Renal Vascular Clamp Placement : A Potential Cause of Incomplete Hilar Control during Partial Nephrectomy
Copyright © 2016 American Urological Association Education and Research, Inc. Published by Elsevier Inc. All rights reserved..
PURPOSE: Previous benchtop studies have shown that robotic bulldog clamps provide incomplete vascular control of a Penrose drain. We determined the efficacy of robotic and laparoscopic bulldog clamps to ensure hemostasis on the human renal artery. The effect of clamp position on vascular control was also examined.
MATERIALS AND METHODS: Fresh human cadaveric renal arteries were used to determine the leak point pressure of 7 bulldog clamps from a total of 3 manufacturers. Five trials were performed per clamp at 4 locations, including the fulcrum, proximal, middle and distal positions. Comparison was done using the Kruskal-Wallis test with p <0.05 considered significant.
RESULTS: None of the bulldog clamps leaked at a pressure less than 215 mm Hg when applied at the proximal, middle or distal position. In general leak point pressure decreased as the artery was positioned more distal along the clamp. The exception was when the vessel was placed at the fulcrum position. At that position 80% to 100% of trials with the Klein laparoscopic, 100% with the Klein robotic (Klein Robotic, San Antonio, Texas) and 60% to 80% with the Scanlan robotic (Scanlan International, Saint Paul, Minnesota) clamp leaked at pressure below 215 mm Hg.
CONCLUSIONS: Each vascular clamp adequately occluded flow at physiological pressure when placed at the proximal, middle or distal position. Furthermore, these results demonstrate that there is leakage at physiological pressure when the artery is placed at the fulcrum of certain clamp types. These results suggest that applying a bulldog clamp at the fulcrum could potentially lead to inadequate vessel occlusion and intraoperative bleeding.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2016 |
---|---|
Erschienen: |
2016 |
Enthalten in: |
Zur Gesamtaufnahme - volume:195 |
---|---|
Enthalten in: |
The Journal of urology - 195(2016), 3 vom: 01. März, Seite 756-62 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Tryon, David [VerfasserIn] |
---|
Links: |
---|
Themen: |
Hemostasis |
---|
Anmerkungen: |
Date Completed 27.06.2016 Date Revised 18.02.2016 published: Print-Electronic Citation Status MEDLINE |
---|
doi: |
10.1016/j.juro.2015.09.080 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM253208807 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM253208807 | ||
003 | DE-627 | ||
005 | 20231224165737.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231224s2016 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1016/j.juro.2015.09.080 |2 doi | |
028 | 5 | 2 | |a pubmed24n0844.xml |
035 | |a (DE-627)NLM253208807 | ||
035 | |a (NLM)26417645 | ||
035 | |a (PII)S0022-5347(15)04906-X | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Tryon, David |e verfasserin |4 aut | |
245 | 1 | 0 | |a Renal Vascular Clamp Placement |b A Potential Cause of Incomplete Hilar Control during Partial Nephrectomy |
264 | 1 | |c 2016 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 27.06.2016 | ||
500 | |a Date Revised 18.02.2016 | ||
500 | |a published: Print-Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a Copyright © 2016 American Urological Association Education and Research, Inc. Published by Elsevier Inc. All rights reserved. | ||
520 | |a PURPOSE: Previous benchtop studies have shown that robotic bulldog clamps provide incomplete vascular control of a Penrose drain. We determined the efficacy of robotic and laparoscopic bulldog clamps to ensure hemostasis on the human renal artery. The effect of clamp position on vascular control was also examined | ||
520 | |a MATERIALS AND METHODS: Fresh human cadaveric renal arteries were used to determine the leak point pressure of 7 bulldog clamps from a total of 3 manufacturers. Five trials were performed per clamp at 4 locations, including the fulcrum, proximal, middle and distal positions. Comparison was done using the Kruskal-Wallis test with p <0.05 considered significant | ||
520 | |a RESULTS: None of the bulldog clamps leaked at a pressure less than 215 mm Hg when applied at the proximal, middle or distal position. In general leak point pressure decreased as the artery was positioned more distal along the clamp. The exception was when the vessel was placed at the fulcrum position. At that position 80% to 100% of trials with the Klein laparoscopic, 100% with the Klein robotic (Klein Robotic, San Antonio, Texas) and 60% to 80% with the Scanlan robotic (Scanlan International, Saint Paul, Minnesota) clamp leaked at pressure below 215 mm Hg | ||
520 | |a CONCLUSIONS: Each vascular clamp adequately occluded flow at physiological pressure when placed at the proximal, middle or distal position. Furthermore, these results demonstrate that there is leakage at physiological pressure when the artery is placed at the fulcrum of certain clamp types. These results suggest that applying a bulldog clamp at the fulcrum could potentially lead to inadequate vessel occlusion and intraoperative bleeding | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a hemostasis | |
650 | 4 | |a kidney neoplasms | |
650 | 4 | |a minimally invasive surgical procedures | |
650 | 4 | |a nephrectomy | |
650 | 4 | |a surgical instruments | |
700 | 1 | |a Myklak, Kristene |e verfasserin |4 aut | |
700 | 1 | |a Alsyouf, Muhannad |e verfasserin |4 aut | |
700 | 1 | |a Conceicao, Carol |e verfasserin |4 aut | |
700 | 1 | |a Peplinski, Brandon |e verfasserin |4 aut | |
700 | 1 | |a Arenas, Javier L |e verfasserin |4 aut | |
700 | 1 | |a Faaborg, Daniel |e verfasserin |4 aut | |
700 | 1 | |a Ruckle, Herbert C |e verfasserin |4 aut | |
700 | 1 | |a Baldwin, D Duane |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t The Journal of urology |d 1945 |g 195(2016), 3 vom: 01. März, Seite 756-62 |w (DE-627)NLM000006890 |x 1527-3792 |7 nnns |
773 | 1 | 8 | |g volume:195 |g year:2016 |g number:3 |g day:01 |g month:03 |g pages:756-62 |
856 | 4 | 0 | |u http://dx.doi.org/10.1016/j.juro.2015.09.080 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 195 |j 2016 |e 3 |b 01 |c 03 |h 756-62 |