Synchrony dual-optic accommodating intraocular lens. Part 2 : pilot clinical evaluation
PURPOSE: To evaluate the clinical outcomes of an accommodating dual-optic intraocular lens (IOL).
SETTING: Private practice and university centers.
METHODS: A prospective noncomparative case series with retrospective control comprised 21 patients (26 eyes) scheduled for small-incision extracapsular cataract extraction by phacoemulsification with implantation of the Synchrony dual-optic accommodating IOL (Visiogen) (accommodating IOL group) and 10 patients who had small-incision extracapsular phacoemulsification with implantation of a monofocal, single-optic IOL at least 6 months previously (control group). Patients were examined 1, 3, 6, and 12 months after surgery. Defocus curves in the accommodating IOL group were compared with those in the control group. The main outcome measures were postoperative distance uncorrected and best corrected visual acuity; near uncorrected, distance corrected, and near corrected visual acuity; and accommodative range based on defocus curves.
RESULTS: Twenty-four eyes were available at the 6-month follow-up visit. All eyes had best corrected distance visual acuity of 20/40 or better, and 19 eyes (79%) had an uncorrected distance visual acuity of 20/40 or better. Uncorrected near visual acuity was 20/40 or better in all eyes. With distance correction, 23 eyes (96%) had an acuity of 20/40 or better at near. Defocus curve analysis suggested a mean accommodative range of 3.22 diopters (D) +/- 0.88 (SD) (range 1.00 to 5.00 D) in the accommodating IOL group and 1.65 +/- 0.58 D in the control group (range 1.00 to 2.50 D) (P<.05).
CONCLUSION: The Synchrony dual-optic IOL shows promise as an option to provide accommodative function in pseudophakic patients.
Errataetall: |
CommentIn: J Cataract Refract Surg. 2008 Jan;34(1):8-9; author reply 9. - PMID 18165063 |
---|---|
Medienart: |
Artikel |
Erscheinungsjahr: |
2007 |
---|---|
Erschienen: |
2007 |
Enthalten in: |
Zur Gesamtaufnahme - volume:33 |
---|---|
Enthalten in: |
Journal of cataract and refractive surgery - 33(2007), 1 vom: 15. Jan., Seite 47-52 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Ossma, Ivan L [VerfasserIn] |
---|
Themen: |
Clinical Trial |
---|
Anmerkungen: |
Date Completed 12.03.2007 Date Revised 16.03.2022 published: Print CommentIn: J Cataract Refract Surg. 2008 Jan;34(1):8-9; author reply 9. - PMID 18165063 Citation Status MEDLINE |
---|
Förderinstitution / Projekttitel: |
|
---|
PPN (Katalog-ID): |
NLM167385984 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM167385984 | ||
003 | DE-627 | ||
005 | 20231223112825.0 | ||
007 | tu | ||
008 | 231223s2007 xx ||||| 00| ||eng c | ||
028 | 5 | 2 | |a pubmed24n0558.xml |
035 | |a (DE-627)NLM167385984 | ||
035 | |a (NLM)17189792 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Ossma, Ivan L |e verfasserin |4 aut | |
245 | 1 | 0 | |a Synchrony dual-optic accommodating intraocular lens. Part 2 |b pilot clinical evaluation |
264 | 1 | |c 2007 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ohne Hilfsmittel zu benutzen |b n |2 rdamedia | ||
338 | |a Band |b nc |2 rdacarrier | ||
500 | |a Date Completed 12.03.2007 | ||
500 | |a Date Revised 16.03.2022 | ||
500 | |a published: Print | ||
500 | |a CommentIn: J Cataract Refract Surg. 2008 Jan;34(1):8-9; author reply 9. - PMID 18165063 | ||
500 | |a Citation Status MEDLINE | ||
520 | |a PURPOSE: To evaluate the clinical outcomes of an accommodating dual-optic intraocular lens (IOL) | ||
520 | |a SETTING: Private practice and university centers | ||
520 | |a METHODS: A prospective noncomparative case series with retrospective control comprised 21 patients (26 eyes) scheduled for small-incision extracapsular cataract extraction by phacoemulsification with implantation of the Synchrony dual-optic accommodating IOL (Visiogen) (accommodating IOL group) and 10 patients who had small-incision extracapsular phacoemulsification with implantation of a monofocal, single-optic IOL at least 6 months previously (control group). Patients were examined 1, 3, 6, and 12 months after surgery. Defocus curves in the accommodating IOL group were compared with those in the control group. The main outcome measures were postoperative distance uncorrected and best corrected visual acuity; near uncorrected, distance corrected, and near corrected visual acuity; and accommodative range based on defocus curves | ||
520 | |a RESULTS: Twenty-four eyes were available at the 6-month follow-up visit. All eyes had best corrected distance visual acuity of 20/40 or better, and 19 eyes (79%) had an uncorrected distance visual acuity of 20/40 or better. Uncorrected near visual acuity was 20/40 or better in all eyes. With distance correction, 23 eyes (96%) had an acuity of 20/40 or better at near. Defocus curve analysis suggested a mean accommodative range of 3.22 diopters (D) +/- 0.88 (SD) (range 1.00 to 5.00 D) in the accommodating IOL group and 1.65 +/- 0.58 D in the control group (range 1.00 to 2.50 D) (P<.05) | ||
520 | |a CONCLUSION: The Synchrony dual-optic IOL shows promise as an option to provide accommodative function in pseudophakic patients | ||
650 | 4 | |a Clinical Trial | |
650 | 4 | |a Journal Article | |
650 | 4 | |a Research Support, Non-U.S. Gov't | |
700 | 1 | |a Galvis, Andrea |e verfasserin |4 aut | |
700 | 1 | |a Vargas, Luis G |e verfasserin |4 aut | |
700 | 1 | |a Trager, Michelle J |e verfasserin |4 aut | |
700 | 1 | |a Vagefi, M Reza |e verfasserin |4 aut | |
700 | 1 | |a McLeod, Stephen D |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Journal of cataract and refractive surgery |d 1996 |g 33(2007), 1 vom: 15. Jan., Seite 47-52 |w (DE-627)NLM013278363 |x 1873-4502 |7 nnns |
773 | 1 | 8 | |g volume:33 |g year:2007 |g number:1 |g day:15 |g month:01 |g pages:47-52 |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 33 |j 2007 |e 1 |b 15 |c 01 |h 47-52 |