Injection Pressure Evaluation of the New Venous Catheter with Side Holes for Contrast-enhanced CT/MRI
The simulation study was conducted for the new venous catheter with side holes of contrast-enhanced computed tomography (CT) and magnetic resonance imaging (MRI) to evaluate the infusion pressure on four contrast media and several injection speeds. All infusion pressure of the new venous catheter with side holes were less than 15 kg/cm2 as limitation of extension tube and also reduced the infusion pressure by 15% at the maximum compared to the catheter with single hole. The results suggest that the new venous catheter with side holes can reduce the infusion pressure by power injection of contrast-enhanced CT and MRI.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2018 |
---|---|
Erschienen: |
2018 |
Enthalten in: |
Zur Gesamtaufnahme - volume:74 |
---|---|
Enthalten in: |
Nihon Hoshasen Gijutsu Gakkai zasshi - 74(2018), 4 vom: 20., Seite 362-367 |
Sprache: |
Japanisch |
---|
Beteiligte Personen: |
Fukuda, Junya [VerfasserIn] |
---|
Links: |
---|
Themen: |
Closed intravenous catheter |
---|
Anmerkungen: |
Date Completed 01.03.2019 Date Revised 10.12.2019 published: Print Citation Status MEDLINE |
---|
doi: |
10.6009/jjrt.2018_JSRT_74.4.362 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM283295236 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM283295236 | ||
003 | DE-627 | ||
005 | 20231225040327.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231225s2018 xx |||||o 00| ||jpn c | ||
024 | 7 | |a 10.6009/jjrt.2018_JSRT_74.4.362 |2 doi | |
028 | 5 | 2 | |a pubmed24n0944.xml |
035 | |a (DE-627)NLM283295236 | ||
035 | |a (NLM)29681603 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a jpn | ||
100 | 1 | |a Fukuda, Junya |e verfasserin |4 aut | |
245 | 1 | 0 | |a Injection Pressure Evaluation of the New Venous Catheter with Side Holes for Contrast-enhanced CT/MRI |
264 | 1 | |c 2018 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 01.03.2019 | ||
500 | |a Date Revised 10.12.2019 | ||
500 | |a published: Print | ||
500 | |a Citation Status MEDLINE | ||
520 | |a The simulation study was conducted for the new venous catheter with side holes of contrast-enhanced computed tomography (CT) and magnetic resonance imaging (MRI) to evaluate the infusion pressure on four contrast media and several injection speeds. All infusion pressure of the new venous catheter with side holes were less than 15 kg/cm2 as limitation of extension tube and also reduced the infusion pressure by 15% at the maximum compared to the catheter with single hole. The results suggest that the new venous catheter with side holes can reduce the infusion pressure by power injection of contrast-enhanced CT and MRI | ||
650 | 4 | |a Evaluation Study | |
650 | 4 | |a Journal Article | |
650 | 4 | |a closed intravenous catheter | |
650 | 4 | |a in-vitro study | |
650 | 4 | |a infusion pressure | |
650 | 4 | |a power injection | |
650 | 4 | |a side holes | |
650 | 7 | |a Contrast Media |2 NLM | |
700 | 1 | |a Arai, Keisuke |e verfasserin |4 aut | |
700 | 1 | |a Miyazawa, Hitomi |e verfasserin |4 aut | |
700 | 1 | |a Kobayashi, Kyouko |e verfasserin |4 aut | |
700 | 1 | |a Nakamura, Junpei |e verfasserin |4 aut | |
700 | 1 | |a Suto, Takayuki |e verfasserin |4 aut | |
700 | 1 | |a Tsushima, Yoshito |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Nihon Hoshasen Gijutsu Gakkai zasshi |d 1993 |g 74(2018), 4 vom: 20., Seite 362-367 |w (DE-627)NLM074820923 |x 1881-4883 |7 nnns |
773 | 1 | 8 | |g volume:74 |g year:2018 |g number:4 |g day:20 |g pages:362-367 |
856 | 4 | 0 | |u http://dx.doi.org/10.6009/jjrt.2018_JSRT_74.4.362 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 74 |j 2018 |e 4 |b 20 |h 362-367 |