Temporal registration : a new approach to manage the incomplete recovery of the longitudinal magnetization in the Modified Look-Locker Inversion Recovery sequence (MOLLI) for T1 mapping of the heart

PURPOSE: To correct with post-processing effects of incomplete recovery of the longitudinal magnetization before a new inversion pulse in the Modified Look-Locker Inversion recovery sequence (MOLLI) sequence.

THEORY AND METHODS: We model such effects as a temporal shift ([Formula: see text]) of the signal of the Look-Locker block following next inversion pulses. After using the following equation [Formula: see text], a temporal registration of [Formula: see text] is applied to the signal of the affected block to adjust the sampling time of the recovery signal and correct the underlying effect on quantitative T1. To test our approach, simulations, phantoms, and five volunteers' data were used while applying different MOLLI sampling schemes at different heart rates and compared to the reference three-parameter fit.

RESULTS: The temporal registration of the affected signals allows to reach higher accuracy on long T1 when compared to the reference three parameters fit (10.15 vs 22.12% for T1 = 1785 ms; 8.22 vs 14.65% for T1 = 1278 ms), and lower average variation in case of rest-period deletion (62 vs 231 ms).

CONCLUSION: The proposed approach leads to more accurate T1 in case of incomplete recovery. It is less sensitive to parameters affecting the recovery such as the rest period or the sampling scheme; and, therefore, supports multi-center studies with different MOLLI protocols.

Medienart:

E-Artikel

Erscheinungsjahr:

2020

Erschienen:

2020

Enthalten in:

Zur Gesamtaufnahme - volume:33

Enthalten in:

Magma (New York, N.Y.) - 33(2020), 4 vom: 08. Aug., Seite 569-580

Sprache:

Englisch

Beteiligte Personen:

Rebbah, Habib [VerfasserIn]
Galas, Thierry [VerfasserIn]
Soulat, Gilles [VerfasserIn]
Kachenoura, Nadjia [VerfasserIn]
Menini, Anne [VerfasserIn]
Cuenod, Charles A [VerfasserIn]
Mousseaux, Elie [VerfasserIn]

Links:

Volltext

Themen:

Accuracy
Cardiac MRI
Contrast Media
Journal Article
Myocardial T1 mapping
Reproducibility

Anmerkungen:

Date Completed 15.06.2021

Date Revised 15.04.2022

published: Print-Electronic

Citation Status MEDLINE

doi:

10.1007/s10334-019-00815-6

funding:

Förderinstitution / Projekttitel:

PPN (Katalog-ID):

NLM305152890