Validation of methods for determining pediatric midazolam using wet whole blood and volumetric absorptive microsampling

Aim: Collection and quantitative analysis in dry blood using volumetric absorptive microsampling (VAMS™) potentially offers significant advantages over conventional wet whole blood analysis. This manuscript explores their use for pediatric sampling and explores additional considerations for the validation of the bioanalytical method. Results: HPLC-MS/MS methods for the determination of midazolam and its major metabolite 1-OH midazolam in both whole wet blood, and dry blood collected on VAMS were developed, validated, and used to support an observational clinical study to compare pharmacokinetic parameters in pediatric patients. Conclusion: Validation data met internationally accepted guideline criteria. A strong correlation was observed in calculated concentrations between wet and dry test samples, indicating that VAMS is a suitable technique for use in pediatric clinical studies.

Medienart:

E-Artikel

Erscheinungsjahr:

2019

Erschienen:

2019

Enthalten in:

Zur Gesamtaufnahme - volume:11

Enthalten in:

Bioanalysis - 11(2019), 19 vom: 16. Okt., Seite 1737-1754

Sprache:

Englisch

Beteiligte Personen:

Abu-Rabie, Paul [VerfasserIn]
Neupane, Bikalpa [VerfasserIn]
Spooner, Neil [VerfasserIn]
Rudge, James [VerfasserIn]
Denniff, Philip [VerfasserIn]
Mulla, Hussain [VerfasserIn]
Pandya, Hitesh [VerfasserIn]

Links:

Volltext

Themen:

Blood
Hematocrit
Hypnotics and Sedatives
Journal Article
LC–MS
Microsampling
Midazolam
Mitra® microsamplers
Pediatric
Pharmacokinetics
R60L0SM5BC
Validation Study
Volumetric absorptive sampling (VAMS)
Wet vs dry

Anmerkungen:

Date Completed 25.03.2020

Date Revised 25.03.2020

published: Print-Electronic

Citation Status MEDLINE

doi:

10.4155/bio-2019-0190

funding:

Förderinstitution / Projekttitel:

PPN (Katalog-ID):

NLM30223862X