856 | 4 | 0 | |u http://dx.doi.org/10.1016/j.apradiso.2020.109089 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 159 |j 2020 |b 22 |c 05 |h 109089 |