The study of aluminium diffuser calibration in the UV

A bi-directional reflectance distribution function (BRDF) measurement setup in the ultraviolet spectral range was established. The BRDF of the aluminium diffusers at a given orientation was measured. The relative accuracy of the BRDF measurement is better than 2.5%. The hemispheric reflectance of the aluminium diffusers was measured in the wavelength range from 250 to 650 nm. It increases with the wavelength, and changes about 6% from 300 to 360 nm. It decreases with the time. Since the diffuser was made (about one year ago), from 250 to 300 nm, the peak decrease in the hemispheric reflectance can reach 2.6%, and the average decrease is 1.5%. From 300 to 360 nm, it has an average decrease of 0.9% decrease, and 0.8% when wavelength is longer than 360 nm.

Medienart:

Artikel

Erscheinungsjahr:

2008

Erschienen:

2008

Enthalten in:

Zur Gesamtaufnahme - volume:28

Enthalten in:

Guang pu xue yu guang pu fen xi = Guang pu - 28(2008), 4 vom: 15. Apr., Seite 865-9

Sprache:

Chinesisch

Beteiligte Personen:

Li, Cong [VerfasserIn]
Wang, Yong-Mei [VerfasserIn]
Zhang, Zhong-Mou [VerfasserIn]

Themen:

English Abstract
Journal Article

Anmerkungen:

Date Completed 15.07.2010

Date Revised 14.07.2008

published: Print

Citation Status PubMed-not-MEDLINE

Förderinstitution / Projekttitel:

PPN (Katalog-ID):

NLM180829491