Agreement between optical coherence tomography and digital stereophotography in vertical cup-to-disc ratio measurement
BACKGROUND: Given the increasing popularity of optical coherence tomography in the diagnosis of glaucoma, we aimed to assess agreement between StratusOCT and stereophotography in measuring the vertical cup-to-disc ratio (VCDR) and examine whether it can be improved by changing the StratusOCT cup-offset.
METHODS: In group 1 (79 eyes), VCDR measurements obtained by StratusOCT (using the default cup-offset of 150 microm above the retinal pigment epithelium) and stereophotography were compared. For each eye, the cup-offset was moved upward or downward until the difference disappeared and an adjusted cup-offset was identified. In group 2 (41 eyes), the stereophotography VCDR was compared to the StratusOCT value obtained by using the mean adjusted cup-offset. StratusOCT was used to classify optic disc cups as shallow or deep.
RESULTS: In group 1, StratusOCT overestimated VCDR in optic discs with deep cups and underestimated it in those with shallow cups (p < 0.0001). Mean adjusted cup-offsets were 72.8 +/- 92 microm and 191.6 +/- 70 microm for eyes with deep and shallow cups. The 95% limits of agreement (LOA) ranged from -0.21 to 0.17. In group 2, the adjusted cup-offsets enabled us to eliminate the statistical differences in VCDR measurements and improve the 95% LOA (from -0.13 to 0.14).
CONCLUSION: Adjusting the StratusOCT cup-offset according to the cup depth improves agreement between StratusOCT and stereophotography in measuring the VCDR.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2009 |
---|---|
Erschienen: |
2009 |
Enthalten in: |
Zur Gesamtaufnahme - volume:247 |
---|---|
Enthalten in: |
Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie - 247(2009), 3 vom: 25. März, Seite 377-83 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Savini, Giacomo [VerfasserIn] |
---|
Links: |
---|
Themen: |
---|
Anmerkungen: |
Date Completed 25.03.2009 Date Revised 13.11.2018 published: Print-Electronic Citation Status MEDLINE |
---|
doi: |
10.1007/s00417-008-0968-3 |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM184034450 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM184034450 | ||
003 | DE-627 | ||
005 | 20231223165827.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231223s2009 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1007/s00417-008-0968-3 |2 doi | |
028 | 5 | 2 | |a pubmed24n0614.xml |
035 | |a (DE-627)NLM184034450 | ||
035 | |a (NLM)18958489 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Savini, Giacomo |e verfasserin |4 aut | |
245 | 1 | 0 | |a Agreement between optical coherence tomography and digital stereophotography in vertical cup-to-disc ratio measurement |
264 | 1 | |c 2009 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 25.03.2009 | ||
500 | |a Date Revised 13.11.2018 | ||
500 | |a published: Print-Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a BACKGROUND: Given the increasing popularity of optical coherence tomography in the diagnosis of glaucoma, we aimed to assess agreement between StratusOCT and stereophotography in measuring the vertical cup-to-disc ratio (VCDR) and examine whether it can be improved by changing the StratusOCT cup-offset | ||
520 | |a METHODS: In group 1 (79 eyes), VCDR measurements obtained by StratusOCT (using the default cup-offset of 150 microm above the retinal pigment epithelium) and stereophotography were compared. For each eye, the cup-offset was moved upward or downward until the difference disappeared and an adjusted cup-offset was identified. In group 2 (41 eyes), the stereophotography VCDR was compared to the StratusOCT value obtained by using the mean adjusted cup-offset. StratusOCT was used to classify optic disc cups as shallow or deep | ||
520 | |a RESULTS: In group 1, StratusOCT overestimated VCDR in optic discs with deep cups and underestimated it in those with shallow cups (p < 0.0001). Mean adjusted cup-offsets were 72.8 +/- 92 microm and 191.6 +/- 70 microm for eyes with deep and shallow cups. The 95% limits of agreement (LOA) ranged from -0.21 to 0.17. In group 2, the adjusted cup-offsets enabled us to eliminate the statistical differences in VCDR measurements and improve the 95% LOA (from -0.13 to 0.14) | ||
520 | |a CONCLUSION: Adjusting the StratusOCT cup-offset according to the cup depth improves agreement between StratusOCT and stereophotography in measuring the VCDR | ||
650 | 4 | |a Comparative Study | |
650 | 4 | |a Journal Article | |
700 | 1 | |a Espana, Edgar M |e verfasserin |4 aut | |
700 | 1 | |a Acosta, Ana Carolina |e verfasserin |4 aut | |
700 | 1 | |a Carbonelli, Michele |e verfasserin |4 aut | |
700 | 1 | |a Bellusci, Costantino |e verfasserin |4 aut | |
700 | 1 | |a Barboni, Piero |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie |d 1985 |g 247(2009), 3 vom: 25. März, Seite 377-83 |w (DE-627)NLM01296803X |x 1435-702X |7 nnns |
773 | 1 | 8 | |g volume:247 |g year:2009 |g number:3 |g day:25 |g month:03 |g pages:377-83 |
856 | 4 | 0 | |u http://dx.doi.org/10.1007/s00417-008-0968-3 |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 247 |j 2009 |e 3 |b 25 |c 03 |h 377-83 |