Diagnostic performance of a high-sensitive troponin T assay and a troponin T point of care assay in the clinical routine of an Emergency Department : A clinical cohort study
Copyright © 2016 Elsevier Ireland Ltd. All rights reserved..
BACKGROUND: A point of care test (POCT) for troponin T (TnT) in the Emergency Department (ED) was compared to a high-sensitivity TnT (hsTnT) central laboratory test (CLT) to determine the influence of test system and different cut-off values on the diagnostic performance in patients with suspected acute coronary syndrome (ACS) under routine conditions.
METHODS: All patients with routine TnT testing in the ED were enrolled. Only internal medicine patients without STEMI and with both troponin values were analyzed. TnT was measured with a contemporary sensitive POCT assay in the ED and with a hs-assay in the central laboratory. The diagnostic performance was analyzed at two different cut-off points (99th percentile and conventional rule-in cut-offs). Primary endpoint was the diagnosis of NSTEMI.
RESULTS: Of all patients (n=3423), 3.6% had a diagnosis of NSTEMI (n=124). For the hsTnT assay, 28.4% of all values were at or below the lower limit of detection (LOD) as compared to 75.7% of the POC-TnT-values. The area under the receiver operating curves did not differ significantly between the assays (hsTnT: 0.912(95%-CI: 0.884-0.940); POC-TnT: 0.896(95%-CI: 0.859-0.933)). The diagnostic performance was very similar for both assays: the positive predictive value was below 50% for troponin values below 100ng/L and hardly increased for values between 100 and 600ng/L for hs and conventional assays.
CONCLUSIONS: In our cohort of emergency patients, the diagnostic performance of conventional POC-testing was comparable to hsTnT. A 99th percentile cut-off may be useful for rule-out of NSTEMI, but seems limited for routine rule-in strategies.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2017 |
---|---|
Erschienen: |
2017 |
Enthalten in: |
Zur Gesamtaufnahme - volume:230 |
---|---|
Enthalten in: |
International journal of cardiology - 230(2017) vom: 01. März, Seite 454-460 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Slagman, Anna [VerfasserIn] |
---|
Links: |
---|
Themen: |
Acute coronary syndrome |
---|
Anmerkungen: |
Date Completed 27.11.2017 Date Revised 28.11.2017 published: Print-Electronic Citation Status MEDLINE |
---|
doi: |
10.1016/j.ijcard.2016.12.085 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM267600577 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM267600577 | ||
003 | DE-627 | ||
005 | 20231224221036.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231224s2017 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1016/j.ijcard.2016.12.085 |2 doi | |
028 | 5 | 2 | |a pubmed24n0892.xml |
035 | |a (DE-627)NLM267600577 | ||
035 | |a (NLM)28040285 | ||
035 | |a (PII)S0167-5273(16)34576-4 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Slagman, Anna |e verfasserin |4 aut | |
245 | 1 | 0 | |a Diagnostic performance of a high-sensitive troponin T assay and a troponin T point of care assay in the clinical routine of an Emergency Department |b A clinical cohort study |
264 | 1 | |c 2017 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 27.11.2017 | ||
500 | |a Date Revised 28.11.2017 | ||
500 | |a published: Print-Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a Copyright © 2016 Elsevier Ireland Ltd. All rights reserved. | ||
520 | |a BACKGROUND: A point of care test (POCT) for troponin T (TnT) in the Emergency Department (ED) was compared to a high-sensitivity TnT (hsTnT) central laboratory test (CLT) to determine the influence of test system and different cut-off values on the diagnostic performance in patients with suspected acute coronary syndrome (ACS) under routine conditions | ||
520 | |a METHODS: All patients with routine TnT testing in the ED were enrolled. Only internal medicine patients without STEMI and with both troponin values were analyzed. TnT was measured with a contemporary sensitive POCT assay in the ED and with a hs-assay in the central laboratory. The diagnostic performance was analyzed at two different cut-off points (99th percentile and conventional rule-in cut-offs). Primary endpoint was the diagnosis of NSTEMI | ||
520 | |a RESULTS: Of all patients (n=3423), 3.6% had a diagnosis of NSTEMI (n=124). For the hsTnT assay, 28.4% of all values were at or below the lower limit of detection (LOD) as compared to 75.7% of the POC-TnT-values. The area under the receiver operating curves did not differ significantly between the assays (hsTnT: 0.912(95%-CI: 0.884-0.940); POC-TnT: 0.896(95%-CI: 0.859-0.933)). The diagnostic performance was very similar for both assays: the positive predictive value was below 50% for troponin values below 100ng/L and hardly increased for values between 100 and 600ng/L for hs and conventional assays | ||
520 | |a CONCLUSIONS: In our cohort of emergency patients, the diagnostic performance of conventional POC-testing was comparable to hsTnT. A 99th percentile cut-off may be useful for rule-out of NSTEMI, but seems limited for routine rule-in strategies | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a Acute coronary syndrome | |
650 | 4 | |a Diagnostic performance | |
650 | 4 | |a High sensitive troponin | |
650 | 4 | |a Non-ST elevation myocardial infarction | |
650 | 4 | |a Point of care testing | |
650 | 7 | |a Troponin T |2 NLM | |
700 | 1 | |a von Recum, Johannes |e verfasserin |4 aut | |
700 | 1 | |a Möckel, Martin |e verfasserin |4 aut | |
700 | 1 | |a Holert, Fabian |e verfasserin |4 aut | |
700 | 1 | |a Meyer Zum Büschenfelde, Dirk |e verfasserin |4 aut | |
700 | 1 | |a Müller, Christian |e verfasserin |4 aut | |
700 | 1 | |a Searle, Julia |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t International journal of cardiology |d 1984 |g 230(2017) vom: 01. März, Seite 454-460 |w (DE-627)NLM012621196 |x 1874-1754 |7 nnns |
773 | 1 | 8 | |g volume:230 |g year:2017 |g day:01 |g month:03 |g pages:454-460 |
856 | 4 | 0 | |u http://dx.doi.org/10.1016/j.ijcard.2016.12.085 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 230 |j 2017 |b 01 |c 03 |h 454-460 |