Technical Note : Investigation of the dosimetric impact of stray radiation on the Common Control Unit of the IBA Blue Phantom2

© 2019 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine..

PURPOSE: This technical note aims to investigate the dosimetric impact of stray radiation on the Common Control Unit (CCU) of the IBA Blue Phantom2 and the measured beam data.

METHODS: Three CCUs of the same model were used for the study. The primary test CCU was placed at five distances from the radiation beam central axis. At each distance, a set of depth dose and beam profiles for two open and two wedge fields were measured. The field sizes were 10 × 10 cm2 and 30 × 30 cm2 for the open fields, and 30 × 30 cm2 and 15 × 15 cm2 for the 30° and 60° wedges, respectively. The other two CCUs were used to cross check the data of the primary CCU. Assuming the effect of stray radiation on the data measured at the farthest reachable distance 4.5 m is negligible, the dosimetric impact of stray radiation on the CCU and consequently on the measured data can be extracted for analysis by comparing it with those measured at shorter distances.

RESULTS: The results of three CCUs were consistent. The dosimetric impact of stray radiation was greater for lower energies at larger field sizes. For open fields, the data variation was up to 4.5% for depth dose curves and 7.1% for beam profiles. For wedge fields, the data variation was up to 9.3% for depth dose curves and 10.6% for beam profiles. Moreover, for wedge field profiles in the wedge direction, they became flatter as the CCU was placed closer to the primary radiation beam, manifesting smaller wedge angles.

CONCLUSION: The stray radiation added a uniform background noise on all measured data. The magnitude of the noise is inversely proportional to the square of the distance of the CCU to the primary radiation beam, approximately following the inverse square law.

Medienart:

E-Artikel

Erscheinungsjahr:

2020

Erschienen:

2020

Enthalten in:

Zur Gesamtaufnahme - volume:21

Enthalten in:

Journal of applied clinical medical physics - 21(2020), 1 vom: 07. Jan., Seite 191-196

Sprache:

Englisch

Beteiligte Personen:

Cui, Guoqiang [VerfasserIn]
Duan, Jun [VerfasserIn]
Yang, Yun [VerfasserIn]
Yin, Fang-Fang [VerfasserIn]

Links:

Volltext

Themen:

Common Control Unit (CCU)
IBA Blue Phantom2
Inverse square law
Journal Article
Stray radiation

Anmerkungen:

Date Completed 26.10.2020

Date Revised 27.03.2024

published: Print-Electronic

Citation Status MEDLINE

doi:

10.1002/acm2.12769

funding:

Förderinstitution / Projekttitel:

PPN (Katalog-ID):

NLM30333066X