Quality assurance multicenter comparison of different MR scanners for quantitative diffusion-weighted imaging

© 2015 Wiley Periodicals, Inc..

PURPOSE: To propose a magnetic resonance imaging (MRI) quality assurance procedure that can be used for multicenter comparison of different MR scanners for quantitative diffusion-weighted imaging (DWI).

MATERIALS AND METHODS: Twenty-six centers (35 MR scanners with field strengths: 1T, 1.5T, and 3T) were enrolled in the study. Two different DWI acquisition series (b-value ranges 0-1000 and 0-3000 s/mm(2) , respectively) were performed for each MR scanner. All DWI acquisitions were performed by using a cylindrical doped water phantom. Mean apparent diffusion coefficient (ADC) values as well as ADC values along each of the three main orthogonal directions of the diffusion gradients (x, y, and z) were calculated. Short-term repeatability of ADC measurement was evaluated for 26 MR scanners.

RESULTS: A good agreement was found between the nominal and measured mean ADC over all the centers. More than 80% of mean ADC measurements were within 5% from the nominal value, and the highest deviation and overall standard deviation were 9.3% and 3.5%, respectively. Short-term repeatability of ADC measurement was found <2.5% for all MR scanners.

CONCLUSION: A specific and widely accepted protocol for quality controls in DWI is still lacking. The DWI quality assurance protocol proposed in this study can be applied in order to assess the reliability of DWI-derived indices before tackling single- as well as multicenter studies.

Medienart:

E-Artikel

Erscheinungsjahr:

2016

Erschienen:

2016

Enthalten in:

Zur Gesamtaufnahme - volume:43

Enthalten in:

Journal of magnetic resonance imaging : JMRI - 43(2016), 1 vom: 15. Jan., Seite 213-9

Sprache:

Englisch

Beteiligte Personen:

Belli, Giacomo [VerfasserIn]
Busoni, Simone [VerfasserIn]
Ciccarone, Antonio [VerfasserIn]
Coniglio, Angela [VerfasserIn]
Esposito, Marco [VerfasserIn]
Giannelli, Marco [VerfasserIn]
Mazzoni, Lorenzo N [VerfasserIn]
Nocetti, Luca [VerfasserIn]
Sghedoni, Roberto [VerfasserIn]
Tarducci, Roberto [VerfasserIn]
Zatelli, Giovanna [VerfasserIn]
Anoja, Rosa A [VerfasserIn]
Belmonte, Gina [VerfasserIn]
Bertolino, Nicola [VerfasserIn]
Betti, Margherita [VerfasserIn]
Biagini, Cristiano [VerfasserIn]
Ciarmatori, Alberto [VerfasserIn]
Cretti, Fabiola [VerfasserIn]
Fabbri, Emma [VerfasserIn]
Fedeli, Luca [VerfasserIn]
Filice, Silvano [VerfasserIn]
Fulcheri, Christian P L [VerfasserIn]
Gasperi, Chiara [VerfasserIn]
Mangili, Paola A [VerfasserIn]
Mazzocchi, Silvia [VerfasserIn]
Meliadò, Gabriele [VerfasserIn]
Morzenti, Sabrina [VerfasserIn]
Noferini, Linhsia [VerfasserIn]
Oberhofer, Nadia [VerfasserIn]
Orsingher, Laura [VerfasserIn]
Paruccini, Nicoletta [VerfasserIn]
Princigalli, Goffredo [VerfasserIn]
Quattrocchi, Mariagrazia [VerfasserIn]
Rinaldi, Adele [VerfasserIn]
Scelfo, Danilo [VerfasserIn]
Freixas, Gloria Vilches [VerfasserIn]
Tenori, Leonardo [VerfasserIn]
Zucca, Ileana [VerfasserIn]
Luchinat, Claudio [VerfasserIn]
Gori, Cesare [VerfasserIn]
Gobbi, Gianni [VerfasserIn]
Italian Association of Physics in Medicine (AIFM) Working Group on MR Intercomparison [VerfasserIn]

Links:

Volltext

Themen:

Apparent diffusion coefficient
Comparative Study
Diffusion weighted imaging
Evaluation Study
Journal Article
Magnetic resonance imaging
Multicenter Study
Quality controls
Validation Study

Anmerkungen:

Date Completed 11.10.2016

Date Revised 10.12.2019

published: Print-Electronic

Citation Status MEDLINE

doi:

10.1002/jmri.24956

funding:

Förderinstitution / Projekttitel:

PPN (Katalog-ID):

NLM249369443