OBJECTIVES: In this study we aimed to evaluate the performance effects of chemiluminescence assay (CLIA) for Treponema pallidum specific antibodies detection, and to compare T. pallidum specific antibodies detection accuracy between CLIA and ELISA with TPPA (T. pallidum particle agglutination assay) as a confirmatory test.

METHODS: A total of 865 samples from suspected syphilis patients and preoperative patients were included, in which T. pallidum specific antibodies were simultaneously detected by CLIA and ELISA. Among them, 457 samples were determined by TPPA.

RESULTS: All coefficients of variation (CVs) of ELISA in high-, median-, and low-level samples were more than 5% and the maximum CV was 54.39% in the low-level sample. CVs of CLIA in different-level samples were all below 5%. Among the three assays the Spearman correlation and Kappa coefficients were 0.771 (P ≤ 0.001) and 0.854 (P ≤ 0.001, CLIA vs. ELISA), 0.806 (P ≤ 0.001) and 0.897 (P ≤ 0.001, ELISA vs. TPPA), 0.937 (P ≤ 0.001) and 0.967 (P ≤ 0.001, CLIA vs. TPPA), respectively. The area under the receiver operating characteristic curve (AUC) of CLIA was higher than that of ELISA (0.994 vs. 0.989) with TPPA as the confirmatory test. In 18 discrepant samples the consistency rate between CLIA and TPPA was elevated compared with that between ELISA and TPPA (72.22% vs. 27.78%, P = 0.008). In gray zone, the consistency rate of CLIA with TPPA was higher than that of ELISA with TPPA (90.91% vs. 41.67%, P = 0.027).

CONCLUSIONS: Compared with ELISA, CLIA is more reliable, sensitive and accurate to detect serum T. pallidum specific antibodies. In the future it may be an alternative test with higher sensitivity to ELISA.

Medienart:

E-Artikel

Erscheinungsjahr:

2016

Erschienen:

2016

Enthalten in:

Zur Gesamtaufnahme - volume:30

Enthalten in:

Journal of clinical laboratory analysis - 30(2016), 3 vom: 18. Mai, Seite 216-22

Sprache:

Englisch

Beteiligte Personen:

Li, Lixin [VerfasserIn]
Cai, Bei [VerfasserIn]
Tao, Chuanmin [VerfasserIn]
Wang, Lanlan [VerfasserIn]

Links:

Volltext

Themen:

Antibodies, Bacterial
Chemiluminescence assay (CLIA)
Enzyme immunoassay (EIA)
Evaluation Study
Journal Article
Syphilis
Treponema pallidum particle agglutination assay (TPPA)
Treponema pallidum specific antibodies

Anmerkungen:

Date Completed 13.02.2017

Date Revised 10.04.2022

published: Print-Electronic

Citation Status MEDLINE

doi:

10.1002/jcla.21839

funding:

Förderinstitution / Projekttitel:

PPN (Katalog-ID):

NLM246568186