Prosthetic Aortic Valve Fixation Study : 48 Replacement Valves Analyzed Using Digital Pressure Mapping
OBJECTIVE: Prostheses attachment is critical in aortic valve replacement surgery, yet reliable prosthetic security remains a challenge. Accurate techniques to analyze prosthetic fixation pressures may enable the use of fewer sutures while reducing the risk of paravalvular leaks (PVL).
METHODS: Customized digital thin film pressure transducers were sutured between aortic annulus models and 21-mm bioprosthetic valves with 15 × 4-mm, 12 × 4-mm, or 9 × 6-mm-wide pledgeted mattress sutures. Simulating open and minimally invasive access, 4 surgeons, blinded to data acquisition, each secured 12 valves using manual knot-tying (hand-tied [HT] or knot-pusher [KP]) or automated titanium fasteners (TFs). Real-time pressure measurements and times were recorded. Two-dimensional (2D) and 3D pressure maps were generated for all valves. Pressures less than 80 mm Hg were considered at risk for PVL.
RESULTS: Pressures under each knot (intrasuture) fell less than 80 mm Hg for 12 of 144 manual knots (5/144 HT, 7/144 KP) versus 0 of 288 TF (P < 0.001). Pressures outside adjacent sutures (extrasuture) were less than 80 mm Hg in 10 of 60 HT, zero of 60 KP, and zero of 120 TF sites for 15 × 4-mm valves; 17 of 48 HT, 25 of 48 KP, and 12 of 96 TF for 12 × 4-mm valves; and 15 of 36 HT, 17 of 36 KP, and 9 and 72 TF for 9 × 6-mm valves; P < 0.001 all manual versus TF. Annular areas with pressures less than 80 mm Hg ranged from 0% of the sewing-ring area (all open TF) to 31% (12 × 4 mm, KP). The average time per manual knot, 46 seconds (HT, 31 seconds; KP, 61 seconds), was greater than TF, 14 seconds (P < 0.005).
CONCLUSIONS: Reduced operative times and PVL risk would fortify the advantages of surgical aortic valve replacement. This research encourages continued exploration of technical factors in optimizing prosthetic valve security.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2016 |
---|---|
Erschienen: |
2016 |
Enthalten in: |
Zur Gesamtaufnahme - volume:11 |
---|---|
Enthalten in: |
Innovations (Philadelphia, Pa.) - 11(2016), 5 vom: 06. Sept., Seite 327-336 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Lee, Candice Y [VerfasserIn] |
---|
Themen: |
---|
Anmerkungen: |
Date Completed 22.05.2017 Date Revised 22.05.2017 published: Print Citation Status MEDLINE |
---|
Förderinstitution / Projekttitel: |
|
---|
PPN (Katalog-ID): |
NLM263754936 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM263754936 | ||
003 | DE-627 | ||
005 | 20231224204606.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231224s2016 xx |||||o 00| ||eng c | ||
028 | 5 | 2 | |a pubmed24n0879.xml |
035 | |a (DE-627)NLM263754936 | ||
035 | |a (NLM)27561176 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Lee, Candice Y |e verfasserin |4 aut | |
245 | 1 | 0 | |a Prosthetic Aortic Valve Fixation Study |b 48 Replacement Valves Analyzed Using Digital Pressure Mapping |
264 | 1 | |c 2016 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 22.05.2017 | ||
500 | |a Date Revised 22.05.2017 | ||
500 | |a published: Print | ||
500 | |a Citation Status MEDLINE | ||
520 | |a OBJECTIVE: Prostheses attachment is critical in aortic valve replacement surgery, yet reliable prosthetic security remains a challenge. Accurate techniques to analyze prosthetic fixation pressures may enable the use of fewer sutures while reducing the risk of paravalvular leaks (PVL) | ||
520 | |a METHODS: Customized digital thin film pressure transducers were sutured between aortic annulus models and 21-mm bioprosthetic valves with 15 × 4-mm, 12 × 4-mm, or 9 × 6-mm-wide pledgeted mattress sutures. Simulating open and minimally invasive access, 4 surgeons, blinded to data acquisition, each secured 12 valves using manual knot-tying (hand-tied [HT] or knot-pusher [KP]) or automated titanium fasteners (TFs). Real-time pressure measurements and times were recorded. Two-dimensional (2D) and 3D pressure maps were generated for all valves. Pressures less than 80 mm Hg were considered at risk for PVL | ||
520 | |a RESULTS: Pressures under each knot (intrasuture) fell less than 80 mm Hg for 12 of 144 manual knots (5/144 HT, 7/144 KP) versus 0 of 288 TF (P < 0.001). Pressures outside adjacent sutures (extrasuture) were less than 80 mm Hg in 10 of 60 HT, zero of 60 KP, and zero of 120 TF sites for 15 × 4-mm valves; 17 of 48 HT, 25 of 48 KP, and 12 of 96 TF for 12 × 4-mm valves; and 15 of 36 HT, 17 of 36 KP, and 9 and 72 TF for 9 × 6-mm valves; P < 0.001 all manual versus TF. Annular areas with pressures less than 80 mm Hg ranged from 0% of the sewing-ring area (all open TF) to 31% (12 × 4 mm, KP). The average time per manual knot, 46 seconds (HT, 31 seconds; KP, 61 seconds), was greater than TF, 14 seconds (P < 0.005) | ||
520 | |a CONCLUSIONS: Reduced operative times and PVL risk would fortify the advantages of surgical aortic valve replacement. This research encourages continued exploration of technical factors in optimizing prosthetic valve security | ||
650 | 4 | |a Journal Article | |
700 | 1 | |a Wong, Joshua K |e verfasserin |4 aut | |
700 | 1 | |a Ross, Ronald E |e verfasserin |4 aut | |
700 | 1 | |a Liu, David C |e verfasserin |4 aut | |
700 | 1 | |a Khabbaz, Kamal R |e verfasserin |4 aut | |
700 | 1 | |a Martellaro, Angelo J |e verfasserin |4 aut | |
700 | 1 | |a Gorea, Heather R |e verfasserin |4 aut | |
700 | 1 | |a Sauer, Jude S |e verfasserin |4 aut | |
700 | 1 | |a Knight, Peter A |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Innovations (Philadelphia, Pa.) |d 2005 |g 11(2016), 5 vom: 06. Sept., Seite 327-336 |w (DE-627)NLM174002238 |x 1559-0879 |7 nnns |
773 | 1 | 8 | |g volume:11 |g year:2016 |g number:5 |g day:06 |g month:09 |g pages:327-336 |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 11 |j 2016 |e 5 |b 06 |c 09 |h 327-336 |