Injection Pressure Evaluation of the New Venous Catheter with Side Holes for Contrast-enhanced CT/MRI

The simulation study was conducted for the new venous catheter with side holes of contrast-enhanced computed tomography (CT) and magnetic resonance imaging (MRI) to evaluate the infusion pressure on four contrast media and several injection speeds. All infusion pressure of the new venous catheter with side holes were less than 15 kg/cm2 as limitation of extension tube and also reduced the infusion pressure by 15% at the maximum compared to the catheter with single hole. The results suggest that the new venous catheter with side holes can reduce the infusion pressure by power injection of contrast-enhanced CT and MRI.

Medienart:

E-Artikel

Erscheinungsjahr:

2018

Erschienen:

2018

Enthalten in:

Zur Gesamtaufnahme - volume:74

Enthalten in:

Nihon Hoshasen Gijutsu Gakkai zasshi - 74(2018), 4 vom: 20., Seite 362-367

Sprache:

Japanisch

Beteiligte Personen:

Fukuda, Junya [VerfasserIn]
Arai, Keisuke [VerfasserIn]
Miyazawa, Hitomi [VerfasserIn]
Kobayashi, Kyouko [VerfasserIn]
Nakamura, Junpei [VerfasserIn]
Suto, Takayuki [VerfasserIn]
Tsushima, Yoshito [VerfasserIn]

Links:

Volltext

Themen:

Closed intravenous catheter
Contrast Media
Evaluation Study
In-vitro study
Infusion pressure
Journal Article
Power injection
Side holes

Anmerkungen:

Date Completed 01.03.2019

Date Revised 10.12.2019

published: Print

Citation Status MEDLINE

doi:

10.6009/jjrt.2018_JSRT_74.4.362

funding:

Förderinstitution / Projekttitel:

PPN (Katalog-ID):

NLM283295236