How Do Contrast Agents Affect Cardiac Markers and Coagulation Tests? Experimental Study
Copyright© Bentham Science Publishers; For any queries, please email at epubbenthamscience.net..
BACKGROUND: The discovery that biotin interferes with results of troponin and Nt-proBNP led some commercial firms to update their measurement methods. In particular, the clinical incompatibility of cardiac test results may affect the risk of morbidity and mortality.
OBJECTIVE: The aim of this study is to investigate the interference effects of 7 different contrast agents on cardiac markers (Troponin-I, Nt-proBNP, Mass CK-MB, CK, AST, LDH) and in coagulation tests (PT, APTT).
METHODS: Seven different contrast media were added into control materials by using interference protocol. The concentrations of PT, APTT, CK, AST, LDH, Mass CK-MB, Troponin-I, and Nt-proBNP were measured by Sysmex CS-2100, Abbott c16000, Siemens Centaur XP and AFİAS-6 analyzer. The number of deviations from target values was calculated.
RESULTS: The 7 different contrast media caused negative interference in troponin levels between 57.43% and 62.87%. It was found that different contrast media produced false negativity in the Nt-proBNP test, ranging from 6.11% to 96.01%. Enzymes and coagulation tests were less affected.
CONCLUSION: Different contrast media may cause false negatives in cTnI and Nt-proBNP. The contrast medium that causes the least interference should be preferred. The results of samples taken in the first hour after contrast imaging should be interpreted with care.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2019 |
---|---|
Erschienen: |
2019 |
Enthalten in: |
Zur Gesamtaufnahme - volume:22 |
---|---|
Enthalten in: |
Combinatorial chemistry & high throughput screening - 22(2019), 5 vom: 08. Aug., Seite 355-360 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Tascanov, Mustafa Begenc [VerfasserIn] |
---|
Links: |
---|
Anmerkungen: |
Date Completed 20.07.2020 Date Revised 20.07.2020 published: Print Citation Status MEDLINE |
---|
doi: |
10.2174/1386207322666190603170438 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM297794507 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM297794507 | ||
003 | DE-627 | ||
005 | 20231225092752.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231225s2019 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.2174/1386207322666190603170438 |2 doi | |
028 | 5 | 2 | |a pubmed24n0992.xml |
035 | |a (DE-627)NLM297794507 | ||
035 | |a (NLM)31161978 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Tascanov, Mustafa Begenc |e verfasserin |4 aut | |
245 | 1 | 0 | |a How Do Contrast Agents Affect Cardiac Markers and Coagulation Tests? Experimental Study |
264 | 1 | |c 2019 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 20.07.2020 | ||
500 | |a Date Revised 20.07.2020 | ||
500 | |a published: Print | ||
500 | |a Citation Status MEDLINE | ||
520 | |a Copyright© Bentham Science Publishers; For any queries, please email at epubbenthamscience.net. | ||
520 | |a BACKGROUND: The discovery that biotin interferes with results of troponin and Nt-proBNP led some commercial firms to update their measurement methods. In particular, the clinical incompatibility of cardiac test results may affect the risk of morbidity and mortality | ||
520 | |a OBJECTIVE: The aim of this study is to investigate the interference effects of 7 different contrast agents on cardiac markers (Troponin-I, Nt-proBNP, Mass CK-MB, CK, AST, LDH) and in coagulation tests (PT, APTT) | ||
520 | |a METHODS: Seven different contrast media were added into control materials by using interference protocol. The concentrations of PT, APTT, CK, AST, LDH, Mass CK-MB, Troponin-I, and Nt-proBNP were measured by Sysmex CS-2100, Abbott c16000, Siemens Centaur XP and AFİAS-6 analyzer. The number of deviations from target values was calculated | ||
520 | |a RESULTS: The 7 different contrast media caused negative interference in troponin levels between 57.43% and 62.87%. It was found that different contrast media produced false negativity in the Nt-proBNP test, ranging from 6.11% to 96.01%. Enzymes and coagulation tests were less affected | ||
520 | |a CONCLUSION: Different contrast media may cause false negatives in cTnI and Nt-proBNP. The contrast medium that causes the least interference should be preferred. The results of samples taken in the first hour after contrast imaging should be interpreted with care | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a Contrast media | |
650 | 4 | |a Nt-proBNP | |
650 | 4 | |a cardiac markers | |
650 | 4 | |a coagulation test | |
650 | 4 | |a interference | |
650 | 4 | |a troponin. | |
650 | 7 | |a Biomarkers |2 NLM | |
650 | 7 | |a Contrast Media |2 NLM | |
650 | 7 | |a Peptide Fragments |2 NLM | |
650 | 7 | |a Troponin I |2 NLM | |
650 | 7 | |a pro-brain natriuretic peptide (1-76) |2 NLM | |
650 | 7 | |a Natriuretic Peptide, Brain |2 NLM | |
650 | 7 | |a 114471-18-0 |2 NLM | |
700 | 1 | |a Gönel, Ataman |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Combinatorial chemistry & high throughput screening |d 1998 |g 22(2019), 5 vom: 08. Aug., Seite 355-360 |w (DE-627)NLM097489352 |x 1875-5402 |7 nnns |
773 | 1 | 8 | |g volume:22 |g year:2019 |g number:5 |g day:08 |g month:08 |g pages:355-360 |
856 | 4 | 0 | |u http://dx.doi.org/10.2174/1386207322666190603170438 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 22 |j 2019 |e 5 |b 08 |c 08 |h 355-360 |