Portable gas chromatography for determining airborne acrylonitrile in workplaces

OBJECTIVE: To establish a method for rapid determination of airborne acrylonitrile using a portable gas chromatograph.

METHODS: A single standard sample of acrylonitrile was prepared in a laboratory and sampled by the built-in constant flow pump of the portable gas chromatograph. The sample was then preconcentrated by the preconcentrator, thermally desorbed, separated by capillary columns, and detected by a micro argon ionization detector to determine the retention time. Retention time was then used to perform qualitative analysis. Under the set condition of gas chromatography, the external standard method was used to create a standard curve for quantitative analysis of acrylonitrile.

RESULTS: The linear range of acrylonitrile on the portable gas chromatograph was 0.25 to 3.00 mg/m(3). The regression equation was y = 10(-5) x-0.0275, r = 0.9977. The limit of detection was 0.005 mg/m(3), and the lower limit of quantification was 0.25 mg/m(3). The relative standard deviation was lower than 7.09%, and the degree of accuracy was 91.09-105.54%.

CONCLUSION: Portable gas chromatography is a simple, repeatable, and accurate method for rapid determination of airborne acrylonitrile.

Medienart:

Artikel

Erscheinungsjahr:

2013

Erschienen:

2013

Enthalten in:

Zur Gesamtaufnahme - volume:31

Enthalten in:

Zhonghua lao dong wei sheng zhi ye bing za zhi = Zhonghua laodong weisheng zhiyebing zazhi = Chinese journal of industrial hygiene and occupational diseases - 31(2013), 10 vom: 01. Okt., Seite 777-8

Sprache:

Chinesisch

Beteiligte Personen:

Liu, Jian [VerfasserIn]
Zhang, Feng [VerfasserIn]
Zhu, Bao-li [VerfasserIn]

Themen:

Acrylonitrile
Air Pollutants, Occupational
English Abstract
Journal Article
MP1U0D42PE

Anmerkungen:

Date Completed 22.04.2015

Date Revised 23.10.2013

published: Print

Citation Status MEDLINE

Förderinstitution / Projekttitel:

PPN (Katalog-ID):

NLM231915705